BD ED FD S (E1) 06 2026

Technical Reference Summary

Covered Models & Document Scope

Covers Solid.Line™ BD-S / BD-S-UL incubators with natural convection, ED-S / ED-S-UL drying and heating ovens with natural convection, and FD-S / FD-S-UL drying and heating ovens with forced convection, all with R-S microprocessor temperature controller. Model sizes 56 and 115, in 230 V (non-UL) and 120 V (UL) versions. Art. Nos.: BD-S 56 9090-0016/9190-0016, BD-S-UL 56 9090-0017/9190-0017, BD-S 115 9090-0022/9190-0022, BD-S-UL 115 9090-0023/9190-0023, ED-S 56 9090-0014/9190-0014, ED-S-UL 56 9090-0015/9190-0015, ED-S 115 9090-0020/9190-0020, ED-S-UL 115 9090-0021/9190-0021, FD-S 56 9090-0018/9190-0018, FD-S-UL 56 9090-0019/9190-0019, FD-S 115 9090-0024/9190-0024, FD-S-UL 115 9090-0025/9190-0025.

Section-by-Section Technical Breakdown

pp. 14
Chamber description

Describes the Solid.Line chamber concept: electronic PID controller with digital display, natural convection for BD-S/ED-S and forced-air circulation for FD-S, stainless steel interior (1.4016 / AISI 430), polyester-coated housing RAL 9003, and built-in overtemperature protection. Temperature display resolution is 0.1 °C for BD-S incubators and 1 °C for ED-S/FD-S ovens.

  • BD-S/BD-S-UL indicate temperature to a tenth of a degree; ED-S/ED-S-UL and FD-S/FD-S-UL to one degree.
  • BD-S and ED-S are naturally ventilated; FD-S uses fan-assisted forced-air circulation.
  • Inner chamber and door interior: stainless steel 1.4016 (AISI 430); housing organically coated polyester RAL 9003.
  • Overtemperature safety device class 1 per DIN 12880:2007 plus safety controller class 2 (ED-S/FD-S) or class 3.1 (BD-S).
  • Above 150 °C / 302 °F, ED-S/FD-S metallic surfaces may discolour by oxidation; this is harmless.
pp. 19-20
Installation

Covers rack installation, optional exhaust/ventilation connection, and electrical connection. Electrical connection tables specify plug type, current type, and chamber fuse for each model and voltage variant.

  • 230 V models: grounded plug, 1N~, 6.3 A chamber fuse.
  • UL 120 V models: NEMA 5-15P plug, 1N~, 12.5 A chamber fuse.
  • Observe local/national regulations when connecting.
  • Optional connection to an exhaust/ventilation system is described.
pp. 16, 21-22
R-S controller overview

Introduces the R-S microprocessor temperature controller with digital display and functional buttons on the triangular instrument panel, and gives the menu structure overview.

  • Controller display (6) and functional controller buttons (7) on triangular instrument panel.
  • Menu structure overview provided for navigation of controller functions.
pp. 22
Start up and adjusting air change

Start-up procedure and adjustment of the air change rate via the exhaust air flap.

  • Air change is adjusted at start-up (chap. 6.1).
pp. 23
Temperature set-point entry

Describes entering the temperature set-point on the R-S controller.

  • Set-point entry via controller buttons (chap. 7).
pp. 23
Selecting the temperature unit

Describes how to set the displayed temperature unit (°C/°F) on the controller.

  • Temperature unit setting procedure in chap. 8.1.
pp. 24-25
Overtemperature protection

Describes the overtemperature protective device class 1 and the safety controller (class 2 for ED-S/FD-S, class 3.1 for BD-S), including setting the safety controller and responding to alarms.

  • Overtemperature protective device class 1 per DIN 12880:2007.
  • Safety controller class 2 (ED-S/ED-S-UL, FD-S/FD-S-UL) or class 3.1 (BD-S/BD-S-UL).
  • Alarm message handling and proceeding in case of an alarm (chap. 9.2.2).
pp. 26
Timer function “delayed off”

Describes the delayed-off timer function and how to set the timer run-time.

  • Timer run-time setting in chap. 10.1.
pp. 26-29
Cleaning and decontamination

Covers cleaning and chemical disinfection/decontamination of the chamber, including precautions for disinfectant spray and drying/aeration after use.

  • Wear sealed protective goggles when using disinfectant spray.
  • Do NOT empty disinfectant into drains.
  • Allow chamber to dry thoroughly and aerate sufficiently after disinfectant use.
pp. 29-30
Maintenance and service

Defines personnel qualification, maintenance intervals, and electrical testing requirements. Regular maintenance at least once a year; annual repeat electrical inspection and tests before initial start-up and after maintenance/repair.

  • Maintenance at least once a year by licensed electricians or BINDER-authorized experts.
  • Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 recommended.
  • Disconnect power plug before maintenance; do NOT remove rear panel.
  • Replace door gasket only when cold.
  • Warranty void if maintenance performed by non-authorized personnel.
pp. 30
Simple troubleshooting

Fault table for operating personnel covering chamber without function with dark display or with “OFF” on lower display, listing possible causes and required measures.

  • Chamber without function, controller display dark: check power supply; overtemperature protective device class 1 may have tripped; controller may be defective.
  • Chamber without function, controller shows “OFF” on lower display and actual temperature on upper display: chamber in standby mode; press confirmation button.
  • Only BINDER-authorized qualified service personnel may perform repair.
pp. 35-44
Technical description

Covers factory calibration and adjustment, definition of usable volume, overcurrent protection, technical data, equipment/options, accessories/spare parts, and dimensioned drawings for all model sizes.

  • Factory calibration and adjustment (chap. 14.1).
  • Definition of usable volume (chap. 14.2).
  • Overcurrent protection (chap. 14.3).
  • Technical data (chap. 14.4).
  • Equipment and options extract (chap. 14.5); accessories and spare parts extract (chap. 14.6).
  • Dimensioned drawings for BD-S 56/115, ED-S 56/115, FD-S 56/115 (chap. 14.7).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
230 V models (BD-S/ED-S/FD-S 56 and 115) use a grounded plug, 1N~ current type, and 6.3 A chamber fuse. p.4 ↗ Provide a 230 V single-phase supply with protective earth and 6.3 A fusing per chamber. BD-S 56/115, ED-S 56/115, FD-S 56/115 (230 V versions)
UL 120 V models (BD-S-UL/ED-S-UL/FD-S-UL 56 and 115) use a NEMA 5-15P plug, 1N~ current type, and 12.5 A chamber fuse. p.4 ↗ Provide a 120 V single-phase NEMA 5-15 receptacle with 12.5 A fusing per chamber. BD-S-UL 56/115, ED-S-UL 56/115, FD-S-UL 56/115
Nominal voltage tolerance is ±10 % at the indicated power frequency (230 V / 50 Hz or 230 V / 60 Hz). p.8 ↗ Supply voltage must remain within ±10 % of nominal; verify mains quality before installation. 230 V models
Type plate example BD-S 115: nominal power 1.30 kW, nominal current 5.7 A, nominal temperature 70 °C / 158 °F, IP protection 20, temperature safety device class 3.1 per DIN 12880. p.1 ↗ Use these values for electrical supply sizing and for verifying the delivered configuration against the type plate. BD-S 115 (example type plate)
Overtemperature safety device class 1 per DIN 12880:2007 is fitted as standard; safety controller is class 2 for ED-S/FD-S and class 3.1 for BD-S. p.2 ↗ Select and configure the safety controller class according to chamber type; class 3.1 on BD-S provides the higher protection level. whole series
Temperature display resolution: 0.1 °C for BD-S/BD-S-UL incubators; 1 °C for ED-S/ED-S-UL and FD-S/FD-S-UL ovens. p.2 ↗ Do not expect sub-degree readout on ED-S/FD-S; use BD-S where tenth-degree indication is required. BD-S vs ED-S/FD-S
Inner chamber and door interior are stainless steel 1.4016 (AISI 430); housing is polyester-coated RAL 9003. p.2 ↗ Avoid corrosive charging materials (acids, halides) that can damage stainless steel; corrosion damage is excluded from liability. whole series
At operating temperatures above 150 °C / 302 °F, ED-S/FD-S metallic surfaces may discolour (yellowish-brown or blue) by natural oxidation. p.2 ↗ Discolouration is harmless and does not impair function or quality; no corrective action required. ED-S/ED-S-UL, FD-S/FD-S-UL
Permissible ambient temperature range during transport and storage: -10 °C to +60 °C / 14 °F to 140 °F; storage humidity max. 70 % r.h., non-condensing. p.3 ↗ Respect these limits for shipping and intermediate storage; after cold storage, wait at least one hour at ambient temperature before start-up. whole series
After storage in a cold location, wait at least one hour before start-up until the chamber has attained ambient temperature and is completely dry. p.3 ↗ Prevents condensation-related electrical hazard and damage at commissioning. whole series
Lift the chamber at its four lower corners with 2 people; do NOT lift or transport using the door handle or door. p.3 ↗ Prevents injury and chamber damage during unpacking and relocation. whole series
Regular maintenance at least once a year; annual repeat electrical inspection and tests before initial start-up and after maintenance/repair, recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.12 ↗ Plan annual service and electrical safety testing; document results to satisfy legal requirements. whole series
Disconnect the power plug before maintenance; do NOT remove the rear panel; chamber must NOT become wet during operation or maintenance. p.12 ↗ Live maintenance is prohibited; electrical hazard from water ingress. whole series
Replace the door gasket only when cold. p.12 ↗ Hot gasket replacement may damage the gasket. whole series
Chambers are produced per VDE regulations and routinely tested per VDE 0411-1 (IEC 61010-1); UL chambers certified to UL 61010-2-10 and CAN/CSA-C22.2 No. 61010-2-10. p.1 ↗ Electrical safety compliance basis for acceptance testing and inspection. whole series; UL chambers for UL/CSA standards
Do NOT install the chamber in unventilated recesses; observe prescribed minimum distances (chap. 3.4) and ensure sufficient ventilation for heat dispersal. p.1 ↗ Prevents overheating damage; installation clearance must be respected. whole series
The chamber has no explosion protection measures; do not introduce combustible or explosive substances, and operating temperature must lie below the flash point or sublimation point of the charging material. p.1 ↗ Intended use is limited to harmless materials; verify material properties before loading. whole series
BD-S incubators are especially suited for incubation of cultures at a standard temperature of 37 °C / 98.6 °F. p.1 ↗ Typical application set-point for BD-S; use precise temperature accuracy of the incubator line. BD-S/BD-S-UL

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Chamber without function. Controller display is dark. p.30 ↗ No power supply. Check connection to power supply. chap. 12.3
Chamber without function. Controller display is dark. p.30 ↗ Overtemperature protective device class 1 has turned off the chamber. Contact BINDER service. chap. 12.3
Chamber without function. Controller display is dark. p.30 ↗ Controller defective. Contact BINDER service. chap. 12.3
Chamber without function. Controller shows “OFF” on lower display and actual temperature value on upper display. p.30 ↗ Chamber in standby mode. Press the confirmation button. chap. 12.3