BINDER KBF-S ECO 720
BINDER ·Made in Germany · In Active Production

BINDER KBF-S ECO 720

BINDER KBF-S ECO 720 — Climate chambers with Temperature 0…70 °C.

The KBF-S ECO 720 is a 700 L constant climate chamber using Peltier thermoelectric cooling and the RD4 microprocessor controller, with a 0...70 °C temperature range and 10...80 % RH humidity range (5...70 °C with humidity active). It is the mid-large volume member of the KBF-S ECO series, distinguished from the 240/400/1020 siblings by its 700 L interior, intended for long-term and accelerated stability testing under continuous duty.

Temperature
0…70 °C

BINDER KBF-S ECO Series — Model Comparison & Selection

Model Temperature, °CHumidity fluctuation, % rhInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
KBF-S ECO 720 Current 0–700.3700454001.2
KBF-S ECO 1020 0–700.41020454001.2
KBF-S ECO 240 0–700.3247302500.8
KBF-S ECO 400 0–700.3400303001.1

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 700 L (KBF-S ECO 720)
Load per rack 45 kg
Maximum heat compensation 400 W
Nominal power 1.2 kW
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Article Number mfr 9020
Recovery time after door was opened for 30 s at 25°C and 60% RH mfr 7 min
Recovery time after door was opened for 30 s at 40°C and 75% RH mfr 17 min
Dimensions, Weight & Mechanical Design
Depth 885 mm
Height 1925 mm
Width 1254 mm
Electrical, Power & Drive Specifications
Power frequency 50/60 Hz
Rated voltage 200...230 V (KBFSECO720-230V)
Unit fuse 16.0 A
Operating Conditions, Environment & Protection
Humidity fluctuation 0.3 % RH
Humidity range 10...80 % RH

Description

Construction and series positioning

Behavior under load and in different media

Installation, control and operation

Integration & Operating Notes

Electrical
Match the supply to the ordered voltage variant before connection

The 230 V variant requires 200...230 V 1~ 50/60 Hz with a CEE 7/7 plug and a 16.0 A unit fuse; the UL variant requires 100...120 V with a NEMA 5-20 plug. Connecting a variant to the wrong voltage can cause malfunction, and the supply circuit must be rated for the 1.2 kW nominal power.

Data Sheet Model KBF-S ECO 720 en.pdf; Operating Manual, Type plate p.9
Process Limits
Stay inside the optimum temperature/humidity diagram for cooling

Combinations of temperature and humidity outside the optimum range in the manual's temperature-humidity diagram (Figure 19) can produce low or no refrigerating performance. With humidity active the working range narrows to 5...70 °C, and the 400 W maximum heat compensation must not be exceeded by the load.

Operating Manual, Simple troubleshooting p.72; Factory Specification Sheet
Mounting
Keep ambient at or below 25 °C for full refrigerating performance

Ambient temperature above 25 °C / 77 °F reduces refrigerating performance, so the installation site must be selected and ventilated accordingly. Wall spacers for distance, the condensate collection pan and a suitable freshwater supply must be provided per the installation section.

Operating Manual, Simple troubleshooting p.72; Installation and connections
Maintenance
Plan annual maintenance and electrical retesting

Regular maintenance work and steam humidifier maintenance must be performed at least once a year, with humidifier interval depending on water quality and steam output. Repeat electrical inspection and test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after maintenance or repair; warranty is void if maintenance is done by non-authorized personnel.

Operating Manual, Maintenance intervals p.70; Electrical testing p.70
Maintenance
A repeatedly blown chamber fuse requires BINDER service

If the chamber fuse has responded, check it and replace with the correct rating; if it responds again, contact BINDER service rather than continuing to replace it, since a repeated trip indicates a fault in the chamber.

Operating Manual, Simple troubleshooting p.72

Selection & Integration Q&A

When should I choose the KBF-S ECO 720 rather than the 400 or 1020 sibling?

All KBF-S ECO models share the 0...70 °C range, so select on volume and load: the 720 offers 700 L with 400 W maximum heat compensation and 45 kg per rack. Choose the 400 for smaller batches or the 1020 when more chamber volume is needed; the 720 fits mid-size stability studies.

What supply and protection must be provided for the 230 V variant, and what does the UL variant change?

The KBFSECO720-230V needs 200...230 V 1~ 50/60 Hz via CEE 7/7 with a 16.0 A unit fuse and 1.2 kW nominal power. The KBFSECO720UL-120V instead requires 100...120 V 1~ with a NEMA 5-20 plug; the supply circuit must be sized for the variant ordered.

What site conditions and ambient limits apply when installing the chamber?

Install with wall spacers for distance, a condensate collection pan and a suitable freshwater supply for humidification. Ambient temperature above 25 °C reduces refrigerating performance, so the room must stay at or below that; IP 20 means the chamber must not become wet.

What maintenance and accessories are required beyond the base chamber?

Annual maintenance and annual steam humidifier service are required, plus electrical retesting per EN 50678/VDE 0701 and EN 50699/VDE 0702. APT-COM™ 4 software and the BINDER Pure Aqua Service water-treatment accessory are options, not part of base delivery.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KBF S ECO 720 en
Datasheet ·710,2 КБ ·8 pp.
View PDF
PDF
KBF S ECO (E6) 07 2026 en
Manual ·9,8 МБ ·99 pp.
View PDF