BINDER KB PRO 470 — Incubators with Temperature -20…100 °C.

The KB PRO 470 is a 472 L cooling incubator with a -20 to +100 °C working range, part of the KB PRO series but distinguished by its 470-size cabinet, signal-triangle control terminal, and flexible tilt protection kit. It is intended for temperature-controlled incubation and storage where 0.1 K fluctuation and 0.2 K uniformity are required, and it supports rack loading up to 40 kg per rack and 200 kg total.

Temperature
-20…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB PRO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB PRO 470 Current -20–10090195.147265640
KB PRO 1060 -20–100117.5195.11070101140
KB PRO 130 -20–10071100.512765630
KB PRO 1600 -20–100117.5195.11610152140
KB PRO 260 -20–10090139.626265640
KB PRO 720 -20–10090195.1727101140

Full specifications

Core Performance & Functional Specifications
Temperature -20...100 °C
Interior volume 472 L
Load per rack 40 kg
Max. heat compensation 800 W
Nominal power 1.5 kW
Permitted load 200 kg
Temperature fluctuation 0.1 K
Temperature uniformity 0.2 K
Back wall clearance 100 mm
Side wall clearance 180 mm
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 4 min
Dimensions, Weight & Mechanical Design
Depth 900 mm
Height 1951 mm
Interior width 656 mm
Width 930 mm
Net weight of the unit (empty) mfr 233 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 208...240 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behavior and load

Control, installation and operation

Integration & Operating Notes

Mounting
Wall spacer and tilt protection kit are mandatory for the 470

The KB PRO 470 requires the wall spacer (from size 260 upward) to maintain 100 mm rear and 180 mm side clearance, and it must be fitted with the flexible tilt protection kit. Installing without these risks overheating from insufficient heat dispersal and loss of stability; the chamber must not be placed in unventilated recesses.

Instruction Manual KB PRO (E7), Sections 1.8, 4.1 and 4.3
Electrical
Size branch protection for the 16.0 A unit fuse at 208–240 V

The unit is single-phase 1 N~, 208–240 V, 50 Hz, with a 16.0 A unit fuse and 1.5 kW nominal power. The supply circuit and upstream protection must be rated for the unit fuse current, and voltage/frequency must be verified before connection; the manual requires electrical safety testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after maintenance.

Factory Specification Sheet; Instruction Manual KB PRO (E7), Section 22.1
Process Limits
Keep self-heating load below the 800 W heat-compensation ceiling

Maximum heat compensation is 800 W, so the combined heat input from samples and any internal equipment must stay within this budget to hold setpoint across the -20 to +100 °C range. Rack loading is separately limited to 40 kg per rack and 200 kg total permitted load; exceeding either limit compromises temperature uniformity and mechanical safety.

Factory Specification Sheet
Control & I/O
Configure the safety controller class to match process risk

The temperature safety device is class 2/3.3 per DIN 12880:2007; the safety controller can be set as class 2 (temperature limiter) or class 3.3 (temperature protection), with class 3.2 undertemperature protection available. The selected class and its limit/offset settings determine the alarm state and function check, so they must be defined before the chamber is released to the process.

Instruction Manual KB PRO (E7), Sections 13.1–13.3
Maintenance
Plan annual service and BINDER reset before 8760 operating hours

A service message appears after 8760 operating hours or two years and repeats every two weeks until reset by BINDER Service. Annual maintenance is required, and all refrigeration work must be documented and performed by technicians trained per EN 13313:2010. In dusty ambient air, clean the condenser several times a year; replace the door gasket only when cold.

Instruction Manual KB PRO (E7), Section 22.2

Selection & Integration Q&A

When should I choose the KB PRO 470 rather than the 260 or 720?

Pick the 470 for a 472 L working volume with 656 mm interior width and 900 mm depth. The 260 is smaller; the 720 and larger use the same signal-triangle control terminal but add volume. The 470 alone shares the flexible tilt protection kit with the 130, so plan for that kit and the wall spacer required from size 260 upward.

What supply and protection does the KB PRO 470 need, and what alarm/recording options must be ordered separately?

It needs single-phase 208–240 V, 50 Hz, with a 16.0 A unit fuse and 1.5 kW nominal power. Zero-voltage alarm contacts, analog temperature outputs, internal LED light, door lock, Pt 100 object sensor and APT-COM 4 software are options, not base delivery; specify them if remote alarm or recording is required.

What maintenance and consumables are not included with the base unit?

Annual maintenance is required, with a service message at 8760 operating hours or two years that only BINDER Service can reset. Refrigeration work needs EN 13313:2010-trained personnel. Condenser cleaning is needed several times a year in dusty air; door gaskets, racks (U-rail or telescopic), APT-COM 4 and interface options are accessories, not base delivery.