BINDER KB 470
BINDER ·Made in Germany · In Active Production

BINDER KB 470

BINDER KB 470 — Incubators with Temperature 0…70 °C.

The BINDER KB 470 is a 472 L cooling incubator (0…70 °C) in the KB series, distinguished by its 530 W heat compensation and 200 kg load capacity. It provides 0.1 K fluctuation and 0.2 K uniformity for precise temperature control. Its 1.3 kW single-phase design suits standard laboratory installations.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB 470 Current 0–7090195.147265640
KB 1060 0–70117.5195.11070101140
KB 130 0–7071100.512765630
KB 260 0–7090139.626265640
KB 65 0–707190.56340630
KB 720 0–7090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 472 L
Load per rack 40 kg
Max. heat compensation 530 W
Nominal power 1.3 kW
Permitted load 200 kg (KB470-230V / KB470UL-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm (KB470-230V / KB470UL-120V)
Side wall clearance 180 mm (KB470-230V / KB470UL-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 900 mm (KB470-230V / KB470UL-120V)
Height 1951 mm (KB470-230V / KB470UL-120V)
Interior width 656 mm (KB470-230V / KB470UL-120V)
Width 930 mm (KB470-230V / KB470UL-120V)
Net weight of the unit (empty) mfr 212 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 220...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and Series Positioning

Thermal Performance

Installation and Operation

Integration & Operating Notes

Process Limits
Do not exceed 200 kg total load or 40 kg per rack

Exceeding these limits may damage the rack system or compromise temperature uniformity. Distribute samples evenly and verify weight before loading.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose the KB 470 over smaller KB series models like the KB 260?

Choose the KB 470 when you need 472 L capacity and higher load handling (200 kg total, 40 kg per rack) or greater heat compensation (530 W). Smaller models suit lower volume or lighter loads but share the same temperature range.

What maintenance and accessories are needed beyond the base unit?

Annual electrical safety inspection is mandatory. LED plant light module and CO2 control module are optional accessories. Racks are not included and must be ordered separately if needed.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB 470 en
Datasheet ·495,3 КБ ·7 pp.
View PDF
PDF
BINDER LED Pflanzenlichtmodul (E7) 09 2025 en
Manual ·5,5 МБ ·55 pp.
View PDF