BINDER BD-S 56
BINDER BD-S 56 — Incubators with Temperature 5…70 °C.
The BD-S 56 is a 62 L natural-convection incubator with a stainless steel interior (1.4016) and PID controller, offering 5–70 °C operation with 0.1 °C display resolution. It is the smaller of two BD-S sizes (56 vs 115), distinguished by class 3.1 overtemperature protection and 0.3 K fluctuation. It serves culture incubation at 37 °C and other temperature-controlled laboratory processes.
Direct drop-in replacements
BINDER BD-S Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| BD-S 56 Current | 5–70 | 57.7 | 62.5 | 62 | 400 | 15 |
| BD-S 115 | 5–70 | 61.6 | 73.5 | 118 | 550 | 20 |
Full specifications
| Temperature | 5...70 °C |
| Interior volume | 62 L |
| Load per rack | 15 kg |
| Nominal power | 0.3 kW |
| Permitted load | 45 kg (BDS056-230V) |
| Temperature fluctuation | 0.3 K |
| Temperature uniformity | 0.5 K |
| Back wall clearance | 160 mm (BDS056-230V) |
| Side wall clearance | 100 mm (BDS056-230V) |
| Article Number mfr | 9090 |
| Recovery time after door was opened for 30 s at 37°C mfr | 15 min |
| Depth | 577 mm (BDS056-230V) |
| Height | 625 mm (BDS056-230V) |
| Interior width | 400 mm (BDS056-230V) |
| Width | 614 mm (BDS056-230V) |
| Net weight of the unit (empty) mfr | 35 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The BD-S 56 is a 62 L incubator with natural convection, stainless steel interior (1.4016 / AISI 430) and polyester-coated RAL 9003 housing. It uses the R-S microprocessor controller with 0.1 °C display resolution.
- Within the BD-S series, it is the 56-size model; the sibling BD-S 115 offers 115 L. Both share the 5–70 °C range, but the BD-S 56 has a 0.3 kW nominal power and 6.3 A fuse (230 V variant).
- Unlike ED-S/FD-S ovens (safety class 2), BD-S incubators are equipped with a safety controller class 3.1 per DIN 12880, providing higher overtemperature protection.
Behavior and performance
- Temperature fluctuation is 0.3 K and uniformity 0.5 K, suitable for incubation at 37 °C. Heating-up time to 100 °C is 230 min (from ambient).
- Recovery time after a 30 s door opening at 37 °C is 15 min, reflecting the natural-convection design and limited 0.3 kW heater power.
- The chamber has no explosion protection; charging materials must be non-combustible and operating temperature must remain below their flash point.
Installation and control
- The 230 V variant (BDS056-230V, Art.-No. 9090-0016) requires a 230 V 1~ 50/60 Hz supply with protective earth, CEE 7/7 plug, and 6.3 A fusing. Voltage tolerance is ±10 %.
- Wall clearances: 160 mm at rear, 100 mm at sides. Do not install in unventilated recesses. Lift at four lower corners with two people; never use the door handle.
- The R-S controller supports set-point entry, °C/°F selection, delayed-off timer, and safety controller adjustment. An exhaust air flap allows manual adjustment of air change rate.
Integration & Operating Notes
The 230 V variant requires a single-phase 230 V supply with protective earth, tolerance ±10 %, and a 6.3 A chamber fuse. Verify mains quality and plug type (CEE 7/7) before installation; incorrect voltage or missing earth can damage the controller or create a shock hazard.
The chamber requires unobstructed air circulation for heat dispersal. Installing it in a recess or closer than the specified clearances risks overheating and reduced temperature performance. Ensure the floor or bench can support the 35 kg net weight plus 45 kg load.
The chamber has no explosion protection. Charging materials must be harmless and their flash point or sublimation point must lie above the operating temperature. Corrosive substances (acids, halides) can damage the stainless steel interior and are excluded from liability.
The BD-S is equipped with an overtemperature safety device class 3.1 per DIN 12880. The safety controller set-point must be adjusted to the application to prevent excessive temperatures. Alarms must be acknowledged and the cause rectified before restarting.
Regular maintenance at least once a year and repeat electrical tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. Disconnect the power plug before maintenance; do not remove the rear panel. Replace the door gasket only when cold.
If the chamber was stored or transported below ambient temperature, condensation may form internally. Allow at least one hour at ambient temperature and ensure it is completely dry before connecting power to avoid electrical hazard or damage.
Selection & Integration Q&A
Choose the BD-S 56 for incubation at up to 70 °C with 62 L capacity and 0.1 °C display resolution; the BD-S 115 offers 115 L for larger loads. ED-S/FD-S ovens are for drying/heating up to higher temperatures but have 1 °C resolution and class 2 safety, not class 3.1.
It requires a 230 V 1~ 50/60 Hz supply with protective earth, tolerance ±10 %, a CEE 7/7 plug, and a 6.3 A chamber fuse. Nominal power is 0.3 kW. Verify the type plate for the exact delivered configuration.
Maintain 160 mm clearance at the rear and 100 mm at the sides. Do not install in unventilated recesses. Permissible storage temperature is -10 to +60 °C with max. 70 % r.h. non-condensing; after cold storage, wait one hour before start-up.
Annual maintenance and electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702 are recommended. The door gasket is a wear part (replace only when cold). Racks and accessories are optional; the base delivery includes the chamber only. Disconnect power before service.