BINDER KBF-S ECO 240
BINDER ·Made in Germany · In Active Production

BINDER KBF-S ECO 240

BINDER KBF-S ECO 240 — Climate chambers with Temperature 0…70 °C.

The KBF-S ECO 240 is a 247-litre constant climate chamber using Peltier thermoelectric cooling and the RD4 microprocessor controller, with a 0–70 °C range and 10–80 % RH humidity control (5–70 °C with humidity active). It is the smallest of four KBF-S ECO sizes (240/400/720/1020) and is intended for long-term and accelerated stability testing under continuous duty.

Temperature
0…70 °C

BINDER KBF-S ECO Series — Model Comparison & Selection

Model Temperature, °CHumidity fluctuation, % rhInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
KBF-S ECO 240 Current 0–700.3247302500.8
KBF-S ECO 1020 0–700.41020454001.2
KBF-S ECO 400 0–700.3400303001.1
KBF-S ECO 720 0–700.3700454001.2

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 247 L (KBF-S ECO 240, 50/60 Hz)
Load per rack 30 kg
Maximum heat compensation 250 W
Nominal power 0.8 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 5 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 11 min
Dimensions, Weight & Mechanical Design
Depth 796 mm
Height 1461 mm
Width 932 mm
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz
Rated voltage 200...230 V (KBFSECO240-230V)
Unit fuse 16.0 A
Operating Conditions, Environment & Protection
Humidity fluctuation 0.3 % RH
Humidity range 10...80 % RH

Description

Construction and positioning in the series

Behaviour under load and in different media

Control, safety and operation

Integration & Operating Notes

Electrical
Match the variant to the site supply before ordering

The 230 V variant accepts 200–230 V / 50 Hz or 208–230 V / 60 Hz at 3.5 A with a CEE 7/7 plug; the UL variant needs 100–120 V at 6.7 A with a NEMA 5-20 plug. Supplying the wrong voltage causes malfunction, and the higher current of the UL version must be reflected in the branch circuit and its 16 A unit fuse coordination.

Data Sheet Model KBF-S ECO 240; Operating Manual, Type plate p.9
Mounting
Keep the installation site at or below 25 °C ambient

Refrigerating performance falls when ambient temperature exceeds 25 °C / 77 °F, so a climate-controlled room is required for full capability. Wall spacers and the specified installation-site clearances from the manual must be observed, and the chamber must not become wet (IP 20).

Operating Manual, Installation and connections; Simple troubleshooting p.72
Process Limits
Stay inside the temperature/humidity window and heat-load budget

Combinations outside the optimum range of the temperature-humidity diagram (Figure 19) and external heat loads beyond the 250 W maximum heat compensation cause low or no refrigerating performance. With humidity active the working range narrows to 5–70 °C; plan sample heat dissipation and set-points accordingly.

Operating Manual, Simple troubleshooting p.72; Factory Specification Sheet
Process Limits
Limit aggressive substances to protect the humidity sensor

The humidity sensor has stated permanent-load thresholds (e.g. ammonia 5500 ppm, acetone 3300 ppm, formaldehyde 2400 ppm) and a ±2 % RH tolerance. Test articles releasing solvents or bases above these concentrations will degrade humidity measurement, so outgassing loads must be assessed before loading the chamber.

Operating Manual, Resistance of the humidity sensor p.17
Control & I/O
Set the class 3.1 safety controller above the working set-point

A safety controller set too low prevents heat-up and triggers alarms; it must be set and function-checked per chapter 10.2. The class 1 protective device trips at nominal temperature +10 °C, so both layers must be configured consistently with the intended test profile before a run starts.

Operating Manual, Temperature safety devices; Simple troubleshooting p.72
Control & I/O
Network and data paths are configured on the RD4, not by jumpers

Ethernet settings (IP address, subnet mask, gateway, DNS, chamber address) and the DL1 recorder storage rate are set through the controller menus; USB import/export uses a front-panel stick. APT-COM 4 can manage up to 100 networked chambers, so address planning belongs in the integration phase.

Operating Manual, Ethernet network settings; Data recorder; USB menu
Maintenance
Plan annual chamber and humidifier service plus electrical retest

Regular maintenance is required at least once a year, and the steam humidifier at least annually with interval depending on water quality and steam output. Repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before first start-up and after maintenance or repair; work by non-authorized personnel voids warranty.

Operating Manual, Maintenance intervals p.70; Electrical testing p.70

Selection & Integration Q&A

When should I choose the KBF-S ECO 240 rather than the 400, 720 or 1020?

Pick the 240 when 247 L and 30 kg per rack cover the sample load; all four share the 0–70 °C range. The larger 400/720/1020 add anti-condensation protection, which the manual does not list for the 240, so humid programs with heavy condensation risk favor a larger frame.

What supply and protective devices does the 230 V KBF-S ECO 240 need at the wall?

The KBFSECO240-230V takes 200–230 V / 50 Hz or 208–230 V / 60 Hz, 1 N~, 3.5 A, 0.80 kW, with a CEE 7/7 plug and a 16.0 A unit fuse. The UL variant instead needs 100–120 V at 6.7 A with a NEMA 5-20 plug; do not mix the two.

What site conditions and duty limits apply when installing the chamber?

Install where ambient stays at or below 25 °C, since higher ambient reduces refrigerating performance. Respect wall spacers and site clearances, keep the unit dry (IP 20), and stay within the temperature-humidity diagram and the 250 W heat-compensation budget for continuous operation.

What maintenance and accessories are needed beyond the base chamber?

Budget annual chamber and steam-humidifier service plus electrical retest per EN 50678/EN 50699. APT-COM 4 software, the BINDER Pure Aqua water-treatment accessory and USB sticks for data export are options/accessories, not part of base delivery; water quality must suit the humidifier.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF S ECO 240 en
Datasheet ·602,8 КБ ·8 pp.
View PDF
PDF
KBF S ECO (E6) 07 2026 en
Manual ·9,8 МБ ·99 pp.
View PDF