KB PRO (E7) 09 2025

Technical Reference Summary

Covered Models & Document Scope

KB PRO (E7) cooling incubators with program control, models KB PRO 130 (9020-0433, 9120-0433), KB PRO 260 (9020-0434, 9120-0434), KB PRO 470 (9020-0435, 9120-0435), KB PRO 720 (9020-0436, 9120-0436), KB PRO 1060 (9020-0437, 9120-0437), KB PRO 1600 (9020-0438, 9120-0438). Options covered include APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, and flexible Pt 100 object temperature sensor.

Section-by-Section Technical Breakdown

pp. 18-24
Chamber description and overview

Describes the KB PRO cooling incubator construction, instrument panel variants, control terminal, and rear connection panel. KB PRO 130/260 use a triangular instrument panel; KB PRO 470/720/1060/1600 use a signal triangle and ergonomically adjustable control terminal.

  • Triangular instrument panel on KB PRO 130 and 260
  • Signal triangle on KB PRO 470, 720, 1060, 1600
  • Ergonomically adjustable control terminal on KB PRO 470, 720, 1060, 1600
  • Connection panel located on chamber rear
pp. 28-31
Installation and connections

Covers wall spacer requirements from size 260, rack mounting with U-rails or telescopic rails, flexible tilt protection kit for KB PRO 130 and 470, condensate collection pan, and electrical connection.

  • Spacer for wall distance required from size 260 on
  • Racks available with U-rails or telescopic rails as accessory
  • Flexible tilt protection kit for KB PRO 130 and 470
  • Condensate collection pan installation
  • Electrical connection details
pp. 32-42
MB2 chamber controller functional overview

Describes the MB2 controller display views, icons, operating modes, menu structure, entry principles, and behavior during power failures and door opening.

  • Normal display, program display, and chart-recorder display views
  • Controller icons overview
  • Operating modes including fixed value, timer, time program, week program
  • Main menu, Settings submenu, Service submenu
  • Performance during and after power failures
  • Performance when opening the door
pp. 42
Start up

Covers turning on the chamber and controller settings performed upon start up.

  • Turning on the chamber
  • Controller settings upon start up
pp. 43-45
Set-point entry in fixed value operating mode

Describes entering temperature and fan speed setpoints through the Setpoints menu or directly via the Normal display, plus special controller functions.

  • Set-point entry for temperature and fan speed through Setpoints menu
  • Direct setpoint entry for temperature via Normal display
  • Special controller functions
pp. 46-47
Timer program: stopwatch function

Covers starting a timer program, behavior during program delay time, cancelling a running timer program, and performance after program end.

  • Starting a timer program
  • Performance during program delay time
  • Cancelling a running timer program
  • Performance after the end of the program
pp. 47-48
Hot-air disinfection

Covers preparations and the procedure for starting and ending hot-air disinfection.

  • Preparations for hot-air disinfection
  • Starting and ending the hot-air disinfection
pp. 48-60
Time programs

Describes starting, stopping, pausing, and cancelling time programs; creating and managing programs and sections; and entering section duration, setpoint ramps/steps, tolerance ranges, repeats, and saving.

  • Starting an existing time program
  • Stopping, pausing, and cancelling a running time program
  • Creating a new time program and program management
  • Section editor: add, copy, replace, delete sections
  • Value entry: section duration, setpoint ramp and step, tolerance ranges
  • Repeating sections within a time program
  • Saving the time program
pp. 61-69
Week programs

Describes starting and cancelling week programs, creating and managing week programs and sections, and entering weekday, start time, setpoint, and special controller functions.

  • Starting an existing week program
  • Cancelling a running week program
  • Creating a new week program and program management
  • Section editor: add, copy, replace, delete sections
  • Value entry: setpoint ramp/step, weekday, start time, setpoint
  • Special controller functions
pp. 70-73
Notification and alarm functions

Covers notification and alarm message overview, state of alarm, resetting alarms and active alarm list, tolerance range settings, and activating/deactivating the audible alarm.

  • Notifications and alarm messages overview
  • State of alarm
  • Resetting an alarm and list of active alarms
  • Tolerance range settings
  • Activating/deactivating the audible alarm (alarm buzzer)
pp. 73-79
Temperature safety devices

Describes over temperature protective device class 1, overtemperature safety controller class 2/class 3.3, safety controller modes and limit/offset settings, and safety controller class 3.2 undertemperature protection including alarm state and function check.

  • Over temperature protective device (class 1)
  • Overtemperature safety controller class 2 / class 3.3
  • Selecting between class 2 (temperature limiter) and class 3.3 (temperature protection)
  • Safety controller modes and setting the mode
  • Setting limit value for overtemperature and offset value for over-/undertemperature
  • Safety controller class 3.2 undertemperature protection
  • Message and measures in the state of alarm
  • Function check
pp. 79-88
User management

Covers authorization levels and password protection, login/logout, user change, password assignment and change, deleting passwords, new password assignment when password function was deactivated, and activation code.

  • Authorization levels and password protection
  • Log in and log out
  • User change
  • Password assignment and password change
  • Deleting the password for an individual authorization level
  • New password assignment for service or admin authorization level when password function was deactivated
  • Activation code
pp. 89-95
General controller settings

Covers selecting menu language, setting date and time, selecting temperature unit, display configuration and touchscreen calibration, network and communication (Ethernet, E-Mail), and USB data transfer.

  • Selecting the controller's menu language
  • Setting date and time
  • Selecting the temperature unit
  • Display configuration and touchscreen calibration
  • Network and communication: Ethernet, E-Mail
  • USB menu: data transfer via USB interface
pp. 96-98
General information

Covers service contact page, current operating parameters, event list, technical chamber information, and self-test function.

  • Service contact page
  • Current operating parameters
  • Event list
  • Technical chamber information
  • Self-test function
pp. 100-104
Chart recorder display

Describes chart recorder views, showing/hiding legend, switching legend pages, showing/hiding specific indications, history display, and setting parameters.

  • Show and hide legend
  • Switch between legend pages
  • Show and hide specific indications
  • History display
  • Setting the parameters
pp. 105
Defrosting at refrigerating operation

Covers defrosting procedures during refrigerating operation.

  • Defrosting at refrigerating operation
pp. 106
Anti-condensation protection

Covers anti-condensation protection measures.

  • Anti-condensation protection
pp. 106-111
Options and accessories

Describes available options: APT-COM 4 Multi Management Software, analog outputs for temperature, zero-voltage alarm contacts, internal LED light, door lock, and object temperature control with flexible Pt 100 sensor including activation, sensitivity, and maximum deviation settings.

  • APT-COM 4 Multi Management Software (accessory)
  • Analog outputs for temperature (option)
  • Zero-voltage alarm contacts for collective alarm output (option)
  • Internal LED light (option)
  • Door lock (option)
  • Object temperature control with flexible Pt 100 temperature sensor (option)
  • Activating/deactivating object temperature control
  • Setting sensitivity and maximum deviation of object temperature control
pp. 111-113
Cleaning and decontamination

Covers cleaning, decontamination/chemical disinfection, and hot-air disinfection procedures.

  • Cleaning
  • Decontamination / chemical disinfection
  • Hot-air disinfection
pp. 114-117
Maintenance and service, troubleshooting, repair, testing

Covers personnel qualification, maintenance intervals and service, simple troubleshooting, and sending the chamber back to BINDER. Includes annual maintenance requirement, condenser cleaning, door gasket replacement, and electrical testing requirements.

  • Maintenance at least once a year
  • Condenser cleaning by suction several times a year with dusty ambient air
  • Replace door gasket only when cold
  • Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702
  • After 8760 operating hours or two years a service message appears
  • Simple troubleshooting by operating personnel
  • Repair only by BINDER Service or qualified personnel
pp. 121-131
Technical description

Covers factory calibration and adjustment, over current protection, definition of usable volume, technical data, equipment and options, accessories and spare parts, and dimensions.

  • Factory calibration and adjustment
  • Over current protection
  • Definition of usable volume
  • Technical Data
  • Equipment and options (extract)
  • Accessories and spare parts (extract)
  • Dimensions

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal temperature 100 °C / 212 °F; IP protection 20; temperature safety device per DIN 12880:2007 class 2/3.3. p.10 ↗ Chamber is rated for a maximum nominal temperature of 100 °C and has a class 2/3.3 temperature safety device; selection must respect this limit. KB PRO 720 example type plate; series-wide per type plate data
Nominal power 1.70 kW, nominal current 7.1 A, nominal voltage 208-240 V at 50 Hz or 60 Hz, 1 N~. p.10 ↗ Electrical supply must be 208-240 V single-phase with adequate branch protection for 7.1 A; verify voltage/frequency before connection. KB PRO 720 example type plate
Refrigerant R290 filling weight 0.095 kg; maximum operating pressure 28 bar; contains hydrocarbon gases. p.10 ↗ Refrigeration work must be performed by qualified personnel trained per EN 13313:2010; observe flammable refrigerant safety labels and local regulations. KB PRO 720 example type plate
Maintenance work at the refrigeration system only by qualified personnel trained according to EN 13313:2010 (e.g. refrigeration technician with certified expert knowledge acc. to Regulation (EC) no. 303/2008). p.114 ↗ Service contracts and personnel selection must ensure refrigerant-certified technicians for any refrigeration system work. Whole series
Ensure regular maintenance work is performed at least once a year; all work on the refrigeration system must be documented. p.114 ↗ Plan annual preventive maintenance and maintain documentation for refrigeration system inspections and repairs. Whole series
With increased dust in ambient air, clean the condenser of the refrigeration machine by suction several times a year; condenser is behind the machine room cover, removable without tools by lifting upwards and pulling forwards. p.115 ↗ In dusty environments, schedule more frequent condenser cleaning to maintain refrigeration performance. Whole series
Replace the door gasket only when cold; otherwise the door gasket may become damaged. p.115 ↗ Door gasket replacement must be done with the chamber at ambient/cold condition. Whole series
After 8760 operating hours or two years, a service message appears; after confirmation with the Confirm icon, the message window pops up again every two weeks until reset by BINDER Service. p.115 ↗ Plan BINDER service reset at or before 8760 operating hours / two years to avoid recurring service prompts. Whole series
Annual repeat inspection and a test prior to initial startup and prior to resuming operation after maintenance or repair are required; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.114 ↗ Include electrical safety testing in commissioning and annual maintenance schedules. Whole series
Do NOT install the chamber in unventilated recesses; ensure sufficient ventilation for heat dispersal and observe prescribed minimum distances (chap. 3.4). p.12 ↗ Installation location must provide ventilation and wall clearances to prevent overheating damage. Whole series
Spacer for wall distance required from size 260 on. p.28 ↗ KB PRO 260 and larger models require the wall spacer to maintain specified wall clearance. KB PRO 260, 470, 720, 1060, 1600
Flexible tilt protection kit is mounted on KB PRO 130 and 470. p.29 ↗ These models require installation of the tilt protection kit for stability. KB PRO 130 and KB PRO 470
Racks are available with U-rails or with telescopic rails as accessory. p.28 ↗ Select rack type according to loading and access requirements; telescopic rails are an accessory option. Whole series
Temperature safety device class 2/3.3 per DIN 12880:2007; safety controller can be selected as class 2 (temperature limiter) or class 3.3 (temperature protection); class 3.2 undertemperature protection also available. p.73 ↗ Configure safety controller class and mode according to process risk; class 3.2 provides undertemperature protection. Whole series
Object temperature control option uses a flexible Pt 100 temperature sensor with adjustable sensitivity and maximum deviation. p.109 ↗ For object-temperature-critical processes, specify the Pt 100 option and configure sensitivity and maximum deviation. Option, whole series
Zero-voltage alarm contacts for collective alarm output are available as option. p.107 ↗ For remote alarm monitoring, order the zero-voltage alarm contact option. Option, whole series
Analog outputs for temperature are available as option. p.107 ↗ For external temperature recording, specify analog output option. Option, whole series
Internal LED light and door lock are available as options. p.108 ↗ Specify internal LED light for visibility and door lock for access control. Option, whole series
APT-COM 4 Multi Management Software is available as accessory, including Basic Edition. p.106 ↗ For computer communication and multi-unit management, specify APT-COM 4 software. Whole series
Controller supports Ethernet and E-Mail communication settings and USB data transfer. p.93 ↗ Network and USB interfaces can be configured for data logging and remote notification. Whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Service message after 8760 operating hours or two years p.115 ↗ Scheduled service interval reached (8760 operating hours or two years) Confirm with the Confirm icon; message reappears every two weeks until reset by BINDER Service Section 22.2