BINDER KB PRO 260 — Incubators with Temperature -20…100 °C.

The KB PRO 260 is a 262 L single-phase cooling incubator with a -20 to 100 °C working range, 0.1 K fluctuation and 0.2 K uniformity, positioned mid-series between the 130 and 470. Its 800 W heat compensation and 200 kg permitted load support high-dissipation or dense loads, while the 40 kg per-rack limit defines shelf loading.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 260 Current -20–10090139.626265640
KB PRO 1060 -20–100117.5195.11070101140
KB PRO 130 -20–10071100.512765630
KB PRO 1600 -20–100117.5195.11610152140
KB PRO 470 -20–10090195.147265640
KB PRO 720 -20–10090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 262 L
Load per rack 40 kg
Max. heat compensation 800 W
Nominal power 1.5 kW
Permitted load 200 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 180 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 3 min
Dimensions, Weight & Mechanical Design
Depth 900 mm
Height 1396 mm
Interior width 656 mm
Width 930 mm
Net weight of the unit (empty) mfr 190 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in different media

Installation, connection and control

Integration & Operating Notes

Electrical
Provide a 16 A branch circuit for the 1.5 kW single-phase load

The unit is rated 208...240 V, 1~, 50 Hz with 1.5 kW nominal power and a 16.0 A unit fuse. The upstream branch protection and wiring must be sized for this current, and the supply voltage and frequency must be verified before connection; the standard version is fitted with a CEE 7/7 plug.

Factory Specification Sheet and Data Sheet KB PRO 260
Process Limits
Respect 40 kg per rack and 200 kg total permitted load

Rack loading is limited to 40 kg per rack and the total permitted load is 200 kg. Distribute heavy or dense samples accordingly; exceeding the per-rack limit risks deformation of the rack and obstruction of airflow, which would degrade the specified 0.2 K uniformity.

Factory Specification Sheet
Cooling & Media
Refrigeration work only by EN 13313:2010-certified technicians

The refrigerant circuit contains a hydrocarbon refrigerant and operates at up to 28 bar. Any work on the refrigeration system must be performed by personnel trained to EN 13313:2010 and documented; door gasket replacement must be done only with the chamber cold to avoid damaging the gasket.

Instruction Manual KB PRO (E7), Section 22.2

Selection & Integration Q&A

When should I choose the KB PRO 260 rather than the 130 or 470?

Pick the 260 when 262 L interior and 200 kg load capacity are needed but floor space favors a 930 mm width; the 130 is smaller, while the 470 and above add the adjustable control terminal and larger volume. The 260 uses the triangular panel and requires the wall spacer.

What supply and protection does the KB PRO 260 require at the installation point?

A single-phase 208...240 V, 50 Hz supply with a 16 A branch circuit, matching the 1.5 kW nominal power and 16.0 A unit fuse; the standard version is fitted with a CEE 7/7 plug. Verify voltage and frequency before connection.

What mounting clearances and environment does the manual require for this model?

Install with the wall spacer, at least 100 mm rear and 180 mm side clearance, and never in an unventilated recess. In dusty air, clean the condenser several times a year to preserve refrigeration performance.

What maintenance and accessories are not included in the base delivery?

Annual maintenance and the 8760 h/two-year BINDER service reset are not included. Options to order separately: APT-COM 4 software, analog outputs, zero-voltage alarm contacts, internal LED light, door lock, Pt 100 object sensor, and U-rail or telescopic racks.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB PRO 260 en
Datasheet ·587,9 КБ ·8 pp.
View PDF
PDF
KB PRO (E7) 09 2025 en
Manual ·12,7 МБ ·143 pp.
View PDF