BINDER KBF-S ECO 400
BINDER KBF-S ECO 400 — Climate chambers with Temperature 0…70 °C.
The KBF-S ECO 400 is a 400 L constant climate chamber using Peltier thermoelectric cooling and an RD4 microprocessor controller, rated for 0–70 °C dry and 5–70 °C with humidity (10–80 % RH). Within the KBF-S ECO series it is the mid-volume unit with anti-condensation protection, sized for long-term and accelerated testing at up to 30 kg per rack.
BINDER KBF-S ECO Series — Model Comparison & Selection
| Model | Temperature, °C | Humidity fluctuation, % rh | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| KBF-S ECO 400 Current | 0–70 | 0.3 | 400 | 30 | 300 | 1.1 |
| KBF-S ECO 1020 | 0–70 | 0.4 | 1020 | 45 | 400 | 1.2 |
| KBF-S ECO 240 | 0–70 | 0.3 | 247 | 30 | 250 | 0.8 |
| KBF-S ECO 720 | 0–70 | 0.3 | 700 | 45 | 400 | 1.2 |
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 400 L |
| Load per rack | 30 kg |
| Maximum heat compensation | 300 W |
| Nominal power | 1.1 kW |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.1 K |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 25°C and 60% RH mfr | 6 min |
| Recovery time after door was opened for 30 s at 40°C and 75% RH mfr | 15 min |
| Width | 925 mm |
| Power frequency | 50/60 Hz |
| Rated voltage | 200...230 V (KBFSECO400-230V) |
| Unit fuse | 16.0 A |
| Humidity fluctuation | 0.3 % RH |
| Humidity range | 10...80 % RH |
Description
Construction and series positioning
- The chamber is a 400 L constant climate cabinet with Peltier thermoelectric cooling rather than a compressor circuit; the same series spans 240, 400, 720 and 1020 L, all with a 0–70 °C dry range.
- This 400 L modification is one of the three sizes (400, 720, 1020) that carry anti-condensation protection; the 240 L sibling does not.
- Two electrical variants exist: KBFSECO400-230V (Art. 9020-0463, 200–230 V 1~ 50/60 Hz, CEE 7/7 plug) and KBFSECO400UL-120V (Art. 9020-0464, 100–120 V 1~ 50/60 Hz, NEMA 5-20 plug). The 230 V variant is rated 1.1 kW with a 16.0 A unit fuse.
- Interior volume is 400 L; net width is 925 mm; rack load is limited to 30 kg per rack.
Behavior under load and in different media
- Temperature fluctuation and uniformity are both stated as 0.1 K; humidity fluctuation is 0.3 % RH over a 10–80 % RH working range.
- With humidity active the temperature range narrows to 5–70 °C; the full 0–70 °C range applies to dry operation only.
- Recovery after a 30 s door opening is 6 min at 25 °C / 60 % RH and 15 min at 40 °C / 75 % RH, so door-opening frequency directly sets achievable duty cycle.
- Maximum heat compensation is 300 W; external heat load above this, or operation outside the optimum temperature/humidity envelope, degrades or stops refrigerating performance.
- Ambient temperature above 25 °C reduces refrigerating performance, so the installation site must stay at or below that for full capability.
Control, installation and operation
- The RD4 controller provides set-point entry for temperature and humidity, tolerance-range alarms with delay, program start/stop, a DL1 data recorder, USB import/export, Ethernet configuration and a self-test function.
- Temperature safety is two-tier: an over-temperature protective device (class 1) that trips at nominal +10 °C, and an over-temperature safety controller class 3.1 per DIN 12880 that must be set and function-checked.
- Water supply for humidification requires suitable water quality; a BINDER Pure Aqua Service accessory and freshwater supply installation are described, and a condensate collection pan is fitted.
- Installation requires wall spacers and, for size 400, a flexible tilt protection kit; IP protection is 20, so the chamber must not become wet.
- Networking up to 100 chambers is possible with APT-COM 4 Multi Management Software, which is an accessory, not part of base delivery.
Integration & Operating Notes
The 230 V variant (Art. 9020-0463) requires 200–230 V 1~ 50/60 Hz on a CEE 7/7 plug and is rated 1.1 kW with a 16.0 A unit fuse; the UL variant (Art. 9020-0464) requires 100–120 V on a NEMA 5-20 plug. Supplying the wrong voltage causes malfunction, and the branch circuit must be sized for the unit fuse rating.
With the humidification system active the chamber is limited to 5–70 °C; the full 0–70 °C range is available only in dry operation. Test profiles that combine low temperature with humidity must be checked against this limit before programming the RD4 controller.
Maximum heat compensation is 300 W, and ambient temperature above 25 °C reduces refrigerating performance. Excessive external heat load or operation outside the optimum temperature/humidity diagram causes low or no refrigerating performance, so charging material and room conditions must be budgeted against these limits.
The humidity sensor tolerates only specified permanent concentrations of substances such as ammonia (5500 ppm), acetone (3300 ppm) and formaldehyde (2400 ppm); exceeding these damages the sensor. Sensor tolerance is ±2 % RH, so test atmospheres containing solvents or amines must be checked against the resistance table before use.
If the chamber fuse has responded, check and replace it with the correct rating; if it responds again, contact BINDER service rather than continuing to replace it. A recurring trip indicates an internal fault that must be diagnosed before returning the chamber to a test program.
Selection & Integration Q&A
Pick the 400 L when 240 L is too small but 720 L is unnecessary; unlike the 240 it includes anti-condensation protection, shared with the 720 and 1020. All siblings cover 0–70 °C dry; the 400 adds 400 L volume with 30 kg per rack and 300 W heat compensation.
The 230 V unit (Art. 9020-0463) needs 200–230 V 1~ 50/60 Hz on a CEE 7/7 plug, rated 1.1 kW with a 16.0 A unit fuse. The UL variant (Art. 9020-0464) is a separate 100–120 V model with NEMA 5-20 plug; choose it only where that mains is available.
Humidification needs a freshwater supply of suitable water quality, with the BINDER Pure Aqua Service accessory available for water treatment, and a condensate collection pan is fitted. The installation section covers water quality, supply installation and the pan; plan these before siting the chamber.
Chamber maintenance is required at least annually, and the steam humidifier at least once a year depending on water quality and steam output; electrical retest per EN 50678/EN 50699 is required before startup and after repair. APT-COM 4 software and the Pure Aqua Service are accessories, not base delivery.