Peter Huber Kältemaschinenbau HB120 Umwälz-Heizgerät
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau HB120 Umwälz-Heizgerät

Peter Huber Kältemaschinenbau HB120 Umwälz-Heizgerät — Thermostatic baths.

The HB120 is a three-phase, magnetically coupled stainless-steel circulation heater for external open circuits, operating from 60 to 250 °C with a Pilot ONE controller. It has no expansion tank and no active refrigeration, so the application must provide filling, volume compensation and a heat loss between 0.1 and 2.8 kW to reach the 60 °C minimum.

Direct drop-in replacements

Eppendorf
MixMate ®
71%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Umwaelz Heizgeraete Series — Model Comparison & Selection

Model Dimensions w d hPower supplyPump connectionPump typeHeating capacity, WAmbient temperature limit, °C
HB120 Umwälz-Heizgerät Current 479 x 539 x 695380-460V 3~ 50/60HzM30x1, 5 AG malemagnetically coupled, st—5–40
HB240 Umwälz-Heizgerät 570 x 700 x 1171400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st240005–40
HB480 Umwälz-Heizgerät 570 x 700 x 1171400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st480005–40
HB60 Umwälz-Heizgerät 323 x 451 x 498400V 3~N 50HzM30x1, 5 AG malePressure pump, speed not60005–40
HB720 Umwälz-Heizgerät 800 x 1007 x 1607400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st720005–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 479 x 539 x 695 mm
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 60...250 °C
Dimensions, Weight & Mechanical Design
Weight 88 kg (net)
Electrical, Power & Drive Specifications
Current consumption 20 A (max.)
Power supply 380-460V 3~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class II / FL

Description

Construction and series positioning

The HB120 is the mid-capacity member of the HB series (HB45, HB60, HB120, HB240, HB480). It is a circulation heating unit, not a refrigerated thermostat: the working range is 60–250 °C, and the 60 °C lower limit is reached only when ambient temperature is below 60 °C and the application heat loss falls within the model-dependent 0.1–2.8 kW window. The pump is magnetically coupled and stainless steel, with adjustable speed and a maximum delivery of 100 l/min at 3.5 bar. The hydraulic connection is an M30x1.5 male thread. The unit is built for external open operation only and has no expansion tank, so filling and volume compensation must come from the application.

Behavior under load and media

Because there is no active cooling, the achievable minimum temperature is set by the balance between heater input, ambient temperature and the heat loss of the connected circuit. The adjustable overtemperature protection to DIN 12876 must be configured for the thermal fluid and application before commissioning. The controller is a Pilot ONE with 0.01 K display resolution and an external Pt100 sensor connection, allowing the control point to be taken at the application rather than at the unit.

Electrical and control integration

The unit requires a 380–460 V three-phase 50/60 Hz supply with a maximum current of 20 A and 3x25 A fusing. Voltage tolerance is ±10% only if the frequency tolerance does not act in the opposite direction. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this. Three-phase devices below 63 A are delivered with a cable but without a plug. Interfaces include Ethernet, USB host and device, and RS232. The protection rating is IP20, and the equipment is classified Class A, Group 1 to EN55011, which restricts use in residential environments.

Integration & Operating Notes

Process Limits
External circuit must supply filling and expansion volume

The HB120 has no expansion tank and is designed exclusively for external open operation. The engineer must provide a reservoir or open bath with sufficient volume compensation, otherwise thermal expansion of the fluid has nowhere to go and the circuit cannot be filled or vented correctly.

HB120 Heating Bath Datasheet, Overview
Process Limits
60 °C minimum depends on ambient and application heat loss

Reaching 60 °C is not guaranteed by the unit alone: ambient temperature must be below 60 °C and the heat loss of the connected application must lie between 0.1 and 2.8 kW depending on the HB model. A circuit with higher losses will not settle at the minimum temperature, so the load budget must be checked during selection.

HB120 Heating Bath Datasheet, Specifications
Electrical
Leakage current above 3.5 mA dictates earthing and RCD choice

The datasheet states a leakage current greater than 3.5 mA. Protective earthing must be sized accordingly, and residual-current devices must be selected for this leakage level rather than a generic 30 mA assumption without verification.

HB120 Heating Bath Datasheet, Specifications
Electrical
Three-phase unit arrives with cable but no plug

Three-phase devices with current consumption below 63 A are supplied with a cable but without a plug. The installer must provide a suitable plug and a 3x25 A protected connection matching the 380–460 V 50/60 Hz supply.

HB120 Heating Bath Datasheet, Specifications
Process Limits
Overtemperature protection must be set before first heat-up

The unit has adjustable overtemperature protection according to DIN 12876. Commissioning personnel must configure the limit for the specific thermal fluid and application; leaving it at a default risks either nuisance trips or inadequate protection of the fluid and circuit.

HB120 Heating Bath Datasheet, Specifications
Process Limits
External circuit pressure drop must fit 100 l/min at 3.5 bar

The magnetically coupled stainless-steel pump delivers up to 100 l/min and 3.5 bar, with adjustable speed. The pressure drop of hoses, adapters and the application must be checked against this capability; a circuit with higher resistance will not reach the required flow.

HB120 Heating Bath Datasheet, Specifications
Mounting
Mating hoses need M30x1.5 male-thread adapters

The pump connection is an M30x1.5 male thread. Hoses, unions and adapters for the external circuit must match this thread; standard laboratory fittings will not connect without a suitable adapter.

HB120 Heating Bath Datasheet, Specifications
Electrical
Class A Group 1 emission limits restrict residential use

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011. This has implications for use in residential or light-industrial environments, where additional filtering or separation from sensitive instrumentation may be required.

HB120 Heating Bath Datasheet, Specifications

Selection & Integration Q&A

When should I choose the HB120 rather than the HB45, HB60, HB240 or HB480?

Pick the HB120 when the external circuit's heat loss and flow demand sit in the middle of the HB range; the smaller HB45/HB60 suit lower heat-loss loads, while the HB240/HB480 are for larger duties. All share the 60–250 °C range and external-open-only architecture.

What must the external circuit and its fittings provide to work with this unit?

An open reservoir with filling and volume compensation (no expansion tank on the unit), hoses or adapters matching the M30x1.5 male pump thread, and a circuit pressure drop compatible with 100 l/min at 3.5 bar. A Pt100 sensor can be placed in the application.

What is not included in base delivery and must be planned for?

The three-phase cable is supplied but no plug, so the installer provides the plug and protected connection. The external reservoir, volume compensation, hoses and M30x1.5 adapters, and the Pt100 sensor for the application are not part of the unit and must be sourced separately.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
HB120 Heating Bath Datasheet
Datasheet ·17,8 КБ ·2 pp.
View PDF