No photo available yet

Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies GCxGC-based analyzers, aviation turbine fuels

Agilent Technologies GCxGC-based analyzers, aviation turbine fuels — Gas chromatography.

Flow-modulated GC×GC-FID analyzer built on the Agilent 8890 platform with a Reverse Flow Modulator, configured for group-type analysis of conventional and synthetic aviation turbine fuels per ASTM D8396 and NB/SH/T 6078-2023. It runs hydrogen carrier gas with a 7693 sampler, split/splitless inlet at 300 °C, dual-column flow modulation and FID detection at 300 °C over a 63.33 min runtime. It is the jet-fuel-specific member of the energy/chemical GC analyzer family, distinct from the biodiesel, RGA, gasoline/benzene and LPG siblings.

Agilent Technologies Energy Chemical GC Analyzers Series — Model Comparison & Selection

Model Typical quantitation limit, mol%Carrier gasAnalysis time, min
GCxGC-based analyzers, aviation turbine fuels Current —Hydrogen—
Analyzers for biodiesel, biofuels 0.01——
Gasoline analyzer, benzene analysis 0.01——
natural gas analyzer, fast analysis times 0.01—20
NGA Analyzer, natural gas analysis 0.01—20
Refinery gas analyzer, four channel multi-dimensional system —Argon—

Full specifications

Core Performance & Functional Specifications
Acquisition Rate 200 Hz
Air flow 350 mL/min
Carrier gas Hydrogen
Column 1 Flow 0.3 mL/min
Column 2 Flow 15 mL/min
Hydrogen Flow 35 mL/min
Monitor Column Flow Rate 0.63 mL/min
Primary Column Flow Rate 0.5 mL/min
Restrictor Column Flow 0.36 mL/min
Sample volume 0.1 µL
Secondary Column Flow Rate 25 mL/min
Septum Purge 6 mL/min
Sampler mfr 7693 Sampler
Detector mfr FID FID
Runtime mfr 63.33 min
Operating Conditions, Environment & Protection
Initial Oven Temperature 40 °C
Inlet Temperature 300 °C

Description

Construction and positioning in the series

Flow and thermal behavior

Operation

Integration & Operating Notes

Process Limits
Carrier gas choice fixes the method variant and all flows

The application note compares hydrogen and helium carrier configurations; the matched flows here (column 1 0.3 mL/min, column 2 15 mL/min, hydrogen 35 mL/min) belong to the hydrogen variant. Switching carrier without re-deriving flows and runtime invalidates the ASTM D8396 / NB/SH/T 6078-2023 result.

Application Note: Aviation Turbine Fuels ASTM D8396/5994/7929
Process Limits
Column set must match the selected method note

The flow-modulated jet fuel note specifies three distinct column sets (primary, secondary, monitor). The extra-spec column entries are truncated in the source, so the exact dimensions must be read from the note for the chosen set; mixing columns across sets changes modulation and retention behavior.

Application Note: Jet Fuel Flow-Modulated GC×GC-FID 5994-7775EN
Control & I/O
200 Hz acquisition needed to resolve modulated peaks

Flow modulation produces narrow second-dimension peaks; the method uses a 200 Hz acquisition rate. Detector electronics and data system must sustain this rate for the full 63.33 min runtime without dropping samples, otherwise peak assignment for group-type quantitation degrades.

Factory Specification Sheet

Selection & Integration Q&A

When should I choose this aviation turbine fuel GC×GC analyzer instead of the gasoline/benzene or LPG sibling analyzers?

Pick this one only for jet-fuel group-type analysis per ASTM D8396 or NB/SH/T 6078-2023, which requires the flow modulator and dual-column set. Gasoline/benzene and LPG siblings use different methods and column configurations; they will not produce the specified jet-fuel group-type result.

What carrier gas and column set must be fixed before ordering consumables for this configuration?

The matched configuration is hydrogen carrier (35 mL/min hydrogen, 350 mL/min air) with the RFM column set. The note also documents a helium variant and three column sets; confirm which set and carrier your method uses, because the extra-spec column entries are truncated and cannot be ordered from them directly.

What is not included in the base delivery that I must plan for?

The 7693 sampler, split/splitless inlet and FID are listed as configured hardware, but the exact column set (primary, secondary, monitor, restrictor) and carrier gas supply are method-dependent and must be specified separately. No certificates or accessory list were supplied, so confirm columns and gas plumbing with the vendor.

Dimensional Drawings & Technical Documentation

Dimensional drawings

performance curve performance curve performance curve

Click drawing to inspect vector dimensions

Technical documentation