KBF S ECO (E6) 07 2026
Technical Reference Summary
Covered Models & Document Scope
This operating manual (E6, 07/2026) covers the constant climate chambers KBF-S ECO and KBF-S ECO-UL with RD4 controller and Peltier technology. Models: KBF-S ECO 240 (KBFSECO240-230V, Art. 9020-0416/9120-0416), KBF-S ECO 240-UL (KBFSECO240UL-120V, Art. 9020-0420/9120-0420), KBF-S ECO 400 (KBFSECO400-230V, Art. 9020-0463/9120-0463), KBF-S ECO 400-UL (KBFSECO400UL-120V, Art. 9020-0464/9120-0464), KBF-S ECO 720 (KBFSECO720-230V, Art. 9020-0418/9120-0418), KBF-S ECO 720-UL (KBFSECO720UL-120V, Art. 9020-0421/9120-0421), KBF-S ECO 1020 (KBFSECO1020-230V, Art. 9020-0419/9120-0419), KBF-S ECO 1020-UL (KBFSECO1020UL-120V, Art. 9020-0422/9120-0422).
Section-by-Section Technical Breakdown
The KBF-S ECO / KBF-S ECO-UL constant climate chambers are equipped with a multifunctional microprocessor controller (RD4) and Peltier thermoelectric cooling. They meet the requirements for 'Long term testing' and 'Accelerated testing' and are designed for maximum usability, including continuous operation. APT-COM™ 4 Multi Management Software permits networking up to 100 chambers.
- Multifunctional microprocessor controller RD4
- Peltier (thermoelectric) heating and cooling technology
- Suitable for long-term and accelerated testing
- Networking up to 100 chambers via APT-COM™ 4
This section covers spacers for wall distance, mounting the flexible tilt protection kit for size 400, water supply (suitable water quality, BINDER Pure Aqua Service accessory, freshwater supply installation), condensate collection pan, and electrical connection. It specifies ambient conditions and installation site requirements.
- Spacers for wall distance
- Flexible tilt protection kit for chambers size 400
- Water supply: suitable water quality, BINDER Pure Aqua Service (accessory), freshwater supply installation
- Condensate collection pan
- Electrical connection
Describes the menu structure of the RD4 controller and access levels, and the controller performance during and after power failures.
- Menu structure and access levels
- Performance during and after power failures
Procedures for starting up the chamber.
- Start-up procedure
Instructions for entering temperature and humidity set-points on the RD4 controller.
- Temperature set-point entry
- Humidity set-point entry
Special controller functions including turning off the humidity system and idle mode.
- Turning off the humidity system
- Idle mode
Password request and assignment/modification of User and Admin passwords.
- Password request
- Assign and modify the User password
- Assign and modify the Admin password
Covers the over temperature protective device (class 1) and the overtemperature safety controller class 3.1, including setting the safety controller mode and value, messages and measures in the state of alarm, and function check.
- Over temperature protective device (class 1)
- Overtemperature safety controller class 3.1
- Setting the safety controller mode
- Setting the safety controller value
- Message and measures in the state of alarm
- Function check
Settings for menu language, temperature unit, date, time, language selection at restart, chamber address, and display brightness.
- Selecting the controller’s menu language
- Selecting the temperature unit
- Setting the current date
- Setting the current time
- Function “Language selection at restart”
- Setting the chamber address
- Display brightness
Setting the delay time for tolerance range alarm, temperature tolerance range, and humidity tolerance range.
- Setting the delay time for tolerance range alarm
- Setting the temperature tolerance range
- Setting the humidity tolerance range
Information messages, alarm messages, and activating/deactivating the audible alarm (alarm buzzer).
- Information messages
- Alarm messages
- Activating / deactivating the audible alarm (alarm buzzer)
Start time program, start week program, stop program, and performance after power failures.
- Start time program
- Start week program
- Stop program
- Performance after power failures
Showing network settings (MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name) and changing the configuration (IP address assignment, DNS server address assignment, IP address, subnet mask, standard gateway, DNS server address).
- Showing the network settings
- Showing the chamber’s MAC address
- Showing the IP address
- Showing the subnet mask
- Showing the standard gateway
- Showing the DNS server address
- Showing the DNS chamber name
- Changing the configuration of the network settings
- Selecting the type of IP address assignment (automatic / manual)
- Selecting the type of assignment of the DNS server address (automatic / manual)
- Assigning the IP address
- Setting the subnet mask
- Setting the standard gateway
- Assigning the DNS server address
Recorded data, storage capacity, setting the storage rate for the “DL1” recorder data, and deleting the data recorder.
- Recorded data
- Storage capacity
- Setting the storage rate for the “DL1” recorder data
- Deleting the data recorder
Connecting the USB stick, import function, export functions, ongoing data transfer, error during data transmission, and removing the USB stick.
- Connecting the USB stick
- Import function
- Export functions
- Ongoing data transfer
- Error during data transmission
- Removing the USB stick
Activating and deactivating the self-test function.
- Activating the self-test function
- Deactivating the self-test mode
Function of the humidifying and dehumidifying system.
- Function of the humidifying and dehumidifying system
Defrosting procedures during refrigerating operation.
- Defrosting at refrigerating operation
Anti-condensation protection for models KBF-S ECO 400, 720, and 1020.
- Anti-condensation protection for sizes 400, 720, 1020
APT-COM™ 4 Multi Management Software (accessory).
- APT-COM™ 4 Multi Management Software (accessory)
Cleaning and decontamination / chemical disinfection procedures.
- Cleaning
- Decontamination / chemical disinfection
General information and personnel qualification, maintenance intervals and service, service reminder, simple troubleshooting, and sending the chamber back to BINDER GmbH. Includes electrical testing requirements and maintenance intervals.
- General information, personnel qualification
- Maintenance intervals, service
- Service Reminder
- Simple troubleshooting
- Sending the chamber back to BINDER GmbH
- Annual repeat inspection according to EN 50678/VDE 0701 and EN 50699/VDE 0702 recommended
- Regular maintenance work at least once a year
- Maintenance of steam humidifier at least once a year
Factory calibration and adjustment, over current protection, definition of usable volume, technical data, equipment/options/accessories, accessories and spare parts, and dimensions.
- Factory calibration and adjustment
- Over current protection
- Definition of usable volume
- KBF-S ECO / KBF-S ECO-UL technical data
- Equipment, options and accessories (extract)
- Accessories and spare parts (extract)
- Dimensions
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal temperature: 70 °C / 158 °F. p.9 ↗ | Maximum operating temperature for all models. | KBF-S ECO / KBF-S ECO-UL series |
| IP protection: 20 according to EN 60529. p.9 ↗ | Chamber is protected against solid objects >12.5 mm but not against water; must not become wet. | KBF-S ECO / KBF-S ECO-UL series |
| Temperature safety device: DIN 12880, class 3.1. p.9 ↗ | Overtemperature safety controller class 3.1 must be set and checked according to chapter 10.2. | KBF-S ECO / KBF-S ECO-UL series |
| KBF-S ECO 240 (230 V): nominal power 0.80 kW, nominal current 3.5 A, nominal voltage 200-230 V / 50 Hz or 208-230 V / 60 Hz, 1 N~. p.9 ↗ | Electrical supply must match these ratings; wrong voltage can cause malfunction. | KBF-S ECO 240 (9020-0416, 9120-0416) |
| KBF-S ECO 240-UL: nominal power 0.80 kW, nominal current 6.7 A, nominal voltage 100-120 V / 50 Hz or 60 Hz, 1 N~. p.9 ↗ | UL model requires 100-120 V supply; higher current than 230 V version. | KBF-S ECO 240-UL (9020-0420, 9120-0420) |
| Ambient temperature above 25 °C / 77 °F reduces refrigerating performance. p.72 ↗ | Install chamber in a location where ambient temperature does not exceed 25 °C / 77 °F for full refrigerating performance. | KBF-S ECO / KBF-S ECO-UL series |
| Regular maintenance work must be performed at least once a year. p.70 ↗ | Plan annual maintenance; warranty void if maintenance conducted by non-authorized personnel. | KBF-S ECO / KBF-S ECO-UL series |
| Have regular maintenance work on the steam humidifier conducted at least once a year. p.70 ↗ | Humidifier maintenance interval depends on water quality and amount of steam produced. | KBF-S ECO / KBF-S ECO-UL series |
| Annual repeat inspection and test prior to initial startup and prior to resuming operation after maintenance or repair are required; recommended under EN 50678/VDE 0701 and EN 50699/VDE 0702. p.70 ↗ | Electrical safety testing must be performed by qualified personnel according to these standards. | KBF-S ECO / KBF-S ECO-UL series |
| APT-COM™ 4 Multi Management Software permits networking up to 100 chambers. p.18 ↗ | For networked operation, up to 100 chambers can be connected via APT-COM™ 4. | KBF-S ECO / KBF-S ECO-UL series |
| Humidity sensor resistance: maximum work place threshold limit values and tolerated concentrations with permanent load are specified for various substances (e.g., Ammonia NH3: 20 ppm / 14 mg/m³ TLV, 5500 ppm / 4000 mg/m³ permanent load; Acetone: 500 ppm / 1200 mg/m³ TLV, 3300 ppm / 8000 mg/m³ permanent load; Formaldehyde: 0.3 ppm / 0.37 mg/m³ TLV, 2400 ppm / 3000 mg/m³ permanent load). p.17 ↗ | Ensure that substances in the chamber do not exceed the specified concentrations to avoid damaging the humidity sensor. | KBF-S ECO / KBF-S ECO-UL series |
| Tolerated error of humidity sensor: +/- 2 % r.h. p.17 ↗ | Humidity measurement accuracy is within ±2 % relative humidity. | KBF-S ECO / KBF-S ECO-UL series |
| Chamber fuse has responded: check chamber fuse and replace if appropriate; if it responds again, contact BINDER service. p.72 ↗ | A blown fuse may indicate a fault; replacement should be done with correct rating, and repeated blowing requires service. | KBF-S ECO / KBF-S ECO-UL series |
| Nominal temperature exceeded by 10 °C due to chamber failure causes over temperature protective device (class 1) to respond. p.72 ↗ | The class 1 protective device trips at 10 °C above nominal temperature to prevent overheating. | KBF-S ECO / KBF-S ECO-UL series |
| Safety controller set too low can cause chamber not to heat up when turned on; acknowledge alarm and select suitable safety controller value (chap. 10.2). p.72 ↗ | Safety controller value must be set above normal operating temperature to avoid nuisance trips. | KBF-S ECO / KBF-S ECO-UL series |
| Combination of temperature/humidity values not in the optimum range (see temperature humidity diagram, Figure 19) can cause low or no refrigerating performance. p.72 ↗ | Operate within the optimum temperature/humidity range shown in the diagram to ensure refrigerating performance. | KBF-S ECO / KBF-S ECO-UL series |
| Too much external heat load can cause low or no refrigerating performance. p.72 ↗ | Reduce heat load from charging material or environment to maintain refrigerating performance. | KBF-S ECO / KBF-S ECO-UL series |
| Controller not well adjusted, or adjustment interval exceeded, can cause chamber heating permanently and set-point not maintained. p.72 ↗ | Calibrate and adjust controller periodically to maintain temperature control. | KBF-S ECO / KBF-S ECO-UL series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.72 ↗ | No power supply. | Check connection to power supply. Check whether the chamber is turned on at the main power switch. | |
| Chamber without function. p.72 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.5). | |
| Chamber without function. p.72 ↗ | Chamber fuse has responded. | Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. | |
| Chamber without function. p.72 ↗ | Controller defective. | Contact BINDER service. | |
| Nominal temperature exceeded by 10°C due to chamber failure. p.72 ↗ | Over temperature protective device (class 1) responds. | ||
| Chamber heating permanently, set-point not maintained. p.72 ↗ | Controller defective. | Contact BINDER service. | |
| Chamber heating permanently, set-point not maintained. p.72 ↗ | Controller not well adjusted, or adjustment interval exceeded. | Calibrate and adjust controller. | |
| Chamber doesn’t heat up. p.72 ↗ | Pt 100 sensor defective. | Contact BINDER service. | |
| Chamber doesn’t heat up. p.72 ↗ | Peltier module for heating/refrigeration defective. | ||
| Chamber doesn’t heat up. p.72 ↗ | Semiconductor relay defective | ||
| Chamber doesn’t heat up when turned on. p.72 ↗ | Safety controller responds. Inner chamber temperature has reached the safety controller value. Safety controller set too low. | Acknowledge the alarm on the controller. Check temperature setpoint setting. If appropriate, select suitable safety controller value (chap. 10.2). | |
| Chamber doesn’t heat up when turned on. p.72 ↗ | Safety controller (chap. 10.2) defective. | Contact BINDER service. | |
| Low or no refrigerating performance. p.72 ↗ | Ambient temperature above 25 °C / 77 °F (chap.3.4). | Select cooler place of installation. | |
| Low or no refrigerating performance. p.72 ↗ | Combination of temperature/humidity values not in the optimum range (see temperature humidity diagram, Figure 19). | Select combination of temperature/humidity values in the optimum range (chap. 19). | |
| Low or no refrigerating performance. p.72 ↗ | Peltier module for heating/refrigeration defective. | Contact BINDER service. | |
| Low or no refrigerating performance. p.72 ↗ | Too much external heat load. | Reduce heat load. |