BINDER ED 260
BINDER ED 260 — Drying ovens with Temperature 5…300 °C.
The BINDER ED 260 is a 255 L natural-convection drying and heating oven (ED260-230V, Art. 9010-0339) rated +5 to 300 °C with 2.25 kW single-phase heating. Its 810×940×760 mm interior and 40 kg/rack load capacity distinguish it from smaller ED 23/56/115 siblings, while the 3×/h air change suits gentle drying and heat treatment of solid or bulk materials.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 3 x/h |
| Heating time to 150°C | 50 min |
| Interior volume | 255 L |
| Load per rack | 40 kg (ED260-230V / ED260UL-240V) |
| Nominal power | 2.25 kW |
| Permitted load | 270 kg (ED260-230V / ED260UL-240V) |
| Temperature fluctuation | 0.8 K |
| Temperature uniformity | 1.8 K |
| Rear wall clearance | 160 mm (ED260-230V / ED260UL-240V) |
| Side wall clearance | 100 mm (ED260-230V / ED260UL-240V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 20 min |
| Depth | 760 mm (ED260-230V / ED260UL-240V) |
| Height | 940 mm (ED260-230V / ED260UL-240V) |
| Width | 810 mm (ED260-230V / ED260UL-240V) |
| Net weight of the unit (empty) mfr | 85 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 12.5 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Natural-convection oven with stainless steel 1.4301 (AISI 304) inner chamber, pre-heating chamber and door interiors; external surfaces powder-coated.
- 255 L usable volume in a 810 mm W × 940 mm H × 760 mm D net interior; 85 kg empty weight.
- Positioned between the ED 115 (smaller bench-scale) and ED 400 (larger floor unit) in the ED series; all share the +5 to 300 °C range, but the ED 260 is the mid-size single-phase model (230 V 1N~, 2.25 kW, 12.5 A external fuse).
- No fan: airflow relies on thermal convection, giving a gentle, low-turbulence drying environment (3 air changes per hour) suited to powders and bulk materials that must not be dispersed.
Thermal behavior and control
- Temperature fluctuation 0.8 K and uniformity 1.8 K across the usable volume; heating from ambient to 150 °C takes 50 min, and recovery after a 30 s door opening at 150 °C is 20 min.
- Controller display resolution is 1 °C (ED series); set-point entry via membrane keypad with ramp and timer functions.
- Overtemperature protection: class 1 device trips at 330 °C; safety controller class 2 (DIN 12880) with adjustable set-point 10–300 °C, to be set 2–5 °C above the working set-point.
- No fan-speed adjustment (unlike BF/FED); operation is purely set-point temperature and timer.
Installation and operation
- Requires 230 V 1N~ 50/60 Hz supply within ±10 %; IEC connector with shockproof plug (CEE 7/7) on the 230 V variant; unit fuse 12.5 A.
- Minimum clearances: 160 mm rear, 100 mm side; install indoors on a non-flammable surface, ambient +18 to +40 °C, ≤70 % r.h. non-condensing, max altitude 2000 m.
- Load racks up to 40 kg each; total permitted load 270 kg; keep samples ≥10 % of interior dimensions from walls and load ≤50 % of usable volume for uniform temperature.
- Hot-air sterilization possible at 190 °C for ≥30 min; racks removable and autoclavable.
Integration & Operating Notes
At temperatures above 150 °C, stainless steel interior parts can develop a harmless oxidation tint. This does not affect function or safety and requires no corrective action; avoid abrasive cleaning that could damage the passive layer.
Selection & Integration Q&A
Pick the ED 260 for 255 L batches needing up to 300 °C with single-phase 230 V supply and 2.25 kW heating. The ED 115 suits smaller loads on the same voltage; the ED 400 requires 400 V 3N~ and is for larger volumes. All share the +5 to 300 °C range and natural convection.
It needs 230 V 1N~ 50/60 Hz within ±10 %, 2.25 kW nominal, with a 12.5 A time-lag external fuse. The 230 V variant uses an IEC connector with CEE 7/7 shockproof plug; ensure the circuit and outlet are rated for at least 12.5 A and are not shared with high-draw equipment.
Schedule annual maintenance and electrical safety tests per EN 50678/50699 by qualified personnel. Recommended cleaning agents: neutral cleaner Art. 1002-0016 and disinfectant spray Art. 1002-0022. Optional RS 422/APT-COM 4, analog temperature output, and audible over-temperature alarm are not included in base delivery.