Agilent Technologies 5977C GC/MSD
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies 5977C GC/MSD

Agilent Technologies 5977C GC/MSD — Gas chromatography mass spectrometry gc ms.

The 5977C GC/MSD is a single-quadrupole mass-selective detector configured here as an ethylene and trace-gas analysis system with a HydroInert ion source and hydrogen carrier gas. It quantifies permanent gases and greenhouse gases at ppb detection limits (SF6 0.63 ppb, CH4 1.2 ppb) and aromatics per ASTM D5769, distinguishing it from triple-quadrupole and Q-TOF siblings by its scan-mode single-quadrupole architecture.

Agilent Technologies GC MS Instruments Series — Model Comparison & Selection

Model Pumping speed, m³/hCarrier gasTransfer line temperature, °CSample volume, milliliterIon source temperature, °CAnalysis time, min
5977C GC/MSD Current 1Hydrogen2303×10⁻⁴2307.5
7000E GC/TQ 6×10⁻⁵Helium3000.00228010
7010D GCMS —Helium3000.002280—
Quadrupole Time-of-Flight GC/MS, Mass Screening GC/Q-TOF —1.5280———

Full specifications

Core Performance & Functional Specifications
Pumping speed 1 flow
Analysis time 7.5 min
Carrier gas Hydrogen
Column Film Thickness 0.4 µm
Column Flow 0.9 mL/min
Column temperature 250 °C
Foreline Reading 34.7 mTorr
Gas Flow 2 mL/min
Ion source Agilent HydroInert source
Ion Vacuum Gauge Reading 3.18 Torr
Maximum Recommended Gas Flow 1.5 mL/min
Optimal He Column Flow 1 mL/min
Optional Gauge Reading 2.18 Torr
Resolution 4.77 milligram
Sample volume 0.3 µL
Scan speed 2
Septum Purge 3 mL/min
Septum Purge Flow 2 mL/min
Total Run Time nine minutes
Acquisition mode mfr Scan Scan
Gain factor mfr 0.5
Chromatographic separation time mfr 7.5 minutes
Detection limit (MDL), CO2 mfr 2.5 ppb
Detection limit (MDL), N2O mfr 2.0 ppb
Detection limit (MDL), C2H4 mfr 2.1 ppb
Detection limit (MDL), SF6 mfr 0.63 ppb
Detection limit (MDL), CH4 mfr 1.2 ppb
Detection limit (MDL), CO mfr 4.1 ppb
Linearity R2, CO2 mfr 0.99934554
Linearity R2, N2O mfr 0.99980122
Linearity R2, C2H4 mfr 0.99998106
Linearity R2, SF6 mfr 0.99976908
Linearity R2, CH4 mfr 0.99981775
Linearity R2, CO mfr 0.99988739
Inlet mode mfr Multimode inlet , constant flow flow
Inlet liner mfr 4 mm
Equilibrium time mfr 1 min
Ionization mode mfr 70 eV
Number of aromatic species quantified mfr 24 species
Linearity (R2) for toluene mfr 0.99996
Extraction lens diameter mfr 9 mm
Inlet septum mfr Advanced Green, non-stick stick
Benzene average content mfr 1.99 % mass
Benzene %RSD mfr 0.37
Toluene average content mfr 15.17 % mass
Toluene %RSD mfr 1.04
Ethylbenzene average content mfr 3.54 % mass
Ethylbenzene %RSD mfr 0.49
p-Xylene/m-Xylene average content mfr 8.51 % mass
p-Xylene/m-Xylene %RSD mfr 0.42
o-Xylene average content mfr 4.37 % mass
o-Xylene %RSD mfr 0.48
Total aromatics (vol %) mfr 14.46
Total aromatics (vol %) reference mfr 14.82
Benzene (vol %) mfr 0.47
Benzene (vol %) reference mfr 0.50
Toluene (vol %) mfr 2.74
Max column id mfr 0.25 mm
CI capability mfr No No
Inert ion sources available mfr Yes Yes
GC compatibility mfr 9000
Foreline pumps available mfr Pfeiffer Duo Duo
DIP capability (3rd party) mfr Yes Yes
Peak area RSD, CO2 mfr 2.2
Peak area RSD, N2O mfr 2.5
Peak area RSD, C2H4 mfr 3.3
Peak area RSD, SF6 mfr 0.7
Peak area RSD, CH4 mfr 1.5
Peak area RSD, CO mfr 4.9
Linearity (R2), toluene mfr 0.99996
Linearity (R2), benzene mfr 0.99998
Linearity (R2), o-xylene mfr 1.00000
Total aromatics, measured mfr 14.82
Total aromatics, reference mfr 14.46
Benzene, measured mfr 0.50
Benzene, reference mfr 0.47
Toluene, measured mfr 2.74
Benzene, average mfr 1.00 % m/m
Toluene, average mfr 9.00 % m/m
Ethylbenzene, average mfr 3.00 % m/m
p-Xylene/m-Xylene, average mfr 2.98 % m/m
o-Xylene, average mfr 2.98 % m/m
1,2,4-Trimethylbenzene, average mfr 2.95 % m/m
1,2,4,5-Tetramethylbenzene, average mfr 3.00 % m/m
Benzene, %RSD mfr 0.37
Toluene, %RSD mfr 1.04
High vacuum pump mfr Diffusion Diffusion
Instrument mfr 8890 GC
Column Oven Ramp Program mfr 40 min
Carrier gas purity mfr 99.9995
Ionization energy mfr 70 eV
Group 1 mfr 4.25
Group 2 mfr 5.60
Group 3 mfr 8.00
Group 4 mfr 8.80
Group 5 mfr 10.45
Group 6 mfr 12.00
Group 7 mfr 13.70
PCB 118 detection limit mfr 10 ppb
Sampler mfr 7650 Sampler
Hydrogen Leak Sensor mfr 2 GC
Extractor Lens mfr 9 mm
Acquisition Mode mfr Scan Scan
Inlet Septum mfr Advanced Green, non-stick stick
Inlet Liner mfr Deactivated, low pressure drop split liner w/ gl
ALS Syringe mfr 5 cone
Carrier Gas Filter mfr 1 in
Runtime mfr 9 min
Mode mfr 700
Gas Saver mfr 20 minutes
Initial Hold mfr 1 min
Ramp Rate mfr 25 °C/min
Final Hold mfr 1 min
Solvent Delay mfr 2 min
Scan Range mfr 40...200 amu
Tuning Algorithm mfr atune atune
Gain Factor mfr 0.3
Trace Ion Detection (TID) mfr Enabled Enabled
Extraction Lens Diameter mfr 9 mm
Toluene Calibration Concentration mfr 16.3
1,4-Diethylbenzene Concentration mfr 100 ppm
Method Resolution Requirement mfr 2
Total Runtime mfr nine minutes minutes
CI capability‡ mfr No No
DIP** capability (3rd party) mfr Yes Yes
Methane/PCI mfr 17
Methane/NCI mfr 17
Isobutane/PCI mfr 39
Isobutane/NCI mfr 17
Ammonia/PCI mfr 18
Ammonia/NCI mfr 17
Not shown mfr HES assembly assembly
Intuvo column conditioning ramp rate mfr 10...15 °C/min
Intuvo column conditioning hold time mfr 30 minutes
NCI methane flow mfr 40
NCI OFN injection volume mfr 2 μL
NCI OFN concentration mfr 100 fg/μL
Dimensions, Weight & Mechanical Design
Column Inner Diameter 0.18 mm
Column Length 40 m
Foreline Pump Weight 11 kg
Weight 39 kg
Housing height mfr 41 cm
Housing width mfr 30 cm
Housing depth mfr 54 cm
Total system footprint mfr 2 systems
GC footprint mfr 8890 System
MSD housing height mfr 41 cm
MSD housing width mfr 30 cm
MSD housing depth mfr 54 cm
Diffusion pump mainframe weight mfr 39 kg
EI/CI turbo pump mainframe weight mfr 46 kg
Electrical, Power & Drive Specifications
Primary fuse rating mfr 12.5 V
Ionization Voltage mfr 70 eV
Operating Conditions, Environment & Protection
Inlet Temperature 300 °C
Ion Source Temperature 230 °C
Quadrupole Temperature 150 °C
Quad Temperature 150 °C
Sound pressure 70 dB
Sound Pressure Lp 70 dB
Source Temperature 250 °C
Transfer Line Temperature 230 °C
Valve Oven Temperature 50 °C
PCB 118 signal-to-noise ratio mfr 3
Standards, Safety & General Data
Standard EI turbo pump mainframe weight mfr 44 kg

Description

Construction and positioning in the series

The 5977C GC/MSD is the single-quadrupole member of Agilent's GC/MSD line. The sibling models listed — 7000E and 7010D triple quadrupoles and the 7250 GC/Q-TOF — are tandem and high-resolution instruments; the 5977C is the scan-mode single-quadrupole platform used where unit-mass resolution with full-scan acquisition is sufficient. This configuration pairs the MSD with a multimode inlet operated in constant-flow mode, a 4 mm inlet liner, a 20 µm column, and the HydroInert ion source, which is specified for hydrogen carrier gas service.

Behavior in the configured method

Application coverage from the attached documents

The document set covers three distinct uses of the same MSD platform: trace permanent and greenhouse gases with the high-efficiency source (10 species including H2, O2, N2, Ar), PCB determination by US EPA Method 1628 with the HydroInert source and a DB-XLB column on an 8890 GC, and aromatic quantification in gasoline per ASTM D5769 on an 8850 GC with the optional 9 mm extraction lens. In the aromatics method, 24 species are quantified; benzene averages 1.99 % mass at 0.37 %RSD and toluene 15.17 % mass at 1.04 %RSD, with toluene linearity R² 0.99996. These are method results, not instrument ratings, and depend on the column, inlet hardware and carrier gas named in each document.

Control and operation

The digests describe method parameters and results only; they contain no installation, connection or control-section content, so wiring, communication interfaces and startup sequencing cannot be stated from the verified facts.

Integration & Operating Notes

Process Limits
HydroInert source is tied to hydrogen carrier service

The configured ion source is the Agilent HydroInert source, and the method facts specify hydrogen as carrier gas. Method performance figures (MDLs, linearity) were obtained under that carrier/source pairing; substituting helium or another carrier without re-validating the method invalidates the quoted detection limits.

Factory Specification Sheet; Application Note 5994-5743EN
Process Limits
Thermal setpoints are method-specific, not universal maxima

Inlet and transfer line are held at 300 °C, ion source at 230 °C and quadrupole at 150 °C in the configured method. These are operating setpoints for the described analysis; the verified facts give no independent maximum ratings, so do not treat them as absolute limits when developing other methods.

Factory Specification Sheet
Maintenance
Consumables named in the method are not part of base delivery

The method specifies a 4 mm inlet liner, a 20 µm column and an Advanced Green non-stick septum, and the aromatics method uses an optional 9 mm extraction lens. These are method-specific consumables and hardware; confirm each against the ordered configuration, since the verified facts do not establish them as included accessories.

Factory Specification Sheet; Application Note 5994-8386EN
Process Limits
Quoted MDLs and R² belong to the named method only

Detection limits from 0.63 ppb (SF6) to 4.1 ppb (CO) and linearity R² values were generated with the specified column flow, oven program, 70 eV ionization and scan/gain settings. Any change to carrier, column dimensions, inlet mode or acquisition parameters requires re-establishing these figures.

Factory Specification Sheet; Application Note 5994-5743EN

Selection & Integration Q&A

When should I choose the 5977C GC/MSD instead of the 7000E/7010D triple quadrupoles or the 7250 GC/Q-TOF?

Pick the 5977C when unit-mass full-scan data at ppb levels suffice, as in the trace-gas method (SF6 0.63 ppb, CH4 1.2 ppb) and ASTM D5769 aromatics. The triple quadrupoles and Q-TOF siblings target MRM trace quantitation and high-resolution exact-mass work, which this scan-mode configuration does not provide.

What adjacent hardware and consumables must match the configured method?

The method pairs the MSD with a multimode inlet in constant-flow mode at 0.9 mL/min hydrogen, a 4 mm inlet liner, a 20 µm column, an Advanced Green non-stick septum and, for ASTM D5769 aromatics, the optional 9 mm extraction lens. The HydroInert source is specified for hydrogen carrier; confirm column and liner part numbers against the ordered configuration.

Which consumables and accessories are not included in base delivery and must be ordered separately?

The 4 mm inlet liner, 20 µm analytical column, Advanced Green non-stick septum and the optional 9 mm extraction lens are method-specific items named in the application documents rather than confirmed base-delivery contents. The verified facts contain no maintenance-interval or spare-parts schedule, so service intervals must be taken from the instrument manual.

Dimensional Drawings & Technical Documentation

Dimensional drawings

5977C GC/MSD — drawing 1
performance curve

Click drawing to inspect vector dimensions

Technical documentation

PDF
Analysis of Trace Permanent Gases
Brochure ·1,1 МБ ·10 pp.
View PDF
PDF
Improved Determination of PCBs by EPA Method 1628
Brochure ·630,6 КБ ·5 pp.
View PDF
PDF
Analysis of Total Aromatic Content
Brochure ·810,2 КБ ·8 pp.
View PDF
PDF
5977C Series MSD Operating Manual
Manual ·234 pp.
View PDF