BINDER CB 170
BINDER ·Made in Germany · In Active Production

BINDER CB 170

BINDER CB 170 — Incubators with Temperature 4…50 °C.

The BINDER CB 170 is a 170 L CO2 incubator with a sterilizable NDIR CO2 sensor and MB2 touchscreen controller, operating from +4 to 50 °C with 0.1 K fluctuation and 0.3 K uniformity. It is the mid-size chamber of the CB series, positioned between the 56 L CB 56 and 260 L CB 260, and is intended for cell culture and controlled atmosphere incubation requiring precise temperature and CO2 regulation.

Temperature
4…50 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER CB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmHumidity range, % rhInterior volume, LPermitted load, kg
CB 170 Current 4–5071.58790 ...9517030
CB 260 5–5078.510290...9526740
CB 56 4–5054.567.490...955330

Full specifications

Core Performance & Functional Specifications
Temperature 4...50 °C
Interior volume 170 L (CB 170)
Load per rack 10 kg (all variants)
Nominal power 1.4 kW (all variants)
Permitted load 30 kg (all variants)
Temperature fluctuation 0.1 ± K (all variants)
Temperature uniformity 0.3 K
Side wall clearance 50 mm (all variants)
Article Number mfr 9040
Dimensions, Weight & Mechanical Design
Depth 715 mm (all variants)
Height 870 mm (all variants)
Interior width 560 mm (all variants)
Net weight 95 kg (all variants)
Width 680 mm (all variants)
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz (all variants)
Rated voltage 200...230 V (CB170-230V variants)
Unit fuse 10.0 A (200...230 V variants)
Operating Conditions, Environment & Protection
Humidity range 90 ...95 % RH (all variants)

Description

Construction and Series Positioning

The CB 170 uses a 170 L interior (560 mm wide, 715 mm deep, 870 mm high) within a 680 mm wide, 95 kg cabinet. It is the middle size of the CB series, larger than the CB 56 and smaller than the CB 260. Unlike the CBF variants, this model does not include humidity control; the humidity range of 90–95 % RH is achieved via the Permadry water pan system. The standard model is the CB170-230V (article 9040-0199) for 200–230 V supply; left-hinged (-L) and O2-control (-O) variants are available but are separate modifications.

Behavior Under Load and Control

Temperature is controlled from +4 to 50 °C with a fluctuation of ±0.1 K and uniformity of 0.3 K. The chamber supports a permitted load of 30 kg and 10 kg per rack. The MB2 controller manages fixed-value, timer, and week programs with setpoint ramps and steps. CO2 control uses a sterilizable NDIR sensor; O2 control (0.2–20 vol.-% standard, 10–90 vol.-% optional) is only present on -O variants. During hot-air sterilization, CO2/O2 control is deactivated until the chamber cools below 80 °C and sterilization success is confirmed.

Installation and Operation

Side wall clearance of 50 mm is required. Electrical supply is 200–230 V, 50/60 Hz, 1.4 kW nominal, with a 10.0 A unit fuse. Gas connections for CO2 (and O2/N2 if equipped) require appropriate cylinder connection kits. The Permadry water pan must be filled for humidity. The DuoDoor door lock and touchscreen provide access and control. Alarm functions include a zero-voltage relay output and adjustable delay times for temperature, CO2, O2, and door-open events.

Integration & Operating Notes

Electrical
Verify supply voltage and fuse rating before connection

The CB 170 standard variant requires 200–230 V at 50/60 Hz and is protected by a 10.0 A unit fuse. Connecting to a 100–120 V supply without the UL variant will prevent operation and may damage components. Confirm the local supply matches the model code (CB170-230V vs CB170UL-120V) before installation.

Data Sheet Model CB 170 en.pdf; Operating Manual CB+CBF (E7.1) 01-2026, Technical Description
Control & I/O
CO2/O2 control inhibited during and after sterilization

During hot-air sterilization, CO2 and O2 control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas regulation during this period; plan cell culture workflows to avoid relying on CO2 control immediately after a sterilization cycle.

Operating Manual CB+CBF (E7.1) 01-2026, Section 21.3.5 Hot-air sterilization
Maintenance
Annual maintenance and electrical safety test required

Regular maintenance must be performed at least once a year, including electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. After 8760 operating hours or two years, a maintenance message appears and recurs every two weeks until reset by BINDER Service. Schedule service accordingly to avoid interrupted operation.

Operating Manual CB+CBF (E7.1) 01-2026, Section 22.1 and 22.5
Maintenance
Replace door gasket only when chamber is cold

The door gasket must be replaced only when the chamber is at room temperature. Attempting replacement while hot can deform or damage the gasket, leading to sealing failures and temperature/humidity deviations. Allow sufficient cooling time before service.

Operating Manual CB+CBF (E7.1) 01-2026, Section 22.5

Selection & Integration Q&A

When should I choose the CB 170 over the CB 56 or CB 260?

Choose CB 170 for 170 L capacity, balancing the smaller CB 56 and larger CB 260. It offers the same +4 to 50 °C range as CB 56 but more space; CB 260 extends to +5 to 50 °C with greater volume. Select based on sample throughput and available bench space.

What gas supply accessories are needed for CO2 control?

A CO2 gas cylinder connection kit is required (available as accessory). For O2-control variants, additional O2 and N2 supplies and kits are needed. Ensure gas inlet pressure is within specified limits and use appropriate pressure regulators.

What maintenance items and accessories are not included in the base delivery?

Base delivery excludes gas cylinder connection kits, Permadry water pan (for humidity), and optional accessories like APT-COM 4 software, silicone access ports, interior socket, and stacking adapters. Annual maintenance and electrical safety tests are user responsibilities; CO2 sensor replacement requires recalibration.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model CB 170 en
Datasheet ·426,3 КБ ·9 pp.
View PDF
PDF
CB+CBF (E7.1) 01 2026 en
Manual ·19,4 МБ ·201 pp.
View PDF