BINDER ED 720
BINDER ED 720 — Drying ovens with Temperature 5…300 °C.
The ED 720 is a 743-litre natural-convection drying and heating oven in BINDER's ED series, built as a two-door chamber with a 5–300 °C working range and a 4.1 kW three-phase 400 V supply. It is the largest ED size, distinguished from the 23/115/260/400 siblings by its 400 V 3N~ connection, internal 16 A breaker and fixed 1800 mm power cable. It serves bulk drying, tempering and heat treatment of solid or pulverized material where forced convection is not required.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 2 x/h |
| Heating time to 150°C | 85 min |
| Interior volume | 743 L |
| Load per rack | 45 kg |
| Nominal power | 4.1 kW |
| Permitted load | 315 kg |
| Temperature fluctuation | 0.8 K |
| Temperature uniformity | 3.2 K |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 25 min |
| Depth | 870 mm |
| Height | 1590 mm |
| Width | 1165 mm |
| Net weight of the unit (empty) mfr | 169 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and series positioning
- The ED 720 is the size-720 member of the ED natural-convection line; interior, pre-heating chamber and door interiors are stainless steel 1.4301 (AISI 304).
- Unlike the 230 V single-phase ED 23/115/260, this modification runs on 400 V 3N~/50 Hz with 4.1 kW nominal power and an internal 16 A miniature circuit breaker instead of an external chamber fuse.
- It is a two-door chamber; the maximum adjustable temperature set-point depends on ambient temperature, so the 300 °C ceiling is not unconditional.
- Net weight 169 kg empty; permitted load 315 kg, 45 kg per rack.
Behaviour under load and in operation
- Natural convection (no fan) gives an air-change rate of about 2 x/h; temperature uniformity is 3.2 K and fluctuation 0.8 K, so do not expect forced-convection homogeneity.
- Heating from cold to 150 °C takes 85 min; after a 30 s door opening at 150 °C the chamber recovers in about 25 min.
- Controller display resolves to 1 degree on ED ovens; the safety controller set-point range is 10–300 °C at 1-degree accuracy.
- Above 150 °C metallic surfaces may discolour by oxidation; this is harmless and does not impair function.
Installation, connection and control
- Fixed power cable of at least 1800 mm; supply tolerance ±10 % of nominal voltage; pollution degree 2, over-voltage category II, IP 20.
- Ambient limits: +18 to +40 °C, max 70 % r.h. non-condensing, installation altitude max 2000 m; not for unventilated recesses or explosive areas.
- Controller supports temperature set-point with ramp and timer functions, °C/°F change and chamber addressing; RS 422 with APT-COM 4 and analog temperature output are options.
- Overtemperature protection is class 1 with a class 2 safety controller (DIN 12880); class 3.1 is available as an option for ED/FD.
Integration & Operating Notes
Samples must stay inside the usable volume with at least 10 % of the interior dimensions as clearance to the walls, and the usable volume must not be loaded beyond half. With natural convection only, overloading blocks airflow and destroys the 3.2 K uniformity; respect the 45 kg per rack and 315 kg total load limits.
Only harmless materials may be processed; the charge must not be explosive or flammable, must not form an explosive mixture with air, and the operating temperature must remain below its flash point or sublimation point. The chamber is not qualified for sterilization of medical devices.
Selection & Integration Q&A
Pick the ED 720 when you need the 743 L interior and 315 kg total load for bulk batches. It is also the only ED size on 400 V 3N~ with 4.1 kW and an internal 16 A breaker; the 23/115/260 run on 230 V 1N~. If your batch fits a smaller chamber, the single-phase sizes avoid a three-phase feed.
It requires 400 V 3N~/50 Hz within ±10 %, 4.1 kW nominal, protected by its internal 16 A miniature circuit breaker rather than an external chamber fuse. A fixed power connection cable of at least 1800 mm is factory-fitted, so plan the installation position so the cable reaches the supply; use only original BINDER connection cables.
Install indoors on a non-flammable surface at +18 to +40 °C, max 70 % r.h. non-condensing, below 2000 m altitude, with 160 mm rear and 100 mm side clearance and no unventilated recess. Note that on this two-door 720 the maximum adjustable temperature set-point depends on ambient temperature.
Annual maintenance by BINDER-authorized personnel and yearly electrical testing per EN 50678/EN 50699 are required but not part of delivery. Options include the disconnectable audible over-temperature alarm, APT-COM 4 software with RS 422, analog temperature output and safety class 3.1; cleaning agent 1002-0016 and disinfectant 1002-0022 are ordered separately.