BINDER ED 115
BINDER ED 115 — Drying ovens with Temperature 5…300 °C.
The BINDER ED 115 is a 114 L natural-convection drying and heating oven with a +5 to 300 °C working range, 1.25 kW single-phase heating, and 6 air changes per hour. Its 710 × 705 × 605 mm net chamber and 150 kg total load capacity (30 kg per rack) position it between the smaller ED 56 and larger ED 260 in the ED series. It is intended for tempering, drying, and heat treatment of non-flammable solids and bulk materials where forced convection is not required.
BINDER ED Series — Model Comparison & Selection
Full specifications
| Temperature | 5...300 °C |
| Air change rate | 6 x/h |
| Heating time to 150°C | 45 min |
| Interior volume | 114 L |
| Load per rack | 30 kg (ED115-230V / ED115UL-120V) |
| Nominal power | 1.25 kW |
| Permitted load | 150 kg (ED115-230V / ED115UL-120V) |
| Temperature fluctuation | 0.4 K |
| Temperature uniformity | 1.5 K |
| Rear wall clearance | 160 mm (ED115-230V / ED115UL-120V) |
| Side wall clearance | 100 mm (ED115-230V / ED115UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 150°C mfr | 16 min |
| Depth | 605 mm (ED115-230V / ED115UL-120V) |
| Height | 705 mm (ED115-230V / ED115UL-120V) |
| Width | 710 mm (ED115-230V / ED115UL-120V) |
| Net weight of the unit (empty) mfr | 54 kg |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The ED 115 uses natural convection (no fan), unlike the FD series, and is the mid-size single-door variant in the ED line; siblings ED 23, ED 56, ED 260, and ED 400 share the +5 to 300 °C range but differ in volume and load rating.
- Interior, pre-heating chamber, and door inner surfaces are stainless steel 1.4301 (AISI 304); housing protection is IP 20 per IEC 60529, pollution degree 2, over-voltage category II per IEC 61010-1.
- Temperature is displayed with 1 °C resolution (ED series), and the safety controller is class 2 per DIN 12880 with a 330 °C class 1 overtemperature trip.
Behavior under load and in operation
- Temperature fluctuation is 0.4 K and uniformity 1.5 K; recovery after a 30 s door opening at 150 °C is 16 min, and heat-up to 150 °C takes 45 min.
- The usable volume must not be loaded by more than half, and samples must keep at least 10 % of interior dimensions as wall clearance to maintain airflow and homogeneity.
- Above 150 °C, metallic surfaces may discolor due to oxidation; this is harmless and does not impair function.
Installation, connection, and control
- The standard ED115-230V variant (Art. 9010-0335) operates on 230 V 1~ 50/60 Hz with a CEE 7/7 plug and a 6.3 A external chamber fuse; the UL variant ED115UL-120V (Art. 9010-0336) uses 120 V 1~ 60 Hz with a NEMA 5-15 plug.
- The controller supports temperature set-point entry with optional ramp function, timer/continuous operation, and °C/°F selection; the safety controller set-point must be set 2–5 °C above the working set-point within the 10–300 °C range.
- Optional RS 422 interface enables APT-COM 4 software communication; an analog temperature output and a disconnectable audible over-temperature alarm are also available.
Integration & Operating Notes
The ED 115 draws 1.25 kW at 230 V 1~; fluctuations beyond ±10 % may affect control accuracy and safety-device operation. Confirm supply quality and use a 6.3 A time-lag external fuse (5×20 mm, 250 V) in the upstream circuit.
Exceeding half-loading or placing samples closer than 10 % of interior dimensions to walls prevents sufficient natural convection, causing temperature stratification and non-uniform results. Plan rack placement and batch size accordingly.
The safety controller acts as an absolute maximum limit; after any change to the temperature set-point, re-adjust the safety limit within its 10–300 °C range. Failure to do so may either trip the heating prematurely or leave the chamber without effective overtemperature protection.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, including repeat electrical tests per EN 50678/VDE 0701 and EN 50699/VDE 0702. Replace the door gasket only when the chamber is cold to avoid damage.
The ED series can be hot-air sterilized at 190 °C for at least 30 minutes; remove all flammable goods from the interior beforehand. Racks are removable and autoclavable; use neutral cleaning agent Art. 1002-0016 and avoid it on zinc-coated hinge parts and rear wall.
Selection & Integration Q&A
Choose the ED 115 when your batch fits within 114 L and requires up to 150 kg total load (30 kg per rack). The ED 56 offers less volume for smaller loads; the ED 260 provides more capacity but requires a larger footprint and higher nominal power (2.25 kW).
Annual maintenance by licensed electricians is required, including electrical safety tests per EN 50678/VDE 0701 and EN 50699/VDE 0702. Optional accessories include RS 422 interface with APT-COM 4 software, analog temperature output, and disconnectable audible over-temperature alarm. Neutral cleaning agent Art. 1002-0016 and disinfectant spray Art. 1002-0022 are recommended consumables.