Peter Huber Kältemaschinenbau KISS K6s Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau KISS K6s Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau KISS K6s Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.

The KISS K6s is an air-cooled refrigerated bath circulator with a 4.5 L bath and a non-variable pressure/suction pump, covering -25 to 200 °C with ±0.05 K stability. It is the smallest refrigerated model in the KISS series up to 30 °C, distinguished by its compact 210 mm width and 0.05 kW cooling capacity at -20 °C. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory processes.

Noise level
57 dB

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau KISS Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
KISS K6s Kälte-Badumwälzthermostat Current 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–200250.002
KISS K12 Kälte-Badthermostat —110-115V 1~ 50/60Hz(Option, #19607)-20–20028—
KISS K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.01
KISS K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200390.015
KISS K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–200250.002

Full specifications

Core Performance & Functional Specifications
Noise level 57 dB(A) (sound pressure level)
Filling volume 2 l (min)
Maximum delivery flow rate 14 l/min (pressure pump max, 0, 25 bar)
Pump connection M16x1 AG male
Pump speed adjustable no
Pump type Pressure and suction pump, speed not variable
Temperature range -25...200 °C
Temperature stability mfr ±0, 05 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 210 x 400 x 546 mm
Heating capacity mfr 1, 6 - 2, 1 kW
Delivery capacity suction pump max. mfr 10, 5 l/min; 0, 17 bar
Bath opening WxD/BD mfr 140 x 120 / 150 mm
Bath volume mfr 4, 5 l
Dimensions, Weight & Mechanical Design
Weight 25 kg (net)
Electrical, Power & Drive Specifications
Power supply 208-240V 1~/2~ 50/60Hz
Current consumption max. mfr 10 + 1, 3 A (max)
Interfaces, Connections & Control
Controller KISS
Digital interface USB (Device), RS232 Interface
Temperature display OLED Display
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C (permissible)
Protection rating IP20
Cooling capacity mfr 20:0, 26; 0:0, 21; -10:0, 15; -20:0, 05 °C / kW (at 20/0/-10/-20 °C)
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal behavior and media

Pump, connection and control

Integration & Operating Notes

Electrical
Supply tolerance pairs are restricted; verify before connection

The unit accepts 208–240 V 1~/2~ 50/60 Hz with a 10 A minimum and 16 A maximum fuse, but voltage and frequency tolerances are linked: ±5% voltage is permitted only with ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are explicitly not allowed. Confirm the site supply against this matrix before commissioning.

KISS K6s Technical Data, page 2
Mounting
Air-cooled condenser needs unobstructed airflow and dry location

Because the refrigerated bath is air-cooled, it requires adequate space around the unit per the manufacturer's space requirements; the IP20 rating means it is not protected against water ingress and must be kept dry. Ambient temperature must remain within 5 to 40 °C, and cooling capacity is rated at 20 °C room temperature and may drop as ambient rises.

KISS K6s Technical Data, pages 1–2
Process Limits
Plastic pump limits fluid choice; avoid acetone and polyglycols

The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these fluids require a more resistant plastic option. For low-temperature operation, water should be mixed with glysantine or ethylene glycol for freeze protection to avoid freezing damage in the bath or circuit.

KISS K6s Technical Data, page 2
Safety
Autostart may restart the unit after power returns

The controller includes an autostart function for power failure, meaning the unit may resume operation automatically when power is restored. Risk assessments and procedures for unattended operation must account for this behavior, alongside the adjustable overtemperature protection per DIN 12876 (safety class III / FL) and optic/acoustic alarms.

KISS K6s Technical Data, page 1
Process Limits
Cooling capacity at -20 °C is only 0.05 kW

At the low end of the -25 to 200 °C range, available cooling is 0.05 kW at -20 °C, falling from 0.15 kW at -10 °C. Process heat loads near this temperature must be kept well below this figure, or the bath will not hold setpoint; the unit is not suitable for high heat loads near its minimum temperature.

KISS K6s Technical Data, page 1
Electrical
Class A EMC classification may affect sensitive lab environments

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1. In residential or EMC-sensitive laboratory settings, this classification may require additional consideration of adjacent sensitive instrumentation or shielding measures.

KISS K6s Technical Data, page 2

Selection & Integration Q&A

When should I choose the KISS K6s rather than a larger sibling such as the K12 or K20?

Choose the K6s for the smallest footprint (210 mm wide, 4.5 L bath) and lowest cooling demand; its 0.05 kW at -20 °C suits light loads only. Larger siblings K12/K15/K20/K25 provide more bath volume, and the K20/K25 are quieter at 50 dB(A) versus 57 dB(A).

What pump connections and accessories are needed to integrate the K6s into an external circuit?

The pump uses M16x1 male connections (NW8/12). Included accessories are hose connector NW12 and sleeve nuts thread M16x1; the pressure/suction pump delivers up to 14 L/min at 0.25 bar pressure and 10.5 L/min at 0.17 bar suction, with no speed adjustment.

What site conditions and duty limits apply to installing the K6s?

Install within 5–40 °C ambient with adequate airflow for the air-cooled condenser, keep dry (IP20), and respect flammable-refrigerant R-290 requirements. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises; supply must be 208–240 V with 10–16 A fuse.

What maintenance and consumable items are not included with the base unit?

The base delivery includes hose connector NW12 and M16x1 sleeve nuts but not heat-transfer fluid; water should be mixed with glysantine or ethylene glycol for freeze protection. The plastic pump is incompatible with acetone and polyglycols unless a resistant plastic option is ordered.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
KISS K6s Technical Data
Datasheet ·136,0 КБ ·2 pp.
View PDF