Peter Huber Kältemaschinenbau HB480 Umwälz-Heizgerät
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau HB480 Umwälz-Heizgerät

Peter Huber Kältemaschinenbau HB480 Umwälz-Heizgerät — Thermostatic baths.

The HB480 is a 48 kW external open-loop circulation heater with a magnetically coupled stainless-steel pump (200 l/min, 5.5 bar) and a 60–250 °C working range. It has no expansion tank or active refrigeration, so it serves high-temperature external circuits where the application itself provides fluid filling, volume compensation and heat loss.

Direct drop-in replacements

Eppendorf
MixMate ®
82%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Umwaelz Heizgeraete Series — Model Comparison & Selection

Model Dimensions w d hPower supplyPump connectionPump typeHeating capacity, WAmbient temperature limit, °C
HB480 Umwälz-Heizgerät Current 570 x 700 x 1171400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st480005–40
HB120 Umwälz-Heizgerät 479 x 539 x 695380-460V 3~ 50/60HzM30x1, 5 AG malemagnetically coupled, st—5–40
HB240 Umwälz-Heizgerät 570 x 700 x 1171400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st240005–40
HB60 Umwälz-Heizgerät 323 x 451 x 498400V 3~N 50HzM30x1, 5 AG malePressure pump, speed not60005–40
HB720 Umwälz-Heizgerät 800 x 1007 x 1607400V 3~ 50HzM38x1, 5 AG malemagnetically coupled, st720005–40

Full specifications

Core Performance & Functional Specifications
Dimensions (w × D × H) 570 x 700 x 1171 mm
Heating capacity 48 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range 60...250 °C
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class II / FL

Description

Construction and series positioning

Behavior in operation

Installation and control

Integration & Operating Notes

Process Limits
Minimum 60 °C depends on application heat loss

Without active refrigeration, the 60 °C lower limit is only reachable when ambient is below 60 °C and the application dissipates between 0.1 kW and 2.8 kW depending on the HB model; a loop with lower or higher loss will not settle at the minimum temperature.

Datasheet HB480 Immersion Circulator, page 2
Electrical
Plug not included for sub-63 A three-phase units

The unit is supplied with a cable but without a plug because three-phase devices drawing less than 63 A are delivered this way; the site must provide a matching 400 V 3~ 50 Hz plug and connection hardware.

Datasheet HB480 Immersion Circulator, page 2
Electrical
Voltage and frequency tolerances must not oppose

Tolerance combinations that run in opposite directions are not permitted — for example -10% voltage together with +3% frequency is explicitly disallowed — so site power quality must be checked against this rule before connection.

Datasheet HB480 Immersion Circulator, page 2
Electrical
Class A group 1 EMC points to industrial siting

EN 55011 Class A Group 1 classification typically requires an industrial environment; in residential or light-industrial areas additional filtering or shielding may be needed to meet disturbance limits.

Datasheet HB480 Immersion Circulator, page 2
Electrical
Account for leakage current from external temperature control

Leakage current from the external temperature control circuit must be factored into protective earthing design and residual current device selection; undersized RCDs may nuisance-trip or fail to provide intended protection.

Datasheet HB480 Immersion Circulator, page 2
Mounting
Respect published space requirements around the unit

Layout must leave the space around the unit according to the operating conditions published by the manufacturer; inadequate clearance affects ventilation and access for service.

Datasheet HB480 Immersion Circulator, page 2

Selection & Integration Q&A

When should I choose the HB480 rather than the HB240 or HB120?

Pick the HB480 when the external circuit needs the series' highest heating capacity (48 kW) and the 60–250 °C range; the smaller HB45/HB60/HB120/HB240 suit lower heat loads. All share the no-expansion-tank external open-operation architecture, so the choice is driven by heating power and loop heat loss.

What must the external circuit and power installation provide to mate with the HB480?

The loop must match the M38x1.5 male pump connection and stay within the pump envelope of 200 l/min at 5.5 bar. Power is 400 V 3~ 50 Hz; since units below 63 A ship with cable but no plug, the site supplies the plug. EMC Class A Group 1 favors an industrial installation.

What ambient and duty conditions apply to the HB480 installation?

Ambient must stay within 5–40 °C, and the 60 °C minimum temperature is only reachable with ambient below 60 °C and application heat loss between 0.1 kW and 2.8 kW depending on model. Voltage and frequency tolerances must not oppose, and space requirements around the unit must be respected.

What is not included in base delivery and what service provisions should be planned?

The unit ships with a cable but without a plug for sub-63 A three-phase supply, and it has no expansion tank, so the application must provide fluid filling and volume compensation. External Pt100 sensing is supported via the dedicated sensor connection; plan earthing and RCD selection around the external temperature-control leakage current.