BD-S 56 drawing
BINDER ·Made in Germany · In Active Production

BINDER BD-S 56

BINDER BD-S 56 — Incubators with Temperature 5…70 °C.

The BD-S 56 is a 62 L natural-convection incubator with a stainless steel interior (1.4016) and PID controller, offering 5–70 °C operation with 0.1 °C display resolution. It is the smaller of two BD-S sizes (56 vs 115), distinguished by class 3.1 overtemperature protection and 0.3 K fluctuation. It serves culture incubation at 37 °C and other temperature-controlled laboratory processes.

Temperature
5…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD-S Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
BD-S 56 Current 5–7057.762.56240015
BD-S 115 5–7061.673.511855020

Full specifications

Core Performance & Functional Specifications
Temperature 5...70 °C
Interior volume 62 L
Load per rack 15 kg
Nominal power 0.3 kW
Permitted load 45 kg (BDS056-230V)
Temperature fluctuation 0.3 K
Temperature uniformity 0.5 K
Back wall clearance 160 mm (BDS056-230V)
Side wall clearance 100 mm (BDS056-230V)
Article Number mfr 9090
Recovery time after door was opened for 30 s at 37°C mfr 15 min
Dimensions, Weight & Mechanical Design
Depth 577 mm (BDS056-230V)
Height 625 mm (BDS056-230V)
Interior width 400 mm (BDS056-230V)
Width 614 mm (BDS056-230V)
Net weight of the unit (empty) mfr 35 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 6.3 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior and performance

Installation and control

Integration & Operating Notes

Electrical
Supply must be 230 V 1~ with PE and 6.3 A fusing

The 230 V variant requires a single-phase 230 V supply with protective earth, tolerance ±10 %, and a 6.3 A chamber fuse. Verify mains quality and plug type (CEE 7/7) before installation; incorrect voltage or missing earth can damage the controller or create a shock hazard.

Instruction Manual, chap. 4.3, p. 19-20; type plate, p. 8
Mounting
Maintain 160 mm rear and 100 mm side clearance

The chamber requires unobstructed air circulation for heat dispersal. Installing it in a recess or closer than the specified clearances risks overheating and reduced temperature performance. Ensure the floor or bench can support the 35 kg net weight plus 45 kg load.

Instruction Manual, chap. 3.4, p. 17; factory specification sheet
Process Limits
Do not load combustible or explosive materials

The chamber has no explosion protection. Charging materials must be harmless and their flash point or sublimation point must lie above the operating temperature. Corrosive substances (acids, halides) can damage the stainless steel interior and are excluded from liability.

Instruction Manual, chap. 1.9, p. 10-11
Control & I/O
Safety controller class 3.1 must be set before operation

The BD-S is equipped with an overtemperature safety device class 3.1 per DIN 12880. The safety controller set-point must be adjusted to the application to prevent excessive temperatures. Alarms must be acknowledged and the cause rectified before restarting.

Instruction Manual, chap. 9, p. 24
Maintenance
Annual electrical inspection and maintenance required

Regular maintenance at least once a year and repeat electrical tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.

Instruction Manual, chap. 12.1-12.2, p. 29-30
Maintenance
After cold storage, wait one hour before start-up

If the chamber was stored or transported below ambient temperature, condensation may form internally. Allow at least one hour at ambient temperature and ensure it is completely dry before connecting power to avoid electrical hazard or damage.

Instruction Manual, chap. 3.3, p. 17

Selection & Integration Q&A

When should I choose the BD-S 56 over the BD-S 115 or an ED-S/FD-S oven?

Choose the BD-S 56 for incubation at up to 70 °C with 62 L capacity and 0.1 °C display resolution; the BD-S 115 offers 115 L for larger loads. ED-S/FD-S ovens are for drying/heating up to higher temperatures but have 1 °C resolution and class 2 safety, not class 3.1.

What are the electrical supply and fusing requirements for the BD-S 56 230 V variant?

It requires a 230 V 1~ 50/60 Hz supply with protective earth, tolerance ±10 %, a CEE 7/7 plug, and a 6.3 A chamber fuse. Nominal power is 0.3 kW. Verify the type plate for the exact delivered configuration.

What installation clearances and ambient conditions must be respected?

Maintain 160 mm clearance at the rear and 100 mm at the sides. Do not install in unventilated recesses. Permissible storage temperature is -10 to +60 °C with max. 70 % r.h. non-condensing; after cold storage, wait one hour before start-up.

What maintenance and service parts are required, and what is not included in base delivery?

Annual maintenance and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. The door gasket is a wear part (replace only when cold). Racks and accessories are optional; the base delivery includes the chamber only. Disconnect power before service.

Dimensional Drawings & Technical Documentation

Dimensional drawings

BD-S 56 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model BD S 56 en
Datasheet ·399,8 КБ ·5 pp.
View PDF
PDF
BD ED FD S (E1) 06 2026 en
Manual ·10,8 МБ ·65 pp.
View PDF