BINDER KB PRO 130

BINDER KB PRO 130 — Incubators with Temperature -20…100 °C.

The KB PRO 130 is a 127 L single-phase cooling incubator with a −20…100 °C working range, 0.1 K fluctuation and 0.2 K uniformity, housed in a 900 mm wide cabinet. It is the smallest KB PRO model and the only one that uses a triangular instrument panel and requires a flexible tilt protection kit instead of a wall spacer. It is intended for temperature-controlled incubation and storage where precise, reproducible thermal conditions are required.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 130 Current -20–10071100.512765630
KB PRO 1060 -20–100117.5195.11070101140
KB PRO 1600 -20–100117.5195.11610152140
KB PRO 260 -20–10090139.626265640
KB PRO 470 -20–10090195.147265640
KB PRO 720 -20–10090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 127 L
Load per rack 30 kg
Max. heat compensation 450 W
Nominal power 1 kW
Permitted load 150 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 100 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 2 min
Dimensions, Weight & Mechanical Design
Depth 710 mm
Height 1005 mm
Interior width 656 mm
Width 900 mm
Net weight of the unit (empty) mfr 135 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior and load

Installation, control and operation

Integration & Operating Notes

Mounting
Tilt protection kit must be fitted before operation

The KB PRO 130 requires the flexible tilt protection kit to be mounted; operating without it leaves the cabinet without the stability safeguard specified for this model. The kit is supplied separately and must be installed during commissioning.

Instruction Manual KB PRO (E7), Section 4.3
Mounting
Maintain 100 mm clearance at rear and sides

A minimum wall clearance of 100 mm at the back and sides is required for heat dispersal. Do not install the chamber in unventilated recesses; insufficient clearance can cause overheating and invalidate thermal performance.

Instruction Manual KB PRO (E7), Section 1.8 and Section 3.4
Electrical
Branch circuit must be rated for 16 A fuse

The unit is protected by a 16.0 A fuse and draws 1 kW nominal at 208…240 V single-phase. The supply circuit and upstream protection must be sized accordingly; verify voltage and frequency (50 Hz) before connection.

Data Sheet Model KB PRO 130; Instruction Manual KB PRO (E7), Section 1.6
Control & I/O
Alarm contacts and analog outputs are optional

Zero-voltage alarm contacts and analog temperature outputs are not part of the standard delivery. If remote alarm monitoring or external recording is required, these options must be specified at order; the controller otherwise provides Ethernet and USB interfaces for data transfer.

Instruction Manual KB PRO (E7), Sections 20.2 and 20.3
Process Limits
Do not exceed 150 kg total load or 30 kg per rack

The permitted load is 150 kg for the chamber and 30 kg per rack. Exceeding these limits can damage the rack supports or affect temperature uniformity; distribute loads evenly across racks.

Factory Specification Sheet; Instruction Manual KB PRO (E7), Technical Data
Maintenance
Annual maintenance and electrical safety test required

Regular maintenance must be performed at least once a year, and an electrical safety test per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial startup and after any repair. A service message appears after 8760 operating hours or two years and must be reset by BINDER Service.

Instruction Manual KB PRO (E7), Sections 22.1 and 22.2
Maintenance
Clean condenser more often in dusty environments

With increased dust in ambient air, clean the condenser several times a year by suction. The condenser is behind the machine room cover, which can be removed without tools by lifting upwards and pulling forwards. Refrigeration system work must be performed by personnel trained to EN 13313:2010.

Instruction Manual KB PRO (E7), Section 22.2

Selection & Integration Q&A

When should I choose the KB PRO 130 instead of a larger KB PRO model?

Choose the 130 when 127 L interior volume and a 150 kg load capacity suit your workload and you want the compact 900 mm wide cabinet. Larger models (260 and up) add volume but require a wall spacer; the 130 instead uses a tilt protection kit.

What maintenance and accessories are not included in the base delivery?

Annual maintenance and electrical safety testing are required but not included. Racks with U-rails or telescopic rails, APT-COM 4 software, analog outputs, alarm contacts, LED light, door lock and Pt 100 sensor are options/accessories. A service message appears after 8760 h or two years.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB PRO 130 en
Datasheet ·470,2 КБ ·7 pp.
View PDF
PDF
KB PRO (E7) 09 2025 en
Manual ·12,7 МБ ·143 pp.
View PDF