BINDER KBF-S ECO 400
BINDER ·Made in Germany · In Active Production

BINDER KBF-S ECO 400

BINDER KBF-S ECO 400 — Climate chambers with Temperature 0…70 °C.

The KBF-S ECO 400 is a 400 L constant climate chamber using Peltier thermoelectric cooling and an RD4 microprocessor controller, rated for 0–70 °C dry and 5–70 °C with humidity (10–80 % RH). Within the KBF-S ECO series it is the mid-volume unit with anti-condensation protection, sized for long-term and accelerated testing at up to 30 kg per rack.

Temperature
0…70 °C

BINDER KBF-S ECO Series — Model Comparison & Selection

Model Temperature, °CHumidity fluctuation, % rhInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
KBF-S ECO 400 Current 0–700.3400303001.1
KBF-S ECO 1020 0–700.41020454001.2
KBF-S ECO 240 0–700.3247302500.8
KBF-S ECO 720 0–700.3700454001.2

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 400 L
Load per rack 30 kg
Maximum heat compensation 300 W
Nominal power 1.1 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 6 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 15 min
Dimensions, Weight & Mechanical Design
Width 925 mm
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz
Rated voltage 200...230 V (KBFSECO400-230V)
Unit fuse 16.0 A
Operating Conditions, Environment & Protection
Humidity fluctuation 0.3 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in different media

Control, installation and operation

Integration & Operating Notes

Electrical
Match the ordered voltage variant before wiring the supply

The 230 V variant (Art. 9020-0463) requires 200–230 V 1~ 50/60 Hz on a CEE 7/7 plug and is rated 1.1 kW with a 16.0 A unit fuse; the UL variant (Art. 9020-0464) requires 100–120 V on a NEMA 5-20 plug. Supplying the wrong voltage causes malfunction, and the branch circuit must be sized for the unit fuse rating.

Data Sheet Model KBF-S ECO 400 en.pdf; Operating Manual, Type plate p.9
Process Limits
Humidity operation narrows the usable temperature range

With the humidification system active the chamber is limited to 5–70 °C; the full 0–70 °C range is available only in dry operation. Test profiles that combine low temperature with humidity must be checked against this limit before programming the RD4 controller.

Factory Specification Sheet; Operating Manual, Technical description
Process Limits
Keep heat load and ambient below the Peltier cooling budget

Maximum heat compensation is 300 W, and ambient temperature above 25 °C reduces refrigerating performance. Excessive external heat load or operation outside the optimum temperature/humidity diagram causes low or no refrigerating performance, so charging material and room conditions must be budgeted against these limits.

Operating Manual, Simple troubleshooting p.72
Process Limits
Respect humidity sensor exposure limits for volatile substances

The humidity sensor tolerates only specified permanent concentrations of substances such as ammonia (5500 ppm), acetone (3300 ppm) and formaldehyde (2400 ppm); exceeding these damages the sensor. Sensor tolerance is ±2 % RH, so test atmospheres containing solvents or amines must be checked against the resistance table before use.

Operating Manual, Resistance of the humidity sensor p.17
Maintenance
Treat a repeated fuse trip as a fault, not a consumable

If the chamber fuse has responded, check and replace it with the correct rating; if it responds again, contact BINDER service rather than continuing to replace it. A recurring trip indicates an internal fault that must be diagnosed before returning the chamber to a test program.

Operating Manual, Simple troubleshooting p.72

Selection & Integration Q&A

When should I choose the KBF-S ECO 400 over the 240 or 720 siblings?

Pick the 400 L when 240 L is too small but 720 L is unnecessary; unlike the 240 it includes anti-condensation protection, shared with the 720 and 1020. All siblings cover 0–70 °C dry; the 400 adds 400 L volume with 30 kg per rack and 300 W heat compensation.

What supply and plug does the 230 V KBF-S ECO 400 need, and can I use the UL variant instead?

The 230 V unit (Art. 9020-0463) needs 200–230 V 1~ 50/60 Hz on a CEE 7/7 plug, rated 1.1 kW with a 16.0 A unit fuse. The UL variant (Art. 9020-0464) is a separate 100–120 V model with NEMA 5-20 plug; choose it only where that mains is available.

What water and drain connections are required for humidity operation?

Humidification needs a freshwater supply of suitable water quality, with the BINDER Pure Aqua Service accessory available for water treatment, and a condensate collection pan is fitted. The installation section covers water quality, supply installation and the pan; plan these before siting the chamber.

How often must the chamber and its humidifier be serviced, and what is not included in base delivery?

Chamber maintenance is required at least annually, and the steam humidifier at least once a year depending on water quality and steam output; electrical retest per EN 50678/EN 50699 is required before startup and after repair. APT-COM 4 software and the Pure Aqua Service are accessories, not base delivery.