BINDER KBF-S ECO 1020
BINDER ·Made in Germany · In Active Production

BINDER KBF-S ECO 1020

BINDER KBF-S ECO 1020 — Climate chambers with Temperature 0…70 °C.

The KBF-S ECO 1020 is a 1,020 L constant climate chamber using Peltier thermoelectric cooling, with a working range of 0…70 °C and 10…80 % RH (5…70 °C with humidity). As the largest sibling in the KBF-S ECO series, it offers the same 0.1 K temperature uniformity and 0.4 % RH humidity fluctuation but adds anti-condensation protection and a 400 W max heat compensation. It is intended for long-term and accelerated stability testing under continuous duty in networked laboratory environments.

Temperature
0…70 °C

BINDER KBF-S ECO Series — Model Comparison & Selection

Model Temperature, °CHumidity fluctuation, % rhInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
KBF-S ECO 1020 Current 0–700.41020454001.2
KBF-S ECO 240 0–700.3247302500.8
KBF-S ECO 400 0–700.3400303001.1
KBF-S ECO 720 0–700.3700454001.2

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 1020 L (KBFSECO1020-230V)
Load per rack 45 kg
Maximum heat compensation 400 W
Nominal power 1.2 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 5 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 15 min
Dimensions, Weight & Mechanical Design
Depth 1146 mm
Height 1925 mm
Width 1254 mm
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz
Rated voltage 200...230 V (KBFSECO1020-230V)
Unit fuse 16.0 A
Operating Conditions, Environment & Protection
Humidity fluctuation 0.4 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection, and control specifics

Integration & Operating Notes

Mounting
Install in ambient ≤25 °C for full refrigerating performance

Above 25 °C ambient, refrigerating performance degrades, potentially preventing the chamber from reaching low temperature set-points. Ensure the installation site has adequate ventilation and does not exceed this limit, especially in continuous operation.

Operating Manual, Simple troubleshooting, p.72
Electrical
Supply must match 200–230 V, 1~, 50/60 Hz, 16 A fuse

The 230 V variant requires a 16 A branch circuit protection and a CEE 7/7 outlet. Connecting to a different voltage or using a fuse of lower rating can cause malfunction or nuisance tripping. The UL variant (100–120 V) is a separate model and not interchangeable.

Data Sheet Model KBF-S ECO 1020 en.pdf; Type plate, p.9
Maintenance
Annual maintenance and electrical safety test required

Regular maintenance must be performed at least once a year, including steam humidifier maintenance (interval depends on water quality). Electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after repairs. Warranty may be void if maintenance is done by unauthorized personnel.

Operating Manual, Maintenance intervals, p.70; Electrical testing, p.70

Selection & Integration Q&A

When should I choose the KBF-S ECO 1020 over the smaller 720 or 400 models?

Choose the 1020 when you need the largest capacity (1,020 L) for high-throughput testing or large batches. The 720 and 400 offer the same temperature range but smaller volumes; the 1020 also includes anti-condensation protection (like 400 and 720) but has higher power (1.2 kW vs. lower in smaller models). Select the 1020 if your sample sizes or batch counts require the extra space and you can accommodate the larger footprint (1,254 mm wide).

What are the requirements for the water supply and drain when integrating the humidity system?

The chamber requires a freshwater supply with suitable water quality (conductivity and hardness per manual) and a condensate collection pan. BINDER Pure Aqua Service is available as an accessory for water treatment. Ensure the supply pressure and flow meet the humidifier's demand, and provide a drain for condensate. The steam humidifier needs annual maintenance, with interval depending on water quality.

Can the chamber be installed in a room that occasionally exceeds 30 °C ambient?

No, ambient temperatures above 25 °C reduce refrigerating performance, and prolonged operation above this limit may prevent the chamber from maintaining low temperature set-points. The installation site must be kept below 25 °C for full performance. If occasional excursions occur, ensure they are brief and monitor chamber performance.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance is mandatory, including steam humidifier service (interval depends on water quality) and electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional accessories include APT-COM™ 4 software for networking up to 100 chambers, BINDER Pure Aqua Service for water treatment, and additional racks (max load 45 kg per rack). USB sticks for data transfer are not included.