Agilent Technologies natural gas analyzer, fast analysis times
Agilent Technologies ·Made in USA · In Active Production

Agilent Technologies natural gas analyzer, fast analysis times

Agilent Technologies natural gas analyzer, fast analysis times — Gas chromatography.

The Agilent natural gas analyzer is a factory-configured, chemically tested three-channel GC system built on the 8890 GC or 990 Micro GC platform, equipped with three valves, four columns, and TCD/FID detection. It delivers a 20-minute analysis with a typical quantitation limit of 0.01 mol%, covering H2, O2, N2, and extended hydrocarbons per UOP 1945. It is positioned for fast, method-compliant natural gas composition analysis within the Energy Chemical GC Analyzer series.

Agilent Technologies Energy Chemical GC Analyzers Series — Model Comparison & Selection

Model Typical quantitation limit, mol%Carrier gasAnalysis time, min
natural gas analyzer, fast analysis times Current 0.01—20
Analyzers for biodiesel, biofuels 0.01——
Gasoline analyzer, benzene analysis 0.01——
GCxGC-based analyzers, aviation turbine fuels —Hydrogen—
NGA Analyzer, natural gas analysis 0.01—20
Refinery gas analyzer, four channel multi-dimensional system —Argon—

Full specifications

Core Performance & Functional Specifications
Analysis time 20 minutes
Typical Quantitation Limit 0.01 mol%
Valves mfr 3
Detectors mfr TCD/FID TCD/FID
Columns mfr 4
Analyzer channels mfr Three-channel channel
Configured per published method mfr 1945 UOP
Analysis capacity for H2 mfr Yes Yes
Analysis capacity for O2 and N2 mfr Yes Yes
Extended hydrocarbon analysis mfr Yes Yes

Description

Construction and series positioning

Analytical behavior

Installation, control, and operation

Integration & Operating Notes

Process Limits
Sour gas service requires the dedicated sour gas configuration

The analyzer family is offered in standard, extended, and sour gas configurations. If the sample stream contains sour components, the standard configuration must not be used; the sour gas option must be ordered and the sample conditioning must match it. Confirm the stream composition before selecting the option.

Brochure 'br nga gc analyzer 5994 8260en agilent', Scope and Variants

Selection & Integration Q&A

When should I choose this fast natural gas analyzer instead of the refinery gas (RGA) or gasoline analyzer siblings?

Choose this model when the sample is natural gas and a 20-minute run with 0.01 mol% typical quantitation limit is required, including H2, O2, N2, and extended hydrocarbons per UOP 1945. Select the RGA sibling for refinery gas streams and the gasoline/benzene sibling for fuel analysis; those cover different matrices and methods.

What adjacent system parts must be specified with this analyzer, and what is not included?

The analyzer is factory-configured and chemically tested, but the GC gasifier accessory (G3535A) is a separate part and not included in base delivery. The platform (8890 GC or 990 Micro GC) is fixed by the ordered option, so site utilities and sample-handling hardware must be matched to that platform.

What sample-stream condition determines whether the standard configuration is acceptable?

The family includes standard, extended, and sour gas configurations. If the natural gas stream contains sour components, the standard configuration is not appropriate; the sour gas option must be ordered. Confirm stream composition before selecting the option, because the configuration is fixed at the factory.

What maintenance and accessory planning is required beyond the base analyzer?

Plan for the GC gasifier accessory (G3535A) as a separate order if condensable samples are involved; it is not in the base delivery. Because the unit is factory-configured and chemically tested, service and consumables should follow the platform actually supplied (8890 GC or 990 Micro GC) and the ordered option number.

Dimensional Drawings & Technical Documentation

Technical documentation