BINDER MKT 240
BINDER ·Made in Germany · In Active Production

BINDER MKT 240

BINDER MKT 240 — Climate chambers with Temperature -70…180 °C.

The BINDER MKT 240 is a 228 L alternating climate chamber with a -70 to 180 °C working range, using a two-stage cascade refrigeration system (R452a/R23) for rapid temperature changes. It is the mid-size MKT variant, positioned between the MKT 115 and MKT 720, and is distinguished by its 400 V three-phase supply and 6.5 kW nominal power. It is intended for dynamic temperature cycling of components and materials in laboratory and industrial test environments.

Temperature
-70…180 °C

BINDER MKT Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior height, mmInterior volume, LInterior width, mm
MKT 240 Current -70–18092.5193.5700228735
MKT 115 -70–18086.5172.5480115600
MKT 720 -70–180114200.510207341200

Full specifications

Core Performance & Functional Specifications
Temperature -70...180 °C
Interior volume 228 L
Load per rack 30 kg (MKT240-400V, 50 Hz)
Nominal power 6.5 kW
Article Number mfr 9020
Dimensions, Weight & Mechanical Design
Depth 925 mm (MKT240-400V, 50 Hz)
Height 1935 mm (MKT240-400V, 50 Hz)
Interior height 700 mm (MKT240-400V, 50 Hz)
Interior width 735 mm (MKT240-400V, 50 Hz)
Width 1115 mm (MKT240-400V, 50 Hz)
Net weight of the unit (empty) mfr 380 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 400 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 3 N/PE
Operating Conditions, Environment & Protection
Cooling down time from 180°C to -70°C mfr 95 min
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 5.0 K/min

Description

Construction and series positioning

Behavior under load and in different modes

Installation, connection, and operation

Integration & Operating Notes

Electrical
Requires 400 V three-phase supply with 16 A fusing

The chamber draws 11.3 A nominal current at 400 V 3~/N/PE 50 Hz and must be protected by a 16.0 A unit fuse. Verify that the site supply can deliver 6.5 kW nominal power and that the over-current protection is sized accordingly; a voltage deviation beyond ±10% may affect performance and safety.

Type plate example, Chapter 1.6, page 11
Mounting
Install with prescribed clearances for heat dissipation

The chamber must not be installed in unventilated recesses. Sufficient ventilation and the minimum distances specified in chapter 3.4 of the operating manual are required to disperse heat from the refrigeration system and avoid overheating damage.

Chapter 1.8, page 12
Process Limits
Do not operate in explosive or wet environments

The chamber has no explosion protection and must not be operated in potentially explosive areas or near combustible dust/air-solvent mixtures. Water ingress creates a deadly electric shock hazard; keep electrical parts dry during operation, cleaning, and maintenance.

Chapter 1.8, page 13
Control & I/O
Class 1 safety trip indicates chamber failure

The over temperature protective device class 1 responds if the nominal temperature is exceeded by 20 °C due to chamber failure. A class 1 trip requires the chamber to be taken out of operation and BINDER Service informed; do not reset and continue without investigating the cause.

Chapter 12.1, page 76; chapter 22.3, page 120
Maintenance
Plan service at 8760 operating hours or two years

A maintenance message appears after 8760 operating hours or two years. After confirmation, the message window reappears every two weeks until reset by BINDER Service. Schedule service intervention at or before this interval to reset the counter and avoid recurring messages.

Chapter 22.2, page 120
Maintenance
Annual electrical safety inspection is mandatory

An annual repeat electrical inspection is required, as well as a test prior to initial startup and prior to resuming operation after maintenance or repair. Testing under EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended. Document the results.

Chapter 22.1, page 118
Maintenance
On-site staff limited to simple troubleshooting

Simple troubleshooting by operating personnel does not require technical intervention into the chamber or disassembly of chamber parts. Detailed troubleshooting and repair must be performed by BINDER Service or BINDER-qualified service partners per the Service Manual.

Chapter 22.1 and 22.3, pages 118 and 120

Selection & Integration Q&A

When should I choose the MKT 240 over the MKT 115 or MKT 720?

Choose the MKT 240 when you need a 228 L interior volume and a -70 to 180 °C range with 5.0 K/min heating and 4.2 K/min cooling. The MKT 115 offers smaller capacity, while the MKT 720 provides larger capacity; all share the same temperature range. Select based on sample size and required throughput.

What installation clearances and environmental limits apply?

Install the chamber with the minimum distances specified in chapter 3.4 of the manual to allow heat dispersal; do not install in unventilated recesses. Avoid potentially explosive areas, combustible dust/air-solvent mixtures, and damp locations. The chamber must not become wet during operation or cleaning.

What maintenance is required and what accessories are not included?

A maintenance message appears after 8760 operating hours or two years, requiring BINDER Service to reset. Annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is mandatory. Base delivery excludes options such as RS485 interface, analog temperature outputs, Pt 100 object sensor, compressed air connection/dryer, and APT-COM™ 4 software.

Dimensional Drawings & Technical Documentation

Dimensional drawings

MKT 240 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model MKT 240 en
Datasheet ·646,5 КБ ·8 pp.
View PDF
PDF
MK+MKT (E5) 10 2026 en
Manual ·15,9 МБ ·169 pp.
View PDF