BINDER CB 170
BINDER CB 170 — Incubators with Temperature 4…50 °C.
The BINDER CB 170 is a 170 L CO2 incubator with a sterilizable NDIR CO2 sensor and MB2 touchscreen controller, operating from +4 to 50 °C with 0.1 K fluctuation and 0.3 K uniformity. It is the mid-size chamber of the CB series, positioned between the 56 L CB 56 and 260 L CB 260, and is intended for cell culture and controlled atmosphere incubation requiring precise temperature and CO2 regulation.
Direct drop-in replacements
BINDER CB Series — Model Comparison & Selection
Full specifications
| Temperature | 4...50 °C |
| Interior volume | 170 L (CB 170) |
| Load per rack | 10 kg (all variants) |
| Nominal power | 1.4 kW (all variants) |
| Permitted load | 30 kg (all variants) |
| Temperature fluctuation | 0.1 ± K (all variants) |
| Temperature uniformity | 0.3 K |
| Side wall clearance | 50 mm (all variants) |
| Article Number mfr | 9040 |
| Depth | 715 mm (all variants) |
| Height | 870 mm (all variants) |
| Interior width | 560 mm (all variants) |
| Net weight | 95 kg (all variants) |
| Width | 680 mm (all variants) |
| Power frequency | 50/60 Hz (all variants) |
| Rated voltage | 200...230 V (CB170-230V variants) |
| Unit fuse | 10.0 A (200...230 V variants) |
| Humidity range | 90 ...95 % RH (all variants) |
Description
Construction and Series Positioning
The CB 170 uses a 170 L interior (560 mm wide, 715 mm deep, 870 mm high) within a 680 mm wide, 95 kg cabinet. It is the middle size of the CB series, larger than the CB 56 and smaller than the CB 260. Unlike the CBF variants, this model does not include humidity control; the humidity range of 90–95 % RH is achieved via the Permadry water pan system. The standard model is the CB170-230V (article 9040-0199) for 200–230 V supply; left-hinged (-L) and O2-control (-O) variants are available but are separate modifications.
Behavior Under Load and Control
Temperature is controlled from +4 to 50 °C with a fluctuation of ±0.1 K and uniformity of 0.3 K. The chamber supports a permitted load of 30 kg and 10 kg per rack. The MB2 controller manages fixed-value, timer, and week programs with setpoint ramps and steps. CO2 control uses a sterilizable NDIR sensor; O2 control (0.2–20 vol.-% standard, 10–90 vol.-% optional) is only present on -O variants. During hot-air sterilization, CO2/O2 control is deactivated until the chamber cools below 80 °C and sterilization success is confirmed.
Installation and Operation
Side wall clearance of 50 mm is required. Electrical supply is 200–230 V, 50/60 Hz, 1.4 kW nominal, with a 10.0 A unit fuse. Gas connections for CO2 (and O2/N2 if equipped) require appropriate cylinder connection kits. The Permadry water pan must be filled for humidity. The DuoDoor door lock and touchscreen provide access and control. Alarm functions include a zero-voltage relay output and adjustable delay times for temperature, CO2, O2, and door-open events.
Integration & Operating Notes
The CB 170 standard variant requires 200–230 V at 50/60 Hz and is protected by a 10.0 A unit fuse. Connecting to a 100–120 V supply without the UL variant will prevent operation and may damage components. Confirm the local supply matches the model code (CB170-230V vs CB170UL-120V) before installation.
During hot-air sterilization, CO2 and O2 control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas regulation during this period; plan cell culture workflows to avoid relying on CO2 control immediately after a sterilization cycle.
Regular maintenance must be performed at least once a year, including electrical safety inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702. After 8760 operating hours or two years, a maintenance message appears and recurs every two weeks until reset by BINDER Service. Schedule service accordingly to avoid interrupted operation.
The door gasket must be replaced only when the chamber is at room temperature. Attempting replacement while hot can deform or damage the gasket, leading to sealing failures and temperature/humidity deviations. Allow sufficient cooling time before service.
Selection & Integration Q&A
Choose CB 170 for 170 L capacity, balancing the smaller CB 56 and larger CB 260. It offers the same +4 to 50 °C range as CB 56 but more space; CB 260 extends to +5 to 50 °C with greater volume. Select based on sample throughput and available bench space.
A CO2 gas cylinder connection kit is required (available as accessory). For O2-control variants, additional O2 and N2 supplies and kits are needed. Ensure gas inlet pressure is within specified limits and use appropriate pressure regulators.
Base delivery excludes gas cylinder connection kits, Permadry water pan (for humidity), and optional accessories like APT-COM 4 software, silicone access ports, interior socket, and stacking adapters. Annual maintenance and electrical safety tests are user responsibilities; CO2 sensor replacement requires recalibration.