Peter Huber Kältemaschinenbau KISS K6s Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau KISS K6s Kälte-Badumwälzthermostat — Thermostatic baths with Noise level 57 dB.
The KISS K6s is an air-cooled refrigerated bath circulator with a 4.5 L bath and a non-variable pressure/suction pump, covering -25 to 200 °C with ±0.05 K stability. It is the smallest refrigerated model in the KISS series up to 30 °C, distinguished by its compact 210 mm width and 0.05 kW cooling capacity at -20 °C. It serves as a temperature-control source for external circuits or direct bath immersion in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau KISS Series — Model Comparison & Selection
| Model | Noise level, dB | Power supply | Pump connection | Temperature range, °C | Weight, kg | Filling volume, m³ |
|---|---|---|---|---|---|---|
| KISS K6s Kälte-Badumwälzthermostat Current | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
| KISS K12 Kälte-Badthermostat | — | 110-115V 1~ 50/60Hz | (Option, #19607) | -20–200 | 28 | — |
| KISS K20 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 41 | 0.01 |
| KISS K25 Kälte-Badthermostat | 50 | 220-240V 1~/2~ 50/60Hz | (Option, #19607) | -30–200 | 39 | 0.015 |
| KISS K6 Kälte-Badumwälzthermostat | 57 | 208-240V 1~/2~ 50/60Hz | M16x1 AG male | -25–200 | 25 | 0.002 |
Full specifications
| Noise level | 57 dB(A) (sound pressure level) |
| Filling volume | 2 l (min) |
| Maximum delivery flow rate | 14 l/min (pressure pump max, 0, 25 bar) |
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | -25...200 °C |
| Temperature stability mfr | ±0, 05 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 210 x 400 x 546 mm |
| Heating capacity mfr | 1, 6 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min; 0, 17 bar |
| Bath opening WxD/BD mfr | 140 x 120 / 150 mm |
| Bath volume mfr | 4, 5 l |
| Weight | 25 kg (net) |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Current consumption max. mfr | 10 + 1, 3 A (max) |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C (permissible) |
| Protection rating | IP20 |
| Cooling capacity mfr | 20:0, 26; 0:0, 21; -10:0, 15; -20:0, 05 °C / kW (at 20/0/-10/-20 °C) |
| Safety class | III / FL |
Description
Construction and series positioning
- The KISS K6s is a refrigerated bath thermostat with a 4.5 L bath volume and a 140 x 120 mm opening (150 mm depth), housed in a 210 x 400 x 546 mm case weighing 25 kg net.
- Within the KISS series up to 30 °C, it is the smallest refrigerated variant; siblings K12, K15, K20 and K25 offer larger baths, and the K20/K25 are rated at 50 dB(A) versus 57 dB(A) for the K6s.
- Cooling is air-cooled with refrigerant R-290 (A3, H220), requiring flammable-refrigerant site precautions and adequate airflow clearance.
Thermal behavior and media
- Working range is -25 to 200 °C with ±0.05 K stability, displayed on an OLED screen and managed by the KISS controller.
- Cooling capacity falls steeply at low temperature: 0.26 kW at 20 °C, 0.21 kW at 0 °C, 0.15 kW at -10 °C and only 0.05 kW at -20 °C; the unit is therefore limited for high heat loads near the low end.
- Heating capacity is 1.6–2.1 kW. The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glysantine or ethylene glycol.
Pump, connection and control
- A pressure and suction pump (speed not variable) delivers up to 14 L/min at 0.25 bar pressure and 10.5 L/min at 0.17 bar suction, with M16x1 male pump connections (NW8/12).
- Included accessories comprise hose connector NW12 and sleeve nuts thread M16x1.
- Interfaces are USB (Device) and RS232; the controller provides an autostart function after power failure and optic/acoustic alarm messages.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876; protection rating IP20 and EN55011 Class A, Group 1 disturbance classification.
Integration & Operating Notes
The unit accepts 208–240 V 1~/2~ 50/60 Hz with a 10 A minimum and 16 A maximum fuse, but voltage and frequency tolerances are linked: ±5% voltage is permitted only with ±2% frequency tolerance, and combinations such as -5% voltage with +2% frequency are explicitly not allowed. Confirm the site supply against this matrix before commissioning.
Because the refrigerated bath is air-cooled, it requires adequate space around the unit per the manufacturer's space requirements; the IP20 rating means it is not protected against water ingress and must be kept dry. Ambient temperature must remain within 5 to 40 °C, and cooling capacity is rated at 20 °C room temperature and may drop as ambient rises.
The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer), so these fluids require a more resistant plastic option. For low-temperature operation, water should be mixed with glysantine or ethylene glycol for freeze protection to avoid freezing damage in the bath or circuit.
The controller includes an autostart function for power failure, meaning the unit may resume operation automatically when power is restored. Risk assessments and procedures for unattended operation must account for this behavior, alongside the adjustable overtemperature protection per DIN 12876 (safety class III / FL) and optic/acoustic alarms.
At the low end of the -25 to 200 °C range, available cooling is 0.05 kW at -20 °C, falling from 0.15 kW at -10 °C. Process heat loads near this temperature must be kept well below this figure, or the bath will not hold setpoint; the unit is not suitable for high heat loads near its minimum temperature.
The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1. In residential or EMC-sensitive laboratory settings, this classification may require additional consideration of adjacent sensitive instrumentation or shielding measures.
Selection & Integration Q&A
Choose the K6s for the smallest footprint (210 mm wide, 4.5 L bath) and lowest cooling demand; its 0.05 kW at -20 °C suits light loads only. Larger siblings K12/K15/K20/K25 provide more bath volume, and the K20/K25 are quieter at 50 dB(A) versus 57 dB(A).
The pump uses M16x1 male connections (NW8/12). Included accessories are hose connector NW12 and sleeve nuts thread M16x1; the pressure/suction pump delivers up to 14 L/min at 0.25 bar pressure and 10.5 L/min at 0.17 bar suction, with no speed adjustment.
Install within 5–40 °C ambient with adequate airflow for the air-cooled condenser, keep dry (IP20), and respect flammable-refrigerant R-290 requirements. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises; supply must be 208–240 V with 10–16 A fuse.
The base delivery includes hose connector NW12 and M16x1 sleeve nuts but not heat-transfer fluid; water should be mixed with glysantine or ethylene glycol for freeze protection. The plastic pump is incompatible with acetone and polyglycols unless a resistant plastic option is ordered.