BINDER LED Pflanzenlichtmodul (E7) 09 2025
Technical Reference Summary
Covered Models & Document Scope
BINDER LED Plant Light Module, model LED-M (E7), Art. no. 8012-2439, operating manual art. no. 7001-0610, version 09/2025. Accessory for BINDER cooling incubators series KB PRO (E7) and constant climate chambers series KBF / KBF-UL / KBF PRO (E7); also usable as stand-alone solution with other cooling incubators/constant climate chambers if placement, cable routing and heat compensation are suitable. Includes plant light box, 16 LED light bars, 4 connection cables (2 m), 4 distributors, ventilation hose, power supply cable, 60 °C maximum temperature sticker, 2 mounting rails with screws (4 x M8, 2 x M5), and 2 silicone foam plugs.
Section-by-Section Technical Breakdown
The LED plant light module is a stand-alone accessory with its own light controller, independent of the chamber controller. It provides three independently adjustable light channels (cool white, warm white, far red) via 16 freely positionable LED bars mounted on standard racks. A diaphragm pump ventilates the chamber with 8 L/min ambient air to prevent CO2 drop from photosynthesis.
- Three light channels: cool white (6,500 K), warm white (2,700 K), far red (730 nm), adjustable in any ratio
- 16 LED light bars with clips, positionable on one, two or four levels
- Independent controller with resistive touch display, timer and dusk functions, and ventilation control
- Diaphragm pump supplies 8 L/min ambient air via hose connection
- Can be used simultaneously with the BINDER CO2 control module accessory
- Plant light box housing RAL 9003 powder-coated; on/off switch illuminates during operation; screen enters standby after 5 minutes inactivity
Covers mounting the plant light box on the chamber, attaching the maximum temperature sticker, fastening LED bars on racks, connecting LED bar connectors to distributors, connectors on the plant light box, inserting silicone foam plugs, and electrical connection. The accessory is for indoor use only and must not be installed in unventilated recesses or explosive areas.
- Mounting rails with 4 x M8 and 2 x M5 screws are supplied for installing the plant light box
- 60 °C maximum temperature sticker must be attached to the front of the plant light box and to the chamber door
- LED bars are fastened to standard racks with clips and connected to 4 distributors, each for 4 LED bars
- Plant light box connections: IEC connector plug, on/off switch, touch controller, 4 LED bar connectors, fresh-air hose connection 'Out'
- Silicone foam plugs are inserted into the elongated access port
- Electrical supply: 120-240 V wide voltage, 50/60 Hz, 1 N ~
Describes standby mode, functional overview, setting system time and weekday, adjusting fresh air supply, setting light channels, adjusting light intensity, displaying current lighting status, and activating lighting. The controller is independent of the chamber controller.
- System time and weekday must be set for timer and program functions
- Fresh air supply is adjustable via the ventilation control
- Light channels can be set individually and light intensity adjusted
- Current lighting status is displayed on the touch screen
- Lighting is activated via the controller; on/off switch illuminates during operation
Covers switching the program function on/off, the day/night simulation program, the week program, and the twilight function. These allow automatic simulation of natural light conditions.
- Program function can be switched on and off
- Day/night simulation program for automatic light cycles
- Week program for scheduling light sequences across the week
- Twilight function for gradual light transitions
Provides instructions for cleaning and chemical disinfection/decontamination of the accessory. Cleaning must only be performed with the chamber and plant light box switched off and power plugs disconnected.
- Turn off cooling incubator/climate chamber and plant light box and disconnect power plugs before cleaning
- Do not allow water to penetrate the LED plant light box
- Use appropriate cleaning and decontamination agents only
Covers maintenance intervals, replacement of LED light bars, battery status warning, troubleshooting, and return of the accessory to BINDER. Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; electrical repeat inspection is required annually and after maintenance or repair.
- Maintenance at least once a year by licensed electricians or experts authorized by BINDER
- Annual electrical repeat inspection and test before initial startup and after maintenance/repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702
- LED light bars recommended for replacement after approx. 25,000 operating hours and at the latest after 6 years
- Battery for system time: CR1225 button cell; low battery warning below 2.5 V shown as flashing battery icon; re-enter system time and weekday after replacement
- Simple troubleshooting by operating personnel per chap. 8.3; detailed troubleshooting and repair by BINDER Service or qualified service partners
- Return requires an RMA authorization number issued before sending the product
Technical data from the type plate: nominal temperature 60 °C / 140 °F, IP protection 20, nominal power 0.25 kW, nominal current 2.1 A, nominal voltage range 120-240 V at 50 Hz and 60 Hz, single-phase 1 N ~. The unit is CE and UKCA marked and ETL Listed.
- Nominal temperature: 60 °C / 140 °F
- IP protection: 20 according to EN 60529
- Nominal power: 0.25 kW; nominal current: 2.1 A
- Nominal voltage range: 120-240 V, 50 Hz and 60 Hz, +/-10% at indicated power frequency
- Current type: 1 N ~ (single-phase)
- No temperature safety device or class per DIN 12880:2007 (indicated as ---)
- ETL Listed to UL 61010-1:2012 Ed.3+R:06Jun2023, UL 61010-2-012:2022 Ed.2+R:02Nov2023, CSA C22.2 No. 61010-2-012:2019 Ed.2
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal voltage range 120-240 V, 50/60 Hz, single-phase 1 N ~, with +/-10% tolerance at the indicated power frequency. p.7 ↗ | The accessory can be connected to 120 V or 230 V supplies without a voltage selector; verify supply is within +/-10% of nominal. | Art. no. 8012-2439 |
| Nominal power 0.25 kW and nominal current 2.1 A. p.7 ↗ | Size the branch circuit and protective device accordingly; the accessory draws up to 2.1 A. | Art. no. 8012-2439 |
| Nominal temperature 60 °C / 140 °F; maximum permissible temperature with light bars is 60 °C. p.7 ↗ | Do not operate the chamber with LED bars above 60 °C; the 60 °C sticker must be attached to the plant light box front and the chamber door. | Whole series |
| IP protection 20 according to EN 60529. p.8 ↗ | The accessory is not protected against water; it must not become wet during operation, cleaning or maintenance and must be installed splash-proof. | Whole series |
| At full light output (100%) the LED light bars contribute approximately 220 W of heat into the chamber. p.16 ↗ | This heat must be compensated in air-controlled cabinets by an empirically determined offset of temperature and humidity setpoints or by an object temperature control sensor. | Whole series |
| Diaphragm pump provides ventilation with 8 liters/min of ambient air via hose connection. p.16 ↗ | The fresh-air supply prevents CO2 concentration drop due to plant photosynthesis in gas-tight E7 chambers; ensure the ventilation hose is connected. | Whole series |
| 16 LED light bars with clips, positionable on one, two or four levels; 4 connectors each for 4 LED bars. p.16 ↗ | Plan rack layout and cable routing for up to 16 bars; no racks are included and must be ordered separately if needed. | Whole series |
| Permissible ambient temperature range during operation: +18 °C / 64.4 °F to +32 °C / 89.6 °F. p.20 ↗ | Install the plant light box in a room within this temperature range; indoor use only. | Whole series |
| Permissible ambient temperature range during transport and storage: -10 °C / 14 °F to +60 °C / 140 °F. p.19 ↗ | After storage in a cold location, wait at least one hour at the installation site before start-up to avoid condensation. | Whole series |
| LED light bars should be replaced after approx. 25,000 operating hours and at the latest after 6 years. p.42 ↗ | Plan replacement intervals and contact BINDER Service for replacement. | Whole series |
| System time battery: CR1225 button cell; low battery warning below 2.5 V. p.42 ↗ | Replace battery when flashing battery icon appears and re-enter system time and weekday. | Whole series |
| Annual electrical repeat inspection required, plus test before initial startup and after maintenance or repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.41 ↗ | Include the accessory in the chamber's electrical safety testing schedule. | Whole series |
| Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER. p.41 ↗ | Warranty becomes void if maintenance is conducted by non-authorized personnel. | Whole series |
| Mounting rails supplied with 4 x M8 and 2 x M5 screws. p.19 ↗ | Use the supplied fasteners for installing the plant light box on the chamber. | Whole series |
| Do not install the cooling incubator/climate chamber with accessory in unventilated recesses; observe prescribed minimum distances. p.9 ↗ | Ensure sufficient ventilation for heat dispersal to avoid overheating damage. | Whole series |
| The accessory can be used simultaneously with the BINDER CO2 control module accessory. p.15 ↗ | Combined CO2 and light control is permitted; observe both operating manuals. | Whole series |
| Non-ionizing radiation: safe distance of 30 cm from field source to active implants (pacemakers, defibrillators) excludes influence with high probability. p.14 ↗ | Personnel with active implants should maintain at least 30 cm distance from electrical components. | Whole series |
| Do not plug or unplug the light connectors during operation (safety label on top of plant light box). p.7 ↗ | Switch off the plant light box before connecting or disconnecting LED bar connectors. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Flashing battery icon in the upper area of the normal display p.42 ↗ | System time battery low (below 2.5 V) | Replace the battery (CR1225 button cell) and re-enter the system time and weekday after replacing it | Chapter 8.2.2 |