BINDER CB 56
BINDER CB 56 — Incubators with Temperature 4…50 °C.
The BINDER CB 56 is a 53 L CO2 incubator with hot-air sterilization and a heat-sterilizable IR CO2 sensor, available in 200–230 V (CB056-230V) and 100–120 V UL (CB056UL-120V) variants. It operates from +4 to 50 °C with 0.1 K fluctuation and 0.3 K uniformity, and provides 90–95 % RH via the Permadry water pan. It is distinguished within the CB series by its compact 53 L chamber, suited to small-scale cell culture where reproducible CO2 control and in-chamber sterilization are required.
Direct drop-in replacements
BINDER CB Series — Model Comparison & Selection
Full specifications
| Temperature | 4...50 °C |
| CO2 measuring technology | IR |
| Interior volume | 53 L (CB056-230V / CB056UL-120V) |
| Load per rack | 10 kg (CB056-230V / CB056UL-120V) |
| Nominal power | 0.8 kW |
| Permitted load | 30 kg (CB056-230V / CB056UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 100 mm (CB056-230V / CB056UL-120V) |
| Side wall clearance | 50 mm (CB056-230V / CB056UL-120V) |
| Article Number mfr | 9040 |
| Recovery time after door was opened for 30 s at 37°C mfr | 5 min |
| CO2 recovery time after door was opened for 30 s at 5 vol. % CO2 mfr | 5 min |
| Depth | 545 mm (CB056-230V / CB056UL-120V) |
| Height | 674 mm (CB056-230V / CB056UL-120V) |
| Width | 528 mm (CB056-230V / CB056UL-120V) |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
| Humidity range | 90...95 % RH |
Description
Construction and series positioning
- The CB 56 is the smallest chamber in the CB series (53 L), positioned below the CB 170 and CB 260; it shares the same +4 to 50 °C range but differs in chamber volume and net dimensions (528 × 674 × 545 mm).
- It uses a sterilizable NDIR CO2 sensor and an MB2 touchscreen program controller, with hot-air sterilization capability.
- Two voltage variants exist: CB056-230V (200–230 V, CEE 7/7 plug, Art.-No. 9040-0197) and CB056UL-120V (100–120 V, NEMA 5-20 plug, Art.-No. 9040-0198).
Behavior under load and in different modes
- Temperature control achieves 0.1 K fluctuation and 0.3 K uniformity across the 53 L interior; recovery after a 30 s door opening at 37 °C is 5 min, and CO2 recovery after a 30 s door opening at 5 vol.-% CO2 is also 5 min.
- Humidity is maintained at 90–95 % RH by the Permadry water pan system (standard on CB models).
- Hot-air sterilization cycle: effective sterilization phase (heating and holding) completes automatically after 4 hours, followed by a 6-hour cooling phase to 37 °C; total cycle 10 hours. CO2 control is deactivated during sterilization and reactivated only after confirming “Successful sterilization” and when temperature falls below 80 °C.
Installation, connection, and control specifics
- Installation requires 100 mm rear clearance and 50 mm side clearance; the unit is powered by a single phase (200–230 V, 50 Hz, 0.8 kW nominal, 10 A fuse).
- Gas connections for CO2 (and O2/N2 if O2 control option is present) are made via gas cylinder connection kits (accessory). The MB2 controller supports fixed-value, timer, and week programs, with alarm functions including a zero-voltage relay output and adjustable delay times.
- The chamber is equipped with a DuoDoor door lock and an optional divided inner door; shelves are rated for 10 kg each, with a total permitted load of 30 kg.
Integration & Operating Notes
Replacing the door gasket while the chamber is warm can damage the gasket material. Always allow the chamber to cool to ambient temperature before attempting replacement.
The MB2 controller allows setting alarm delay times for temperature, CO2, and door opening. Configure these delays to avoid false alarms during routine door access, but keep them short enough to detect real deviations.
An electrical safety inspection according to EN 50678/VDE 0701 and EN 50699/VDE 0702 must be performed before initial startup and after any maintenance or repair, and is recommended annually. This ensures continued safe operation.
Selection & Integration Q&A
Choose the CB 56 for small-scale cell culture where 53 L capacity suffices and bench space is limited. The CB 170 and CB 260 offer larger volumes (170 L and 260 L) but share the same temperature range and control features. Select the larger models for higher throughput or when working with multiple racks.
A CO2 gas cylinder connection kit is required (available as accessory). For O2 control (if equipped), additional O2 and N2 connection kits are needed. Ensure the gas supply pressure matches the chamber's inlet requirements as specified in the manual.
The sterilization cycle takes 10 hours total: 4 hours effective sterilization (heating and holding) followed by 6 hours cooling to 37 °C. CO2 control is deactivated during this period and reactivates only after successful confirmation and cooling below 80 °C. Plan experiments around this downtime.
The base delivery includes the chamber and standard accessories. Optional accessories such as gas cylinder connection kits, APT-COM 4 software, silicone access ports, and stacking adapters must be ordered separately. Annual maintenance by BINDER Service is recommended, and the CO2 sensor may need periodic replacement and recalibration.