Agilent Technologies 571 Bayard-Alpert Gauge Tube
Agilent Technologies 571 Bayard-Alpert Gauge Tube — Vacuum technologies with Max pressure 1.33×10⁻⁴ megapascal.
The 571 is a Bayard-Alpert ionization gauge tube with a yttria-coated iridium filament, covering 1×10⁻³ to 2×10⁻¹⁰ Torr. It is available in glass, Kovar, CF, and KF flange variants, all sharing the same electrical ratings. It serves as the high-vacuum sensor in systems where a controller provides emission, degassing, and ground integrity.
Agilent Technologies Vacuum Transducers Series — Model Comparison & Selection
| Model | Measurement range | Atomic mass unit, grams | Electron charge, coulomb | Molar volume, liters | Max pressure, megapascal | Sensitivity |
|---|---|---|---|---|---|---|
| 571 Bayard-Alpert Gauge Tube Current | -3...2 | 1.66 | 1602 | 22.41 | 1.33×10⁻⁴ | 17 |
| 531 Thermocouple Gauge, Rough Vacuum Pressure Measurement | — | 1.66 | 1602 | 22.41 | — | — |
| 572 Tungsten Bayard-Alpert Gauge | -3...2 | 1.66 | 1602 | 22.41 | -3...2 | — |
| Cold Cathode Gauge Tube 525, High Vacuum (HV) Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| Hot Filament Gauge MBA2-201, Ionization BA Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-100 Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| IMG-300 UHV Inverted Magnetron Gauge | — | 1.66 | 1602 | 22.41 | — | — |
| UHV-24/UHV-24p | 1 | 1.66 | 1602 | 22.41 | — | 25 |
Full specifications
| Max pressure | 1 Torr |
| Measurement range | -3...2 Torr |
| Sensitivity | 17 |
| Degassing | 21 |
| Electron Charge | 1.602 coulomb |
| Internal Volume | 220 cc |
| Maximum operating pressure | 1 Torr |
| Molar Volume | 22.41 liters |
| Operation | 17 |
| Pumping Speed, Ionic | 0.06 liters/sec |
| X-ray Limit | 2 Torr |
| Acceleration due to gravity mfr | 9.8066 |
| Angstrom unit mfr | 10 m |
| Avogadro's number mfr | 6.02214076 |
| Boltzmann's constant mfr | 1.38064852 |
| Equivalent of heat mfr | 4.185 |
| Natural log base mfr | 2.7183 |
| Velocity of light mfr | 2.9979 |
| Velocity of sound mfr | 343 |
| Magnetic flux density mfr | 1 |
| Pumping speed conversion mfr | 1 L/s |
| Throughput conversion mfr | 1 cc/sec |
| Outgassing rate mfr | 2 |
| Envelope mfr | Nonex Nonex |
| Grid mfr | 0.025 diameter |
| Collector mfr | 0.010 diameter |
| Base Leads mfr | 0.060 diameter |
| Collector Lead mfr | 0.040 diameter |
| Shield mfr | Internally connected to filament filament |
| Electron Bombardment mfr | 50 maximum |
| Gas Sensitivity (Helium) mfr | 1.5 Torr |
| Gas Sensitivity (Nitrogen) mfr | 10 Torr |
| Gas Sensitivity (Air) mfr | 10 Torr |
| Gas Sensitivity Constant S mfr | 10 Torr |
| Description of the 571 Ionization Gauge mfr | 15 |
| Electrical Connections mfr | 16 |
| Technical Specifications mfr | 18 |
| Use of The Gauge Tube mfr | 21 |
| Gas Correction Factors mfr | 23 |
| Accessories and spare parts mfr | 25 |
| Helium mfr | 1.5 Torr |
| Nitrogen mfr | 10 Torr |
| Air mfr | 10 Torr |
| L7430301 mfr | Glass ion gauge connector kit kit |
| R32443010 mfr | -600 ft |
| R32443025 mfr | -600 ft |
| R32443050 mfr | -600 ft |
| NW25 KF mfr | 40.1 |
| NW40 KF mfr | 55.1 |
| base pressure (ion/turbomolecular pumps, quantum) mfr | 10...9 mbar |
| base pressure (ion/turbomolecular pumps, accelerator) mfr | 10...12 mbar |
| system pressure range (accelerator UHV) mfr | 10...9 mbar |
| base pressure (NEG ion pumps) mfr | 10...11 mbar |
| base pressure (turbo pumping systems) mfr | 1 mbar |
| gauge measurement range mfr | 1 mbar |
| scroll pump maintenance interval mfr | 20...30 minutes |
| scroll pump maintenance period mfr | 2...3 years |
| Atomic Mass Unit | 1.6605 grams |
| Mounting | 20 |
| Shipping Weight | 3 lbs. |
| Emission Setting For 0.01 Amp/torr Sensitivity | 6.7 mA |
| Filament Current, Absolute Maximum | 6.0 Amps |
| Filament Voltage mfr | 4.0 VAC |
| Grid Voltage mfr | 6.3 VAC |
| Grid Current mfr | 8.7 Amps |
| Grid to Filament Voltage mfr | 700 VDC |
| Filament voltage mfr | 30 VDC |
| Grid to filament voltage for electron bombardment mfr | 700 VDC |
| Grid current for electron bombardment mfr | 100 DC |
| Electron bombardment power absolute maximum mfr | 50 W |
| Grid voltage for resistance heating mfr | 6.3 VAC |
| Grid current for resistance heating mfr | 8.7 Amps |
| Flange A dimension (2.75 in DN 40 ConFlat flange) mfr | 69.3 mm |
| Flange B dimension (2.75 in DN 40 ConFlat flange) mfr | 25.4 mm |
| Flange A dimension (NW25 KF) mfr | 40.1 mm |
| Flange B dimension (NW25 KF) mfr | 25.4 mm |
| Flange A dimension (NW40 KF) mfr | 55.1 mm |
| Flange B dimension (NW40 KF) mfr | 25.4 mm |
| Flange B dimension (1.00 in tube) mfr | 25.4 mm |
| Flange B dimension (0.75 in tube) mfr | 19.3 mm |
| Flange A dimension mfr | 55.1 mm |
| Flange B dimension mfr | 19.3 mm |
| Filament Temperature, Absolute Maximum | 1400 °C |
| Shield Coating mfr | Platinum, internally connected to filament filam |
| Standard pressure mfr | 101325 Pa |
Description
Construction and series positioning
The 571 is a wide-range, linear Bayard-Alpert gauge with a burn-out-resistant yttria-coated iridium filament. It is calibrated against McLeod gauge standards. Within the Agilent vacuum transducer series, it is the high-vacuum ionization gauge, distinct from the 531 thermocouple gauge (rough vacuum), the 572 tungsten-filament Bayard-Alpert gauge, and the 525 cold-cathode gauge. The 571 is offered in multiple mechanical configurations: 1 in glass (Nonex) and Kovar tubes, 0.75 in glass and Kovar tubes, 2.75 in DN 40 ConFlat, NW25 KF, NW40 KF, and a sealed version at 10⁻⁶ Torr. All variants use the same filament and electrical ratings.
Behavior under load and in different media
The gauge measures equivalent nitrogen pressure from 10⁻¹ Torr down to the X-ray limit of about 2×10⁻¹⁰ Torr. Maximum operating pressure is 1×10⁻³ Torr; continuous operation in argon at 1×10⁻¹ Torr does not affect filament life, but tube life in air may be shortened above 1×10⁻³ Torr. Sensitivity is composition-dependent: 10/Torr for nitrogen and air at 1.0 mA emission, and 1.5/Torr for helium at 6.7 mA emission. Ionic pumping speed is 0.06 L/s for N₂ at 1 mA, which can cause gauge pumping at low pressures. Near the X-ray limit, readings are not unequivocal and degassing effects can be mistaken for the limit.
Installation, connection, and operation
The tube may be mounted vertically or horizontally and operated/degassed in either position. It should be mounted as close as possible to the measurement location through a large tubulation (1 inch OD recommended) to avoid gauge pumping. The vacuum system must be grounded to a common ground with the controller, and the ground must be tested to support at least 10 A before operation and after each maintenance procedure. Normal operating ratings are 4.0 VAC at 3.5 A filament, +30 VDC filament, +180 VDC grid, and 1 mA grid current. Degassing can be by resistance heating (6.3 VAC, 8.7 A grid, 1200 °C grid temperature) or electron bombardment (up to +700 VDC grid-to-filament, 100 mA DC, 50 W maximum). High-temperature baking should not be attempted above 10⁻⁵ Torr. The gauge is not intrinsically safe and must not be operated with combustible gases present.
Integration & Operating Notes
The vacuum system and controller must share a common ground tested to carry at least 10 A. Inadequate grounding can expose lethal voltages up to 1000 V. Verify ground integrity after every maintenance or service procedure before applying power.
The gauge is rated for a maximum operating pressure of 1×10⁻³ Torr. Prolonged exposure above this in air shortens tube life; continuous operation in argon at 1×10⁻¹ Torr is permitted without affecting filament life. Select the gauge only for high-vacuum stages.
High-temperature baking of the gauge should not be attempted at pressures above 10⁻⁵ Torr. Baking at higher pressures can oxidize gauge elements and make ultrahigh vacuum readings difficult to attain. Plan bakeout only after the system is below this threshold.
Use only commercial controls equipped for resistance degassing and rated for 6.3 VAC, 8.7 A grid current, and 1200 °C grid temperature. Exceeding these absolute maximums can cause grid sagging. Electron bombardment degassing requires up to 700 VDC and must not exceed 50 W.
Low conductance between the gauge and the system causes gauge pumping, reading lower than actual system pressure, particularly below 10⁻⁸ Torr. Mount the tube as close as possible to the measurement location through a large tubulation, 1 inch OD recommended.
Resistance heating degassing typically requires one-half hour or more in the 10⁻⁹ Torr range. For measurements below 10⁻⁸ Torr, allow the gauge to degas overnight. Near the X-ray limit of 2×10⁻¹⁰ Torr, distinguish degassing effects from the true X-ray limit before trusting readings.
The 571 is not intrinsically safe. Do not turn on the gauge when combustible gases or mixtures are present. The hot filament and high voltage can act as ignition sources. Ensure the system is free of combustibles before operating.
Do not use silicone oil or silicone grease; silicone contamination degrades vacuum performance. Use powder-free butyl or polycarbonate gloves. If grease is necessary, use Apiezon L grease (Agilent Part Number 695400004).
Selection & Integration Q&A
Choose the 571 for high-vacuum measurement from 1×10⁻³ to 2×10⁻¹⁰ Torr with a burn-out-resistant yttria-coated iridium filament that tolerates accidental atmosphere exposure. The 572 uses a tungsten filament; the 525 cold-cathode gauge suits different pressure ranges. Select the 571 when filament robustness and linear Bayard-Alpert output are required.
Use a commercial controller equipped for resistance degassing and rated for 4.0 VAC/3.5 A filament, +30 VDC filament, +180 VDC grid, and 1 mA grid current. The vacuum system must share a common ground with the controller, tested to support at least 10 A before operation and after each service. Verify controller compatibility with these voltages.
The tube may be mounted and operated/degassed vertically or horizontally for prolonged periods. Mount close to the measurement point via a 1-inch OD tubulation to avoid gauge pumping below 10⁻⁸ Torr. Do not bake above 10⁻⁵ Torr, and do not operate with combustible gases present because the gauge is not intrinsically safe.
Plan for degassing: resistance heating typically takes one-half hour or more at 10⁻⁹ Torr, and overnight degassing is usually needed below 10⁻⁸ Torr. Accessories include a glass ion gauge connector kit and controller cables in 10 ft, 25 ft, and 50 ft lengths. Use Apiezon L grease (P/N 695400004) if grease is necessary; silicone grease is prohibited.