BINDER MKF 400 CO₂
BINDER ·Made in Germany · In Active Production

BINDER MKF 400 CO₂

BINDER MKF 400 CO₂ — Climate chambers.

The MKF 400 CO₂ is a 417 L alternating climate chamber with a working range of -50 to 180 °C and 10–98 % RH, built for 400 V/50 Hz three-phase supply. It is the mid-size CO₂-refrigerant variant of the MKF (E6) series, distinguished from the MKF 240 and MKF 1020 by its 417 L interior and 30 kg per-rack load rating. It serves cyclic temperature-humidity testing in a single programmable chamber.

BINDER MKF Series — Model Comparison & Selection

Model Temperature, °CHeight, cmInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
MKF 400 CO₂ Current -50...180172.74173016007.9
MKF 1020 -40–180200.599140150012.5
MKF 1020 CO₂ -48–180200.599140140011.5
MKF 115 -40–180172.5115304004.8
MKF 240 -40–180171.5228304006.8
MKF 240 CO₂ -50...180171.3228309006.8
MKF 400 -40–180171417308008
MKF 56 -40–180144.560152502.8
MKF 720 -40–180200.573440100011
MKF 720 CO₂ -48–180200.573440110011

Full specifications

Core Performance & Functional Specifications
Temperature -50...180
Interior volume 417 L (MKF400CO2-400V / MKF400-480V)
Load per rack 30 kg (MKF400CO2-400V / MKF400-480V)
Maximum heat compensation 1600 W
Nominal power 7.9 kW (MKF400CO2-400V / MKF400-480V)
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 1400 mm (MKF400CO2-400V / MKF400-480V)
Height 1727 mm (MKF400CO2-400V / MKF400-480V)
Width 1115 mm (MKF400CO2-400V / MKF400-480V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz (MKF400CO2-400V)
Rated voltage 400 V
Unit fuse 16.0 A (MKF400CO2-400V / MKF400-480V)
Operating Conditions, Environment & Protection
Humidity range 10...98 % RH
Cooling down time from 180°C to -40°C mfr 80 min
Humidity fluctuation with humidity dependent on set value mfr 2 % r.F.
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.7 K/min

Description

Construction and series positioning

Thermal and humidity behavior

Control, installation and operation

Integration & Operating Notes

Electrical
Supply must be protected by a type B RCD

The chamber label and manual require operation only with residual current circuit breakers type B; a standard type A or AC RCD is not acceptable because of the CO₂ refrigeration circuit's DC/leakage characteristics. Verify the building distribution before commissioning.

Instruction Manual, Section 1.5
Electrical
Voltage must stay within ±10 % of nominal at 50 Hz

The nominal voltage tolerance is ±10 % at the indicated power frequency. With a 7.9 kW nominal load and 16.0 A unit fuse, sustained deviation outside this band risks compressor and controller faults; size the feeder and upstream protection from the actual unit type plate.

Instruction Manual, Section 1.6 Type plate
Mounting
Room volume must be verified per EN 378-1

Because the refrigeration circuit uses R 744 (CO₂) at up to 120 bar, the installation room volume must be calculated according to EN 378-1:2016+A1:2020. Do not install in unventilated recesses and observe the minimum distances in chapter 3.4 to disperse heat.

Instruction Manual, Section 1.8
Process Limits
Humidity sensor tolerates only listed solvent concentrations

The humidity sensor has tabulated maximum workplace threshold and permanent-load concentrations for substances such as ammonia, acetone, ethanol, formaldehyde, ozone and sulphur dioxide. Test specimens outgassing above these limits will degrade the sensor; check the material list before loading.

Instruction Manual, Section 1.14
Process Limits
Humidity operation narrows the usable temperature band

With humidity active the chamber is limited to 10–95 °C and 10–98 % RH; below 10 °C or above 95 °C humidity control must be switched off. Program setpoints accordingly to avoid condensation or humidifier faults.

Factory Specification Sheet; Instruction Manual, Section 10 Start up
Control & I/O
Safety controller class 2.0 must be set and function-checked

The over-temperature safety device class 2.0 per DIN 12880 requires its setpoint and alarm-state measures to be configured and its function checked per Section 12 before programs run. The optional class 2 over/under device adds under-temperature protection for sensitive specimens.

Instruction Manual, Section 12 Temperature Safety Devices
Maintenance
Service message recurs every two weeks until BINDER reset

After 8760 operating hours or two years the controller displays a service message; after confirmation it reappears every two weeks until reset by BINDER Service. Plan the annual maintenance and the BINDER reset together to avoid recurring interruptions.

Instruction Manual, Section 23.2
Maintenance
Refrigeration repairs are BINDER-only; log all work

Repairs to the CO₂ refrigeration circuit may currently only be carried out at BINDER, and all refrigeration work must be documented in a service log book by personnel trained per EN 13313:2010. Condensers should be cleaned every 1–2 years and the steam humidifier at least annually, with frequency set by water quality.

Instruction Manual, Sections 23.1 and 23.2

Selection & Integration Q&A

When should I choose the MKF 400 CO₂ rather than the MKF 240 or MKF 1020?

Pick the 400 when 417 L and 30 kg per rack fit your specimen batch; the 240 is smaller, the 1020 larger. The CO₂ designation is the refrigerant, not a gas function, so all MKF variants share the same humidity principle.

What supply and protection does the 400 V variant need at the wall?

3 N/PE 400 V 50 Hz, 7.9 kW nominal, 16.0 A unit fuse, voltage within ±10 %, and a type B RCD. Size the feeder from the actual unit type plate, since nominal current can differ from the catalogue value.

What water connections and room conditions must be in place before start-up?

Provide a freshwater pipe connection kit or fill the internal can manually, plus a wastewater drain. The room volume must satisfy EN 378-1, with ventilation and minimum clearances per chapter 3.4; unventilated recesses are prohibited.

What maintenance and accessories are not included in the base delivery?

Annual maintenance, condenser cleaning every 1–2 years and yearly steam-humidifier service are user-scheduled; the RS485 interface, analog outputs, Pt 100 object sensor, safety kit and APT-COM 4 are options. The two-year/8760-hour service message needs a BINDER reset.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model MKF CO₂ 400 en
Datasheet ·833,0 КБ ·9 pp.
View PDF
PDF
0715 EN MKF CO2 E6 UM 10 2026
Manual ·11,3 МБ ·155 pp.
View PDF