Induztr
Engineering Digest Document Viewer Related Models

Solid Phase Microextraction (SPME) Fibers and Arrows

Contents

  1. Overview pp. 1-2
    Introduces SPME fibers and SPME Arrows product line
  2. Config pp. 3-4
    Guidance and general recommendations for selecting SPME phase
  3. Procedure pp. 5-6
    Sample recovery and direct immersion (DI) procedures
  4. Overview p. 7
    Overview table of the Agilent SPME product portfolio
  5. Ordering p. 8
    Tables of Agilent SPME fiber and Arrow variety kits
  6. p. 9
    Reference list for the document
  7. p. 8 table, 8 rows
  8. p. 7 table, 35 rows

Table 2. Agilent SPME fiber and Arrow variety kits.

Part NumberFiber/ArrowSmartProduct DescriptionFiber Gauge/ Arrow odPhase LengthPhase ThicknessColorPack
5191-5879FiberNoSPME fiber selection set 2 – PDMS (7 µm, 30 µm, and 100 µm), DVB/CWR/PDMS (80 µm), DVB/PDMS (65 µm)2310 mmMultipleMultiple5 pack
5610-5879FiberYesSmart SPME fiber selection set 2 – PDMS (7 µm, 30 µm, and 100 µm); polyacrylate (85 µm); CWR/PDMS (95 µm)2310 mmMultipleMultiple5 pack
5191-5878FiberNoSPME fiber selection set 1 – PDMS (100 µm), polyacrylate (85 µm), CWR/PDMS (95 µm), DVB/PDMS (65 µm), DVB/CWR/PDMS (80 µm)2310 mmMultipleMultiple5 pack
5610-5878FiberYesSmart SPME fiber selection set 1 – PDMS (100 µm), polyacrylate (85 µm), CWR/PDMS (95 µm), DVB/PDMS (65 µm), DVB/CWR/PDMS (80 µm)2310 mmMultipleMultiple5 pack
5191-5869ArrowNoSPME Arrow selection set 2 – PDMS (1.10 mm, 100 μm); polyacrylate (1.10 mm, 100 μm); CWR/PDMS (1.10 mm, 120 μm); DVB/PDMS (1.10 mm, 120 μm); DVB/CWR/PDMS (1.10 mm, 120 μm)1.1 mm20 mmMultipleMultiple5 pack
5610-5869ArrowYesSmart SPME Arrow selection set 2 – Smart SPME Arrow, PDMS (1.10 mm, 100 μm); Smart SPME Arrow polyacrylate (1.10 mm, 100 μm); Smart SPME Arrow, CWR/PDMS (1.10 mm, 120 μm); Smart SPME Arrow, DVB/PDMS (1.10 mm, 120 μm); Smart SPME Arrow, DVB/CWR/PDMS (1.10 mm, 120 μm)1.10 mm20 mmMultipleMultiple5 pack
5191-5868ArrowNoSPME Arrow selection set 1 – PDMS (1.10 mm, 100 μm); polyacrylate (1.10 mm, 100 μm); CWR/PDMS (1.10 mm, 120 μm); DVB/PDMS (1.10 mm, 120 μm); PDMS (1.50 mm, 250 μm)Multiple20 mmMultipleMultiple5 pack
5610-5868ArrowYesSmart SPME Arrow selection set 1 – Smart SPME Arrow, PDMS (1.10 mm, 100 μm); Smart SPME Arrow polyacrylate (1.10 mm, 100 μm); Smart SPME Arrow, CWR/PDMS (1.10 mm, 120 μm); Smart SPME Arrow, DVB/PDMS (1.10 mm, 120 μm); Smart SPME Arrow, PDMS, (1.50 mm, 250 µm)Multiple20 mmMultipleMultiple5 pack

Table 1. SPME fibers and Arrows in the Agilent SPME product portfolio.

Part NumberFiber/ArrowSmartProduct DescriptionFiber Gauge/Arrow odPhase LengthPhase ThicknessColorPack
5191-5877, G7371-67001BothNoManual injection kit for SPME fiber and SPME ArrowNANANANA1 each
5191-5876FiberNoAcrylate (polyacrylate)2310 mm85 µmGray3 pack
5610-5876FiberYesAcrylate (polyacrylate)2310 mm86 µmGray3 pack
5191-5858ArrowNoAcrylate (polyacrylate)1.1 mm20 mm100 µmGray3 pack
5610-5858ArrowYesAcrylate (polyacrylate)1.10 mm20 mm100 µmGray3 pack
5191-5875FiberNoCarbon WR/PDMS (carbon wide range/PDMS)2310 mm95 µmDark blue3 pack
5610-5875FiberYesCarbon WR/PDMS (carbon wide range/PDMS)2310 mm95 µmDark blue3 pack
5191-5859ArrowNoCarbon WR/PDMS (carbon wide range/PDMS)1.1 mm20 mm120 µmLight blue3 pack
5191-5863ArrowNoCarbon WR/PDMS (carbon wide range/PDMS)1.5 mm20 mm120 µmLight blue3 pack
5610-5859ArrowYesCarbon WR/PDMS (carbon wide range/PDMS)1.10 mm20 mm120 µmLight blue3 pack
5610-5863ArrowYesCarbon WR/PDMS (carbon wide range/PDMS)1.50 mm20 mm120 µmLight blue3 pack
5191-5874FiberNoDVB/carbon WR/PDMS2310 mm80 µmDark gray3 pack
5610-5874FiberYesDVB/carbon WR/PDMS2310 mm80 µmDark gray3 pack
5191-5861ArrowNoDVB/carbon WR/PDMS1.1 mm20 mm120 µmDark gray3 pack
5191-5864ArrowNoDVB/carbon WR/PDMS1.5 mm20 mm120 µmDark gray3 pack
5610-5861ArrowYesDVB/carbon WR/PDMS1.10 mm20 mm120 µmDark gray3 pack
5610-5864ArrowYesDVB/carbon WR/PDMS1.50 mm20 mm120 µmDark gray3 pack
5191-5870FiberNoPDMS (polydimethylsiloxane)2310 mm7 µmGreen3 pack
5610-5870FiberYesPDMS (polydimethylsiloxane)2310 mm7 µmGreen3 pack
5191-5871FiberNoPDMS (polydimethylsiloxane)2310 mm30 µmYellow3 pack
5610-5871FiberYesPDMS (polydimethylsiloxane)2310 mm30 µmYellow3 pack
5191-5872FiberNoPDMS (polydimethylsiloxane)2310 mm100 µmRed3 pack
5610-5872FiberYesPDMS (polydimethylsiloxane)2310 mm100 µmRed3 pack
5191-5862ArrowNoPDMS (polydimethylsiloxane)1.1 mm20 mm100 µmRed3 pack
5191-5866ArrowNoPDMS (polydimethylsiloxane)1.5 mm20 mm100 µmRed3 pack
5610-5862ArrowYesPDMS (polydimethylsiloxane)1.10 mm20 mm100 µmRed3 pack
5610-5866ArrowYesPDMS (polydimethylsiloxane)1.50 mm20 mm100 µmRed3 pack
5191-5867ArrowNoPDMS (polydimethylsiloxane)1.5 mm20 mm250 µmBlack3 pack
5610-5867ArrowYesPDMS (polydimethylsiloxane)1.50 mm20 mm250 µmBlack3 pack
5191-5873FiberNoPDMS/DVB (PDMS/divinylbenzene)2310 mm65 µmViolet3 pack
5610-5873FiberYesPDMS/DVB (PDMS/divinylbenzene)2310 mm65 µmViolet3 pack
5191-5860ArrowNoPDMS/DVB (PDMS/divinylbenzene)1.1 mm20 mm120 µmViolet3 pack
5191-5865ArrowNoPDMS/DVB (PDMS/divinylbenzene)1.5 mm20 mm120 µmViolet3 pack
5610-5860ArrowYesPDMS/DVB (PDMS/divinylbenzene)1.10 mm20 mm120 µmViolet3 pack
5610-5865ArrowYesPDMS/DVB (PDMS/divinylbenzene)1.50 mm20 mm120 µmViolet3 pack

Engineering Facts

14 facts · every claim linked to a page
Selection

Choose the SPME phase based on analyte molecular weight, polarity, concentration, and matrix complexity.

These four criteria drive the phase selection decision before any hardware is ordered.

Source
“These criteria include molecular weight of the target analytes, polarity of the target analytes, concentration, and the complexity of the matrix.”
Selection

For nonpolar and volatile compounds, select the 100 µm PDMS phase.

Direct phase recommendation for the most common volatile/nonpolar applications.

Source
“Nonpolar and volatile compounds usually require the 100 µm PDMS phase.”
Selection

For polar analytes in polar samples, use the 85 µm polyacrylate phase.

Prevents poor recovery when a nonpolar phase is mistakenly chosen for polar analytes.

Source
“To extract polar analytes from polar samples, use an 85 µm PA coated phase.”
Selection

For trace-level volatile analysis, the 75 µm PDMS/CWR phase is recommended.

Guides phase choice when low detection limits are the primary requirement.

Source
“For trace-level volatile analysis, the 75 µm PDMS/CWR phase is recommended.”
Selection

The PDMS coating is the most robust choice for direct analysis of complex matrices.

Robustness may outweigh sensitivity when samples are dirty or complex.

Source
“The PDMS coating is the most robust option for directly analyzing complex matrices”
Selection

SPME Arrows offer up to 6× larger sorption phase surface and up to 20× larger volume than fibers.

Determines whether a fiber or an Arrow is needed for sensitivity and phase volume.

Source
“SPME Arrows feature larger sorption phase surfaces (up to 6×) and volumes (up to 20×).”
Installation

SPME fibers use a standard 10 mm length and 23-gauge needle, while Arrows use 20 mm length with 1.1 or 1.5 mm od.

Needle gauge/od and length must match the injector, septum, and microseal hardware.

Source
“All Agilent SPME Arrows have a standard length of 20 mm with either a 1.1 or 1.5 mm outside diameter (od) needle”
Installation

Replacement microseals are size-specific for 1.1 mm and 1.5 mm Arrow probes.

Wrong microseal size causes leaks or damage; must be ordered to match the Arrow od.

Source
“Replacement microseals for use with 1.1 mm arrow SPME probes”
Operation

Time control is critical for quantitation, so automation is highly recommended.

Manual timing variability degrades reproducibility; drives the decision to automate.

Source
“Time control is critical for quantitation, thus automation is highly recommended”
Operation

Increasing sample temperature raises sensitivity for high-boiling components but lowers it for low-boiling components.

Temperature must be tuned per analyte volatility rather than simply maximized.

Source
“increasing the sample temperature will increase the sensitivity for the higher boiling components, but decrease the sensitivity for the lower boiling components”
Operation

Salting out with saturated NaCl improves extraction for many analytes but does not help in every application.

Prevents assuming salt addition is universally beneficial during method development.

Source
“salting out does not help in every application”
Operation

DI-SPME greatly improves extraction efficiency for polar and semivolatile compounds that headspace sampling handles poorly.

Drives the choice between headspace and direct immersion modes.

Source
“The extraction efficiency for polar and semivolatile compounds improves drastically when DI-SPME is performed.”
Maintenance

Solid porous phases degrade from prolonged direct exposure to complex matrices and can saturate.

Explains lifetime limits and the need for rinses and proper conditioning in DI mode.

Source
“solid porous phases show certain drawbacks associated with deterioration of the phase surface due to direct exposure to complex matrices for extended periods of time”
Maintenance

SPME phases may swell in certain solvents, and PDMS swells in nonpolar solvents such as toluene and dichloromethane.

Solvent compatibility must be checked to avoid phase damage and artifacts.

Source
“PDMS swells in contact with nonpolar solvents (such as hydrocarbons, toluene, and dichloromethane).”