Agilent Technologies purge and trap, volatile organic compounds analysis
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies purge and trap, volatile organic compounds analysis

Agilent Technologies purge and trap, volatile organic compounds analysis — Gas chromatography.

Purge-and-trap concentrator configuration for VOC water analysis, defined by a timed sequence of purge, dry-purge, desorb and bake phases. It is distinguished within the series by a 20 °C purge at 40 mL/min, 250 °C desorb and 280 °C bake, with a 0.5–50 ppb calibration range. It serves as the sample-introduction front end to a GC or GC/MS for EPA 5030/8260, 524.2 and 624 methods.

Agilent Technologies GC Sample Preparation Introduction Series — Model Comparison & Selection

Model Sample volume, milliliterAnalysis time, minBake flow, mL/minBake temperature, °CBake time, minCalibration concentration, µg/ml
purge and trap, volatile organic compounds analysis Current ——2002802—
GC autosampler, GC sample injection 0.00180———100
PAL3 autosampler, automated solid phase extraction ——————
Reliable, consistent GC introduction of liquefied gases ——————

Full specifications

Core Performance & Functional Specifications
Bake Flow 200 mL/min
Bake Time 2.00 min
Drain Flow 300 mL/min
Dry Purge Flow 100 mL/min
Presweep Time 0.25 min
Purge Flow 40 mL/min
Purge Time 11.00 min
Rinse Loop Time 0.30 min
Sample Loop Time 0.35 min
Sample Transfer Time 0.35 min
Standby Flow 20 mL/min
Sweep Needle Time 0.30 min
Desorb time mfr 1.00 min
Bake rinse drain cycles mfr 1
Bake rinse drain time mfr 0.35 min
Calibration range mfr 0.5...50 ppb
Dry purge time mfr 1.00 min
Operating Conditions, Environment & Protection
Bake Temperature 280 °C
Desorb Preheat Temperature 245 °C
Dry Purge Temperature 20 °C
Mcs Bake Temperature 200 °C
Mcs Purge Temperature 20 °C
Purge Ready Temperature 35 °C
Purge Temperature 20 °C
Sample Mount Temperature 90 °C
Transfer Line Temperature 150 °C
Valve Oven Temperature 150 °C
Water Temperature 90 °C

Description

Construction and positioning

Behavior across the cycle

Timing and control

Integration & Operating Notes

Process Limits
Dry-purge step is disabled in this configuration

Dry-purge flow is set to 0 mL/min at 20 °C with a 0 min dry-purge time, so no trap drying occurs between purge and desorb. For water samples with high moisture load, confirm with the method whether trap drying is needed; otherwise water carryover reaches the desorb step and the GC column.

Factory Specification Sheet
Process Limits
Bake at 280 °C exceeds desorb temperature by 30 °C

The bake phase runs at 280 °C while desorption is at 250 °C with a 245 °C preheat. Trap and transfer hardware must tolerate the 280 °C bake; the transfer line and valve oven are held at 150 °C, so the temperature gradient along the path must be checked to avoid cold spots condensing heavy VOCs.

Factory Specification Sheet
Control & I/O
Purge-ready at 35 °C gates the next cycle start

The sequence uses a 35 °C purge-ready temperature rather than a full cooldown, so cycle timing depends on the controller honoring this setpoint before re-opening the sample loop. The fixed valve timings (presweep 0.25 min, rinse loop 0.30 min, sample loop 0.35 min, transfer 0.35 min, sweep needle 0.30 min) must be sequenced by the GC/autosampler controller.

Factory Specification Sheet
Maintenance
Bake rinse drain cycle is a single 0.35 min pass

The bake rinse drain is configured as 1 cycle of 0.35 min at a 300 mL/min drain flow. If the drain line or condenser fouls, residual water remains in the trap path; verify drain flow and condenser bake (180 °C) at each preventive-maintenance interval rather than adding cycles ad hoc.

Factory Specification Sheet
Process Limits
Calibration range is 0.5–50 ppb for water VOCs

The stated calibration range of 0.5–50 ppb applies to the purge-and-trap water method. Samples above 50 ppb require dilution or a reduced purge volume; the application notes for EPA 8260, 524.2 and 624 document the calibration and MDL data on this concentrator family with an Agilent 7890B GC/MS.

Application notes 5991-9254EN, 5991-9255EN, 5991-9256EN

Selection & Integration Q&A

When should I choose this purge-and-trap concentrator instead of the 7697A headspace or 7693A liquid sampler in the same series?

Pick this unit for volatile organics in water requiring EPA 5030/8260, 524.2 or 624 purge-and-trap concentration at 0.5–50 ppb. The 7697A headspace sampler suits volatile organics in solid or viscous matrices without a purge step, and the 7693A liquid sampler is for direct liquid injection, not trace water VOCs.

What GC/MS and autosampler configuration is required to run this concentrator for EPA methods?

The documented configuration couples the concentrator to an Agilent 7890B GC with a 5977A MS and an AQUATek 100 autosampler. The transfer line and valve oven are held at 150 °C, so the GC inlet and column must accept a 150 °C transfer path; confirm the MS source and column choice against the method tables in the application notes.

What maintenance and consumables are needed beyond the base delivery?

Plan for trap replacement and condenser cleaning: the condenser bakes at 180 °C and the bake rinse drain is a single 0.35 min cycle at 300 mL/min, so drain-line fouling shows up as residual water. The base unit does not include the autosampler, GC/MS, columns or calibration standards; these are separate items per the application notes.

Dimensional Drawings & Technical Documentation

Technical documentation