BINDER MKFT 720
BINDER ·Made in Germany · In Active Production

BINDER MKFT 720

BINDER MKFT 720 — Climate chambers with Temperature -70…180 °C.

The MKFT 720 is a 734 L deep-temperature alternating climate chamber with a -70…180 °C range and 10…98 %RH humidity control, powered by an 11 kW 400 V 3 N/PE 50 Hz supply. It is the largest MKFT variant, distinguished from the MKFT 115/240 by its 734 L interior and 40 kg per-rack load capacity, and is intended for dynamic temperature-humidity cycling of large test loads in qualification and reliability testing.

Temperature
-70…180 °C

BINDER MKFT Series — Model Comparison & Selection

Model Temperature, °CHumidity range, % rhInterior volume, LLoad per rack, kgMax heat compensation, WNominal power, kW
MKFT 720 Current -70–18010...987344080011
MKFT 115 -70–18095115304005
MKFT 240 -70–18010...98228304006

Full specifications

Core Performance & Functional Specifications
Temperature -70...180 °C
Interior volume 734 L (MKFT720-400V)
Load per rack 40 kg (MKFT720-400V)
Maximum heat compensation 800 W
Nominal power 11 kW (MKFT720-400V)
Article Number mfr 9020
Electrical, Power & Drive Specifications
Power frequency 50 Hz (MKFT720-400V)
Rated voltage 400 V
Unit fuse 32.0 A (MKFT720-400V)
Operating Conditions, Environment & Protection
Humidity range 10...98 % RH
Cooling down time from 180°C to -70°C mfr 120 min
Humidity fluctuation with humidity dependent on set value mfr 2.5 % r.F.
Standards, Safety & General Data
Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr 4.8 K/min

Description

Construction and series positioning

Thermal and humidity behavior

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be 400 V 3 N/PE 50 Hz with 32 A fusing

The 11 kW nominal power and 32.0 A unit fuse require a dedicated three-phase supply with neutral and protective earth. Verify the building distribution can deliver the run-up current without nuisance tripping; the -C variant is needed for 480 V 60 Hz sites.

Data Sheet Model MKFT 720 en.pdf; Operating Manual MKF/MKFT (E5) 10/2026, Section 'Electrical connection'
Process Limits
Humidity sensor is damaged by ammonia and ozone above limits

The humidity sensor must not be exposed to ammonia above 20 ppm (14 mg/m³) workplace threshold or 5500 ppm (4000 mg/m³) permanent load, nor ozone above 0.1 ppm (0.2 mg/m³) or 0.5 ppm (1 mg/m³) permanent. Verify process atmospheres before introducing test specimens that outgas these substances.

Operating Manual MKF/MKFT (E5) 10/2026, p.20 table
Control & I/O
Both main and rear power switches must be on for size 720

From size 115 upward, the chamber has a rear power switch in addition to the main power switch. During commissioning or after troubleshooting, confirm both switches are in the on position; otherwise the controller will not energize and no heating, cooling or humidity function will operate.

Operating Manual MKF/MKFT (E5) 10/2026, p.24-25, p.135
Maintenance
Annual maintenance and leak test are mandatory

Schedule at least one maintenance per year, including a leak test per Regulation (EU) 517/2014 Article 4 and Article 10(1)(b). Document the leak test in the service log book. Steam humidifier maintenance is also required at least annually, with intervals depending on water quality and steam output.

Operating Manual MKF/MKFT (E5) 10/2026, p.134
Maintenance
Service message at 8760 h or two years requires BINDER reset

After 8760 operating hours or two years, a service message appears on the controller. It can be confirmed but reappears every two weeks until reset by BINDER Service. Plan a service intervention at that mark to avoid repeated alarms and to perform the required maintenance.

Operating Manual MKF/MKFT (E5) 10/2026, p.135
Maintenance
Condenser fan grid must be checked weekly in dusty air

With increased dust in ambient air, clean the condenser fan several times a year. Check the fan grid behind the left maintenance access flap every week and clean by suction after disconnecting the chamber from power. Condensers are recommended to be cleaned every 1 to 2 years by a qualified technician.

Operating Manual MKF/MKFT (E5) 10/2026, p.134-135
Maintenance
Door gasket must be replaced only when cold

Replace the door gasket only after the chamber has cooled to ambient temperature. Attempting replacement while the chamber is warm can deform or damage the new gasket and compromise the seal, leading to humidity and temperature control errors.

Operating Manual MKF/MKFT (E5) 10/2026, p.134

Selection & Integration Q&A

When should I choose the MKFT 720 over the MKFT 115 or MKFT 240?

Choose the MKFT 720 when you need the largest interior (734 L) and a 40 kg per-rack load with up to 800 W heat compensation. The MKFT 115 and 240 share the -70…180 °C range but have smaller volumes and lower load ratings; select them for smaller specimens or when floor space is limited.

What water and drain connections are needed for humidity operation?

Humidity control requires a freshwater supply (automatic via water pipe or manual via internal freshwater can from size 115 on) and a wastewater drain. If the water cooling option is ordered, cooling water inlet and outlet must also be plumbed. Without these, only temperature operation is possible.

What maintenance and service items are not included in the base delivery?

The base delivery does not include the recommended disinfectant spray (Art. No. 1002-0022), replacement door gaskets, or the BINDER Pure Aqua Service. Annual maintenance, leak test, humidifier service and condenser cleaning are scheduled service actions; the 8760-hour service message requires BINDER Service reset.

Dimensional Drawings & Technical Documentation

Dimensional drawings

MKFT 720 — drawing 1

Click drawing to inspect vector dimensions

Technical documentation

PDF
Data Sheet Model MKFT 720 en
Datasheet ·1,0 МБ ·10 pp.
View PDF
PDF
MKF+MKFT (E5) 10 2026 en
Manual ·18,1 МБ ·179 pp.
View PDF