BINDER KT 53
BINDER KT 53 — Incubators with Temperature 4…100 °C.
The BINDER KT 53 is a 53 L Peltier-cooled incubator (KT053-230V, Art. 9020-0311) operating from 4 to 100 °C with 0.1 K fluctuation and 0.3 K uniformity. It is the smallest chamber in the KT series, distinguished by a 0.4 kW nominal rating and 40 kg permitted load. It serves refrigerated incubation, long-term 4 °C storage, and alternating 37/4 °C protocols in microbiology.
Direct drop-in replacements
BINDER KT Series — Model Comparison & Selection
Full specifications
| Temperature | 4...100 °C |
| Interior volume | 53 L |
| Load per rack | 15 kg |
| Max. heat compensation | 100 W |
| Nominal power | 0.4 kW |
| Permitted load | 40 kg (KT053-230V / KT053UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 100 mm (KT053-230V / KT053UL-120V) |
| Side wall clearance | 240 mm (KT053-230V / KT053UL-120V) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 2 min |
| Depth | 630 mm (KT053-230V / KT053UL-120V) |
| Height | 635 mm (KT053-230V / KT053UL-120V) |
| Interior width | 400 mm (KT053-230V / KT053UL-120V) |
| Width | 660 mm (KT053-230V / KT053UL-120V) |
| Net weight of the unit (empty) mfr | 63 kg |
| Power frequency | 50 Hz |
| Rated voltage | 200...240 V |
| Unit fuse | 10.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Peltier refrigerating technology with HIT insulation and APT.line preheating chamber system; T4.12 program controller with graphical display.
- Interior 400 mm wide × 630 mm deep × 635 mm high (53 L usable); net weight 63 kg; exterior 660 mm wide.
- Smallest of the KT series (siblings KT 115 and KT 170 share the 4...100 °C range but larger volumes); the KT 53 is the choice when bench/footprint and 53 L capacity suffice.
- Two voltage variants: KT053-230V (200...240 V, 1~, 50 Hz, CEE 7/7 plug) and KT053UL-120V (100...120 V, 1~, 50/60 Hz, NEMA 5-20 plug, UL certified).
Behavior under load and in operation
- Temperature fluctuation 0.1 K and uniformity 0.3 K across the 4...100 °C working range; recovery to setpoint after a 30 s door opening at 37 °C is 2 min.
- Maximum heat compensation 100 W; permitted load 40 kg total, 15 kg per rack.
- Intended use covers 4...37 °C for microorganisms with narrow temperature optima: long-term 4 °C storage, 20–25 °C refrigerated incubation, 37 °C incubation, and alternating 37/4 °C cycles.
- Loads must not contain explosive or flammable solvents; operating temperature must stay below the flash or sublimation point of the charge.
Installation, control and operation
- T4.12 controller supports fixed-value mode, time programs and week programs with setpoint ramps/steps, tolerance ranges and repetition; key lock with password.
- Data export/import via USB; optional APT-COM 4 software, Pt 100 object temperature sensor, door heating and zero-voltage relay alarm/control outputs.
- Overtemperature protective device class 1 and overtemperature safety controller class 3.1 per DIN 12880:2007; IP20 protection.
- Alarm system with set/acknowledged/cleared states, event list, buzzer test and adjustable door alarm delay.
Integration & Operating Notes
The Peltier system rejects heat at the rear; insufficient clearance or installation in an unventilated recess causes overheating and chamber damage. Plan the bench or floor position with these minimum distances to walls and adjacent equipment before commissioning.
The KT053-230V is rated 200...240 V, 1~, 50 Hz with a 10.0 A unit fuse and 0.4 kW nominal power. Verify the supply voltage and frequency before connection; the UL variant (KT053UL-120V) is a separate 100...120 V model with NEMA 5-20 plug and must not be connected to a 230 V circuit.
Charging material must not contain explosive or flammable solvents, and the operating temperature must lie below the flash or sublimation point of the load. Assess the physical and chemical properties of every charge before operation to avoid forming an explosive mixture with air inside the chamber.
Alarm states are set (active, buzzer on), acknowledged (active, buzzer off, icon remains) and cleared (cause gone). Acknowledging does not clear the alarm until the cause is remedied; operators must not assume the process is back in specification after pressing acknowledge.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; warranty is void if non-authorized personnel conduct it. An annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is also required, plus a test before initial startup and after any repair.
With increased dust in ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year to maintain cooling performance. Before any maintenance, switch off at the main power switch, disconnect the plug, and do not remove the rear panel.
Replacing the door gasket while the chamber is warm can damage the gasket. Allow the chamber to cool to ambient before gasket replacement, and test the chamber before resuming operation after any maintenance or repair.
The chamber is protected against solid objects larger than 12.5 mm but not against water (IP20 per EN 60529). Install it in a dry, splash-proof location and do not wet the chamber during cleaning or decontamination.
Selection & Integration Q&A
Pick the KT 53 when 53 L interior (400×630×635 mm) and 40 kg load capacity cover your workload and bench space is limited. Move to KT 115 or KT 170 only for larger batches; all three share the 4...100 °C range and Peltier technology.
The KT053-230V needs 200...240 V, 1~, 50 Hz with CEE 7/7 plug and 10.0 A fuse. The KT053UL-120V is a separate 100...120 V model with NEMA 5-20 plug and UL certification; it must not be connected to a 230 V supply.
Keep 100 mm clearance at the rear and 240 mm at the sides for Peltier heat rejection, and do not install in an unventilated recess. The IP20 enclosure must be in a dry, splash-proof location; dusty air requires more frequent fan-grid cleaning.
Annual maintenance by BINDER-authorized personnel and annual electrical repeat inspection per EN 50678/EN 50699 are mandatory. Optional items not in base delivery include APT-COM 4 software, Pt 100 object temperature sensor, door heating and zero-voltage relay outputs.