Peter Huber Kältemaschinenbau KISS K25 Kälte-Badthermostat
Peter Huber Kältemaschinenbau KISS K25 Kälte-Badthermostat — Thermostatic baths with Noise level 50 dB.
The KISS K25 is a refrigerated bath thermostat with a 23.5 l bath and a non-variable pressure/suction pump, covering -30 to 200 °C with ±0.05 K stability. It is the largest bath in the KISS series up to 30 °C, distinguished by its 15 l minimum filling volume and 0.4 kW cooling at 20 °C. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory and pilot processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau KISS Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| KISS K25 Kälte-Badthermostat Current | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 39 | 0.015 |
| KISS K12 Kälte-Badthermostat | — | 110-115V 1~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| KISS K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.01 |
| KISS K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
| KISS K6s Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
Full specifications
| Noise level | 50 dB(A) |
| Filling volume | 15 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump max; 0, 25 bar) |
| Pump connection | (Option, #19607) (Option) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Refrigerator | air-cooled, natural refrigerant |
| Temperature range | -30...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 350 x 555 x 615 mm |
| Heating capacity mfr | 1, 8 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min; 0, 17 bar |
| Bath opening WxD/BD mfr | 290 x 329 / 200 |
| Bath volume mfr | 23, 5 l |
| Weight | 39 kg |
| Power supply | 220-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 2 A (max) |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Refrigerant | R-290 (A3, H220) (1st stage) |
| Refrigerant quantity 1st stage mfr | 0, 072 kg (1st stage) |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Cooling capacity mfr | 20: 0, 4; 0: 0, 35; -10: 0, 27; -20: 0, 16; -30: |
| Safety class | III / FL |
Description
Construction and series positioning
The KISS K25 is an air-cooled refrigerated bath thermostat using R-290 (A3, H220) in the first stage, with a 23.5 l bath volume and a minimum filling volume of 15 l. Its bath opening measures 290 x 329 mm with a 200 mm depth. Within the KISS series up to 30 °C, the K25 is the largest bath variant listed alongside the K6, K12, K15 and K20; the K20 shares the 50 dB(A) noise level, while the K6 is louder at 57 dB(A). The K25 is therefore selected when bath volume and immersion space matter more than the smaller footprints of the K12 or K15.
Thermal behavior under load
Cooling capacity is strongly temperature-dependent: 0.4 kW at 20 °C, 0.35 kW at 0 °C, 0.27 kW at -10 °C, 0.16 kW at -20 °C and 0.065 kW at -30 °C. Heat loads must be sized against the actual set point rather than the 20 °C figure. Heating capacity is 1.8–2.1 kW. Temperature stability is ±0.05 K across the -30 to 200 °C range, and the unit provides adjustable overtemperature protection to DIN 12876 with safety class III/FL. The pressure and suction pump is not speed-variable; maximum delivery is 14 l/min at 0.25 bar on the pressure side and 10.5 l/min at 0.17 bar on the suction side. A pump connection is available as an option (#19607).
Installation, control and operation
The controller is the KISS type with an OLED display and USB (Device) plus RS232 interfaces; alarms are optic and acoustic. Power supply is 220–240 V, 1~ or 2~, 50/60 Hz, with maximum current consumption of 10 + 2 A; the circuit must be fused between 10 A and 16 A. The enclosure is IP20, so a dry, protected location is required. Ambient temperature must remain between 5 and 40 °C, and cooling capacity falls as ambient temperature rises. Voltage and frequency tolerances are ±5% and ±2%, but the combination of -5% voltage with +2% frequency is not permitted. The unit is Class A, Group 1 to EN55011. Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering. The plastic pump is not resistant to acetone or polyglycols; for sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.
Integration & Operating Notes
Upstream overcurrent protection must be sized within the 10 A to 16 A window; the unit draws up to 10 + 2 A maximum. A breaker outside this range either trips during run-up or fails to protect the compressor circuit.
The standard plastic pump is not resistant to acetone or polyglycols. Bath fluid selection must avoid these media, or a more resistant pump must be specified, otherwise wetted pump parts will be attacked.
Heat loads must be sized against the actual set point: capacity is 0.4 kW at 20 °C but only 0.065 kW at -30 °C. Oversizing the process load against the 20 °C figure will prevent the bath from reaching low set points.
The IP20 rating means the unit must be installed away from water spray and in a dry area. Ambient temperature must remain between 5 and 40 °C, and cooling capacity drops as ambient temperature rises, so climate control or derating may be needed.
Single- and two-phase devices are delivered with a power cable and a country-specific plug. The plug type must be stated when ordering, otherwise the unit may arrive with a plug incompatible with the site's electrical system.
Selection & Integration Q&A
Choose the K25 when you need the largest bath in the series: 23.5 l bath volume, 15 l minimum filling and a 290 x 329 mm opening. The K12, K15 and K20 offer smaller footprints; the K20 shares the 50 dB(A) noise level, while the K6 is louder at 57 dB(A).
Supply must be 220–240 V, 1~ or 2~, 50/60 Hz, with ±5% voltage and ±2% frequency tolerance; the combination of -5% voltage and +2% frequency is not allowed. Fuse the circuit between 10 A and 16 A, and specify the country-specific plug at ordering.
Install in a dry, protected location: the enclosure is IP20, so avoid water spray. Ambient temperature must stay between 5 and 40 °C, and cooling capacity falls as ambient temperature rises, so provide climate control or derate the cooling load accordingly.
The base delivery includes the power cable with country-specific plug only when specified at ordering; the pump connection is an option (#19607). For sub-zero operation, glysantine or ethylene glycol must be added to the bath fluid; the plastic pump is not resistant to acetone or polyglycols, so fluid choice is a consumable decision.