BINDER KBF-S ECO 720 — Climate chambers with Temperature 0…70 °C.
The KBF-S ECO 720 is a 700 L constant climate chamber using Peltier thermoelectric cooling and the RD4 microprocessor controller, with a 0...70 °C temperature range and 10...80 % RH humidity range (5...70 °C with humidity active). It is the mid-large volume member of the KBF-S ECO series, distinguished from the 240/400/1020 siblings by its 700 L interior, intended for long-term and accelerated stability testing under continuous duty.
BINDER KBF-S ECO Series — Model Comparison & Selection
| Model | Temperature, °C | Humidity fluctuation, % rh | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| KBF-S ECO 720 Current | 0–70 | 0.3 | 700 | 45 | 400 | 1.2 |
| KBF-S ECO 1020 | 0–70 | 0.4 | 1020 | 45 | 400 | 1.2 |
| KBF-S ECO 240 | 0–70 | 0.3 | 247 | 30 | 250 | 0.8 |
| KBF-S ECO 400 | 0–70 | 0.3 | 400 | 30 | 300 | 1.1 |
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 700 L (KBF-S ECO 720) |
| Load per rack | 45 kg |
| Maximum heat compensation | 400 W |
| Nominal power | 1.2 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.1 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 7 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 17 min |
| Depth | 885 mm |
| Height | 1925 mm |
| Width | 1254 mm |
| Power frequency | 50/60 Hz |
| Rated voltage | 200...230 V (KBFSECO720-230V) |
| Unit fuse | 16.0 A |
| Humidity fluctuation | 0.3 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The KBF-S ECO 720 (700 L interior, 1,254 mm wide, 885 mm deep, 1,925 mm high) sits between the 400 and 1020 models in the KBF-S ECO series; all siblings share the 0...70 °C range, so the choice is driven by chamber volume and load, not by temperature capability.
- Cooling is by Peltier thermoelectric elements rather than a compressor circuit; the RD4 microprocessor controller manages temperature and humidity set-points, programs, alarms and data recording.
- The chamber is rated for continuous operation and meets the requirements for long-term and accelerated testing.
- Two electrical variants exist: KBFSECO720-230V (200...230 V, 1~, 50/60 Hz, CEE 7/7 plug) and KBFSECO720UL-120V (100...120 V, 1~, 50/60 Hz, NEMA 5-20 plug).
Behavior under load and in different media
- Temperature fluctuation and uniformity are both 0.1 K; humidity fluctuation is 0.3 % RH, with a humidity sensor tolerance of ±2 % r.h.
- Maximum heat compensation is 400 W; exceeding the external heat load or operating outside the optimum temperature/humidity range shown in the manual's diagram degrades or stops refrigerating performance.
- Recovery after a 30 s door opening is 7 min at 25 °C / 60 % RH and 17 min at 40 °C / 75 % RH.
- Ambient temperature above 25 °C reduces refrigerating performance, so the installation site must stay at or below that for full performance.
- The humidity sensor is resistant to defined permanent concentrations of substances (e.g. ammonia 5500 ppm, acetone 3300 ppm, formaldehyde 2400 ppm); exceeding these damages the sensor.
Installation, control and operation
- Installation requires wall spacers for distance, a freshwater supply of suitable quality (BINDER Pure Aqua Service accessory available), a condensate collection pan, and compliance with the specified ambient and site requirements.
- The RD4 controller provides access levels, password protection (User/Admin), tolerance-range and alarm-delay settings, program start/stop with defined behavior after power failure, Ethernet network configuration, a DL1 data recorder with settable storage rate, and USB import/export.
- Temperature safety devices are an over-temperature protective device class 1 and an overtemperature safety controller class 3.1 per DIN 12880; the safety controller value must be set above normal operating temperature.
- Anti-condensation protection is provided on the 400, 720 and 1020 models; defrosting procedures apply during refrigerating operation.
- APT-COM™ 4 software can network up to 100 chambers.
Integration & Operating Notes
The 230 V variant requires 200...230 V 1~ 50/60 Hz with a CEE 7/7 plug and a 16.0 A unit fuse; the UL variant requires 100...120 V with a NEMA 5-20 plug. Connecting a variant to the wrong voltage can cause malfunction, and the supply circuit must be rated for the 1.2 kW nominal power.
Combinations of temperature and humidity outside the optimum range in the manual's temperature-humidity diagram (Figure 19) can produce low or no refrigerating performance. With humidity active the working range narrows to 5...70 °C, and the 400 W maximum heat compensation must not be exceeded by the load.
Ambient temperature above 25 °C / 77 °F reduces refrigerating performance, so the installation site must be selected and ventilated accordingly. Wall spacers for distance, the condensate collection pan and a suitable freshwater supply must be provided per the installation section.
Regular maintenance work and steam humidifier maintenance must be performed at least once a year, with humidifier interval depending on water quality and steam output. Repeat electrical inspection and test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after maintenance or repair; warranty is void if maintenance is done by non-authorized personnel.
If the chamber fuse has responded, check it and replace with the correct rating; if it responds again, contact BINDER service rather than continuing to replace it, since a repeated trip indicates a fault in the chamber.
Selection & Integration Q&A
All KBF-S ECO models share the 0...70 °C range, so select on volume and load: the 720 offers 700 L with 400 W maximum heat compensation and 45 kg per rack. Choose the 400 for smaller batches or the 1020 when more chamber volume is needed; the 720 fits mid-size stability studies.
The KBFSECO720-230V needs 200...230 V 1~ 50/60 Hz via CEE 7/7 with a 16.0 A unit fuse and 1.2 kW nominal power. The KBFSECO720UL-120V instead requires 100...120 V 1~ with a NEMA 5-20 plug; the supply circuit must be sized for the variant ordered.
Install with wall spacers for distance, a condensate collection pan and a suitable freshwater supply for humidification. Ambient temperature above 25 °C reduces refrigerating performance, so the room must stay at or below that; IP 20 means the chamber must not become wet.
Annual maintenance and annual steam humidifier service are required, plus electrical retesting per EN 50678/VDE 0701 and EN 50699/VDE 0702. APT-COM™ 4 software and the BINDER Pure Aqua Service water-treatment accessory are options, not part of base delivery.