BINDER CBF 170
BINDER CBF 170 — Incubators with Temperature 6…50 °C.
The CBF 170 is a 170 L CO2 incubator with hot-air sterilization and active humidity regulation (50–95 % RH), operating from +6 to 50 °C with 0.1 K fluctuation and 0.4 K uniformity. It is the mid-size CBF variant (sibling CBF 260 offers larger volume) and is distinguished by integrated steam humidification and a 6-minute recovery time after a 30 s door opening at 37 °C. It serves cell culture processes requiring controlled CO2, humidity, and periodic 10-hour sterilization cycles.
Direct drop-in replacements
BINDER CBF Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Permitted load, kg | Width, cm |
|---|---|---|---|---|---|---|
| CBF 170 Current | 6–50 | 71.5 | 87 | 170 | 30 | 68 |
| CBF 260 | 6–50 | 78.5 | 102 | 267 | 40 | 74 |
Full specifications
| Temperature | 6...50 °C |
| CO2 recovery time | 6 min (all variants) |
| Interior volume | 170 L (CBF170, 50/60 Hz) |
| Load per rack | 10 kg (all variants) |
| Nominal power | 1.4 kW (all variants) |
| Permitted load | 30 kg (all variants) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.4 K |
| Back wall clearance | 100 mm (all variants) |
| Side wall clearance | 50 mm (all variants) |
| Article Number mfr | 9040 |
| Recovery time after door was opened for 30 s at 37°C mfr | 6 min |
| Depth | 715 mm (all variants) |
| Height | 870 mm (all variants) |
| Width | 680 mm (all variants) |
| Power frequency | 50/60 Hz (all variants) |
| Rated voltage | 200...230 V (CBF170-230V) |
| Unit fuse | 10.0 A (CBF170-230V) |
| Humidity range | 50...95 % RH |
Description
Construction and series positioning
- The CBF 170 is a floor-standing CO2 incubator with a 170 L interior, net dimensions 680 mm (W) × 870 mm (H) × 715 mm (D).
- Within the CBF series, it is the smaller of two sizes (CBF 170 and CBF 260); the CBF 260 provides greater volume but shares the same temperature range (+6 to 50 °C).
- Unlike the CB series, the CBF includes an active humidification system (50–95 % RH); the CB series relies on a passive water pan.
- The chamber is equipped with a sterilizable NDIR CO2 sensor and an MB2 touchscreen program controller. Hot-air sterilization is integrated, with a total cycle of 10 hours (4 h effective sterilization + 6 h cooling to 37 °C).
- Available variants include 200–230 V (CBF170-230V, article 9040-0211) and 100–120 V UL models, as well as O2-control (-O) versions with a 0.2–20 vol.-% range (optional 10–90 vol.-%). This card describes the standard 230 V model without O2 control.
Behavior under load and in different modes
- Temperature control: setpoint range +6 to 50 °C; short-term fluctuation 0.1 K and spatial uniformity 0.4 K support stable culture conditions.
- Humidity control: active steam humidification maintains 50–95 % RH; the system includes a dehumidification function for CBF/CBF-UL chambers.
- Recovery after door opening: 6 minutes to return to setpoint after a 30 s door opening at 37 °C, minimizing process interruption.
- Load limits: maximum 30 kg total permitted load, with 10 kg per rack; exceeding these limits may affect temperature uniformity and mechanical integrity.
- During hot-air sterilization, CO2 and humidity control are deactivated; they reactivate only after the chamber cools below 80 °C and the sterilization success message is confirmed.
Installation, connection, and control specifics
- Electrical: 200–230 V, 50/60 Hz, single-phase; nominal power 1.4 kW; unit fuse 10.0 A. A CEE 7/7 plug is provided on the standard 230 V variant.
- Clearances: maintain at least 100 mm at the rear and 50 mm at the sides for ventilation and access.
- Gas supply: CO2 (and O2/N2 for -O variants) connections require appropriate gas cylinder connection kits (accessory). Gas inlet fine filters are available for replacement.
- Water supply for humidification: can be provided via prefilled water bottle (BINDER-Water-Set), freshwater bag (tubular bag set), or freshwater can (water container set) — all accessories.
- Controller: MB2 touchscreen with program controller; supports fixed-value mode, timer programs, week programs, alarm functions, and user management. Network interfaces include RS485 and Ethernet; USB for data transfer and firmware updates.
- Safety: over-temperature protective device (class 1) and overtemperature safety controller class 3.1 are standard; a zero-voltage relay alarm output is available for external alarms.
Integration & Operating Notes
The standard CBF170-230V requires 200–230 V, 50/60 Hz, single-phase supply with a 10.0 A fuse. Connecting to a different voltage (e.g., 100–120 V) will damage the unit. UL variants are available for 100–120 V markets. Ensure the outlet is properly grounded and protected.
The permitted load is 30 kg for the entire chamber and 10 kg per rack. Overloading can compromise temperature uniformity, damage shelves, and affect door sealing. Distribute weight evenly and use appropriate racks for heavy items.
During the hot-air sterilization cycle, CO2 and humidity control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas or humidity control during this period; plan cultures accordingly.
Regular maintenance must be performed at least once a year, including the steam humidifier. An electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after any maintenance or repair. A maintenance message appears after 8760 operating hours or two years and cannot be permanently cleared by the user; BINDER Service must reset it.
The CBF 170 does not include a water supply in base delivery. Humidification requires an external water source such as a BINDER-Water-Set, freshwater bag set, or water container set. Water quality affects humidifier maintenance intervals; use appropriate water to minimize scale.
Selection & Integration Q&A
Choose CBF 170 for 170 L capacity with active humidity control (50–95 % RH) and hot-air sterilization. The CBF 260 offers larger volume (260 L) for higher throughput. CB series models lack active humidification and are suitable for dry CO2 incubation without humidity regulation.
CO2 supply requires a gas cylinder connection kit (accessory) and a pressure regulator. For humidification, an external water source is needed: BINDER-Water-Set, freshwater bag set, or water container set. Ensure gas inlet pressure is within specified limits and use clean water to prevent scale.
Annual maintenance including humidifier service and electrical safety test is required. A maintenance reminder appears after 8760 h or 2 years. Base delivery excludes gas cylinder connection kits, water supply sets, and the BINDER-Water-Set; these must be ordered separately. CO2 sensor replacement requires recalibration.