Peter Huber Kältemaschinenbau HB480 Umwälz-Heizgerät
Peter Huber Kältemaschinenbau HB480 Umwälz-Heizgerät — Thermostatic baths.
The HB480 is a 48 kW external open-loop circulation heater with a magnetically coupled stainless-steel pump (200 l/min, 5.5 bar) and a 60–250 °C working range. It has no expansion tank or active refrigeration, so it serves high-temperature external circuits where the application itself provides fluid filling, volume compensation and heat loss.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Umwaelz Heizgeraete Series — Model Comparison & Selection
| Model | Dimensions w d h | Power supply | Pump connection | Pump type | Heating capacity, W | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| HB480 Umwälz-Heizgerät Current | 570 x 700 x 1171 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 48000 | 5–40 |
| HB120 Umwälz-Heizgerät | 479 x 539 x 695 | 380-460V 3~ 50/60Hz | M30x1, 5 AG male | magnetically coupled, st | — | 5–40 |
| HB240 Umwälz-Heizgerät | 570 x 700 x 1171 | 400V 3~ 50Hz | M30x1, 5 AG male | magnetically coupled, st | 24000 | 5–40 |
| HB60 Umwälz-Heizgerät | 323 x 451 x 498 | 400V 3~N 50Hz | M30x1, 5 AG male | Pressure pump, speed not | 6000 | 5–40 |
| HB720 Umwälz-Heizgerät | 800 x 1007 x 1607 | 400V 3~ 50Hz | M38x1, 5 AG male | magnetically coupled, st | 72000 | 5–40 |
Full specifications
| Dimensions (w × D × H) | 570 x 700 x 1171 mm |
| Heating capacity | 48 kW |
| Pump connection | M38x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | 60...250 °C |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | II / FL |
Description
Construction and series positioning
- The HB480 is the largest heating-capacity member of the HB series (siblings: HB45, HB60, HB120, HB240), built as an external open-operation circulator without an expansion tank.
- Heating power is 48 kW; the pump is a variable-speed, magnetically coupled stainless-steel pressure pump with M38x1.5 male connections.
- Control is via the Pilot ONE controller with 0.01 K display resolution; an external Pt100 sensor connection is provided.
- Electrical supply is 400 V 3~ 50 Hz; protection rating IP20; safety class II / FL.
Behavior in operation
- Working range is 60–250 °C. Because there is no active refrigeration, reaching the 60 °C minimum depends on ambient temperature below 60 °C and on application heat loss between 0.1 kW and 2.8 kW (model-dependent).
- The pump delivers up to 200 l/min at up to 5.5 bar, setting the hydraulic envelope for the external circuit.
- No digitized performance curves are supplied; heat-up and control behavior must be assessed from the 48 kW heating capacity and the loop pressure drop.
Installation and control
- The unit is intended exclusively for external open operation; the application must provide fluid filling and volume compensation.
- Three-phase devices below 63 A are delivered with a cable but without a plug; the customer supplies the plug and connection.
- Interfaces: Ethernet, USB (host and device), RS232.
- EMC classification is Class A, Group 1 (EN 55011), directing installation toward an industrial environment.
- Leakage current from external temperature control must be considered in earthing and RCD selection.
Integration & Operating Notes
Without active refrigeration, the 60 °C lower limit is only reachable when ambient is below 60 °C and the application dissipates between 0.1 kW and 2.8 kW depending on the HB model; a loop with lower or higher loss will not settle at the minimum temperature.
The unit is supplied with a cable but without a plug because three-phase devices drawing less than 63 A are delivered this way; the site must provide a matching 400 V 3~ 50 Hz plug and connection hardware.
Tolerance combinations that run in opposite directions are not permitted — for example -10% voltage together with +3% frequency is explicitly disallowed — so site power quality must be checked against this rule before connection.
EN 55011 Class A Group 1 classification typically requires an industrial environment; in residential or light-industrial areas additional filtering or shielding may be needed to meet disturbance limits.
Leakage current from the external temperature control circuit must be factored into protective earthing design and residual current device selection; undersized RCDs may nuisance-trip or fail to provide intended protection.
Layout must leave the space around the unit according to the operating conditions published by the manufacturer; inadequate clearance affects ventilation and access for service.
Selection & Integration Q&A
Pick the HB480 when the external circuit needs the series' highest heating capacity (48 kW) and the 60–250 °C range; the smaller HB45/HB60/HB120/HB240 suit lower heat loads. All share the no-expansion-tank external open-operation architecture, so the choice is driven by heating power and loop heat loss.
The loop must match the M38x1.5 male pump connection and stay within the pump envelope of 200 l/min at 5.5 bar. Power is 400 V 3~ 50 Hz; since units below 63 A ship with cable but no plug, the site supplies the plug. EMC Class A Group 1 favors an industrial installation.
Ambient must stay within 5–40 °C, and the 60 °C minimum temperature is only reachable with ambient below 60 °C and application heat loss between 0.1 kW and 2.8 kW depending on model. Voltage and frequency tolerances must not oppose, and space requirements around the unit must be respected.
The unit ships with a cable but without a plug for sub-63 A three-phase supply, and it has no expansion tank, so the application must provide fluid filling and volume compensation. External Pt100 sensing is supported via the dedicated sensor connection; plan earthing and RCD selection around the external temperature-control leakage current.