BINDER KBF-S ECO 1020
BINDER KBF-S ECO 1020 — Climate chambers with Temperature 0…70 °C.
The KBF-S ECO 1020 is a 1,020 L constant climate chamber using Peltier thermoelectric cooling, with a working range of 0…70 °C and 10…80 % RH (5…70 °C with humidity). As the largest sibling in the KBF-S ECO series, it offers the same 0.1 K temperature uniformity and 0.4 % RH humidity fluctuation but adds anti-condensation protection and a 400 W max heat compensation. It is intended for long-term and accelerated stability testing under continuous duty in networked laboratory environments.
BINDER KBF-S ECO Series — Model Comparison & Selection
| Model | Temperature, °C | Humidity fluctuation, % rh | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| KBF-S ECO 1020 Current | 0–70 | 0.4 | 1020 | 45 | 400 | 1.2 |
| KBF-S ECO 240 | 0–70 | 0.3 | 247 | 30 | 250 | 0.8 |
| KBF-S ECO 400 | 0–70 | 0.3 | 400 | 30 | 300 | 1.1 |
| KBF-S ECO 720 | 0–70 | 0.3 | 700 | 45 | 400 | 1.2 |
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 1020 L (KBFSECO1020-230V) |
| Load per rack | 45 kg |
| Maximum heat compensation | 400 W |
| Nominal power | 1.2 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.1 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 5 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 15 min |
| Depth | 1146 mm |
| Height | 1925 mm |
| Width | 1254 mm |
| Power frequency | 50/60 Hz |
| Rated voltage | 200...230 V (KBFSECO1020-230V) |
| Unit fuse | 16.0 A |
| Humidity fluctuation | 0.4 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The KBF-S ECO 1020 is the highest-capacity model in the KBF-S ECO series, with 1,020 L interior volume; siblings 240, 400, and 720 share the same temperature range but differ in chamber size and power ratings.
- It uses Peltier thermoelectric cooling rather than a compressor, enabling precise control without refrigerant; the RD4 microprocessor controller manages temperature and humidity set-points.
- Unlike smaller siblings, the 1020 includes anti-condensation protection (also on 400 and 720) to manage moisture on surfaces during humid operation.
- The chamber is rated IP20, so it must be installed in a dry, dust-controlled environment and must not be exposed to water.
Behavior under load and in different media
- Temperature uniformity is 0.1 K and fluctuation 0.1 K, with humidity fluctuation of 0.4 % RH; these figures apply across the working range but may degrade with excessive external heat load or when operating outside the optimum temperature/humidity diagram (Figure 19).
- Recovery after a 30 s door opening is 5 min at 25 °C/60 % RH and 15 min at 40 °C/75 % RH, indicating faster stabilization at lower humidity/temperature.
- The humidity sensor has a tolerated error of ±2 % RH and is resistant to certain substances up to specified concentrations (e.g., ammonia 20 ppm TLV, acetone 500 ppm TLV); exceeding these limits can damage the sensor.
- Refrigerating performance is reduced if ambient temperature exceeds 25 °C; the chamber should be sited accordingly.
Installation, connection, and control specifics
- Electrical supply: 200…230 V, 1~, 50/60 Hz, nominal power 1.2 kW, unit fuse 16 A; the 230 V variant uses a CEE 7/7 plug. A UL variant (100…120 V) exists but is a separate model code.
- Water supply for humidification requires suitable water quality; BINDER Pure Aqua Service is available as an accessory, and a freshwater supply and condensate collection pan must be installed.
- The RD4 controller supports Ethernet networking (MAC, IP, subnet, gateway, DNS) and USB data transfer; APT-COM™ 4 software can network up to 100 chambers.
- Safety devices include an over-temperature protective device (class 1) that trips at nominal temperature +10 °C and an overtemperature safety controller class 3.1 (DIN 12880) that must be set and checked per chapter 10.2.
- The controller includes a data recorder (DL1) with adjustable storage rate, self-test function, and password-protected access levels.
Integration & Operating Notes
Above 25 °C ambient, refrigerating performance degrades, potentially preventing the chamber from reaching low temperature set-points. Ensure the installation site has adequate ventilation and does not exceed this limit, especially in continuous operation.
The 230 V variant requires a 16 A branch circuit protection and a CEE 7/7 outlet. Connecting to a different voltage or using a fuse of lower rating can cause malfunction or nuisance tripping. The UL variant (100–120 V) is a separate model and not interchangeable.
Regular maintenance must be performed at least once a year, including steam humidifier maintenance (interval depends on water quality). Electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after repairs. Warranty may be void if maintenance is done by unauthorized personnel.
Selection & Integration Q&A
Choose the 1020 when you need the largest capacity (1,020 L) for high-throughput testing or large batches. The 720 and 400 offer the same temperature range but smaller volumes; the 1020 also includes anti-condensation protection (like 400 and 720) but has higher power (1.2 kW vs. lower in smaller models). Select the 1020 if your sample sizes or batch counts require the extra space and you can accommodate the larger footprint (1,254 mm wide).
The chamber requires a freshwater supply with suitable water quality (conductivity and hardness per manual) and a condensate collection pan. BINDER Pure Aqua Service is available as an accessory for water treatment. Ensure the supply pressure and flow meet the humidifier's demand, and provide a drain for condensate. The steam humidifier needs annual maintenance, with interval depending on water quality.
No, ambient temperatures above 25 °C reduce refrigerating performance, and prolonged operation above this limit may prevent the chamber from maintaining low temperature set-points. The installation site must be kept below 25 °C for full performance. If occasional excursions occur, ensure they are brief and monitor chamber performance.
Annual maintenance is mandatory, including steam humidifier service (interval depends on water quality) and electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional accessories include APT-COM™ 4 software for networking up to 100 chambers, BINDER Pure Aqua Service for water treatment, and additional racks (max load 45 kg per rack). USB sticks for data transfer are not included.