BINDER M 720
BINDER M 720 — Drying ovens with Temperature 14…300 °C.
The BINDER M 720 is a forced-convection drying and heating oven with a 735 L interior, configured for 400 V 3/N/PE 50 Hz supply and 6 kW nominal power. It reaches +14 to 300 °C with 2.0 K uniformity and 0.6 K fluctuation, distinguishing it from smaller M-series siblings by its 735 L volume and three-phase connection. It serves high-throughput drying, heating, and thermal testing where large batches must be processed within tight temperature tolerances.
BINDER M Series — Model Comparison & Selection
Full specifications
| Temperature | 14...300 °C |
| Air change rate | 40 x/h |
| Heating time to 150°C | 14 min |
| Interior volume | 735 L |
| Load per rack | 45 kg |
| Nominal power | 6 kW |
| Permitted load | 315 kg |
| Temperature fluctuation | 0.6 K |
| Temperature uniformity | 2.0 K |
| Rear wall clearance | 160 mm |
| Side wall clearance | 100 mm |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 6 min |
| Depth | 816 mm |
| Height | 1590 mm |
| Width | 1160 mm |
| Net weight of the unit (empty) mfr | 170 kg |
| Power frequency | 50 Hz |
| Rated voltage | 400 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 3 N/PE |
Description
Construction and series positioning
The M 720 (E3.1, Art. No. 9010-0374) is the largest chamber in the M series, with a 735 L interior volume and net dimensions of 1,160 mm W × 1,590 mm H × 816 mm D. Unlike the M 56 and M 115, which operate on 230 V single-phase and start at +23 °C, the M 720 uses a 400 V 3/N/PE 50 Hz supply with 6 kW nominal power and reaches down to +14 °C. The M 260 shares the +14 °C lower limit but has a smaller chamber. The M 720 is equipped with the MB2 microprocessor chamber controller and forced convection, and its main power switch is located on the rear panel (as on the M 260).
Thermal behavior and load capacity
- Temperature range: +14 to 300 °C; uniformity 2.0 K; fluctuation 0.6 K.
- Heating-up time to 150 °C: 14 min; recovery after 30 s door opening at 150 °C: 6 min.
- Air-change rate: 40 x/h, supporting solvent drying and fresh-air exchange.
- Permitted load: 315 kg total; maximum load per rack: 45 kg.
The chamber maintains tight uniformity across its large volume, but loading must respect the 45 kg per-rack limit and 315 kg total to avoid rack deformation or airflow obstruction.
Installation, connection, and control
- Electrical: 400 V 3/N/PE, 50 Hz, 6 kW, unit fuse 16.0 A; supply must be within ±10% of nominal.
- Clearances: 160 mm at the rear and 100 mm at the sides for heat dispersal; do not install in unventilated recesses.
- Rear interfaces: SUB-D socket 'ALARM / CONTACT' for zero-voltage relay alarm contact; optional SUB-D 'ANALOG OUTPUT' for temperature.
- Controller: MB2 with fixed-value, timer, time-program, and week-program modes; overtemperature protection class 1 standard, class 2/3.1 selectable.
- Options include APT-COM 4 software (up to 100 chambers), flexible Pt100 sensor, HEPA fresh air filter, and inert gas connection with mostly gas-tight version.
Safety and environment
The chamber is IP20 and must not be installed in damp areas or puddles. It is not explosion-protected; combustible dusts, solvent-air mixtures, and substances combustible at working temperature must be excluded. Inner surfaces, exhaust duct, door gaskets, and access ports become hot during operation.
Integration & Operating Notes
The M 720 draws 6 kW at 400 V 3/N/PE 50 Hz and is protected by a 16.0 A unit fuse. Supply voltage outside ±10% of nominal can cause controller faults, heater stress, or tripping. Verify phase rotation and neutral connection before first start-up; the main power switch is on the rear panel, so ensure accessible isolation.
Insufficient clearance prevents heat dispersal and can overheat the chamber or adjacent surfaces. Do not install in unventilated recesses. The chamber must be set up splash-proof and away from damp areas because IP20 offers no water protection.
The chamber has no explosion protection. Do not introduce substances that are combustible or explosive at working temperature; the set-point must remain below the flash point or sublimation point of the charge. Solvent-air mixtures must be kept away from the chamber.
The rear SUB-D 'ALARM / CONTACT' provides a zero-voltage relay alarm contact. Connect it to a plant alarm or interlock to signal overtemperature or power failure. The optional analog output can feed a chart recorder. APT-COM 4 can network up to 100 chambers.
Maintenance must be performed at least once a year by licensed electricians or experts authorized by BINDER; otherwise warranty is void. Before maintenance, disconnect power via the main switch and pull the plug. Replace door gasket only when cold. After maintenance, test before resuming operation.
If the class 1 overtemperature protective device trips, disconnect the chamber from power for at least 10 seconds and let it cool. If it trips again, contact BINDER service. Do not bypass or repeatedly reset without investigating the cause.
Selection & Integration Q&A
Choose the M 720 for large batches requiring 735 L volume and 315 kg total load. The M 260 offers the same +14 °C lower limit but smaller volume; the M 115 and M 56 are 230 V single-phase and start at +23 °C, suiting smaller loads without three-phase supply.
The M 720 requires 400 V 3/N/PE 50 Hz, 6 kW nominal power, and a 16.0 A unit fuse. Supply must be within ±10% of nominal. The main power switch is on the rear, so provide accessible isolation. Verify phase rotation and neutral before start-up.
Maintain 160 mm rear and 100 mm side clearance for heat dispersal; do not install in unventilated recesses. IP20 requires a dry, splash-proof location away from puddles. The chamber is not explosion-protected; exclude combustible dusts and solvent-air mixtures.
Annual maintenance by BINDER-authorized personnel is mandatory to preserve warranty, plus annual electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional accessories include APT-COM 4 software, flexible Pt100 sensor, analog output, HEPA fresh air filter, and inert gas connection; these are not included in the standard model.