Agilent Technologies Refinery gas analyzer, four channel multi-dimensional system
Agilent Technologies Refinery gas analyzer, four channel multi-dimensional system — Gas chromatography.
The Agilent four-channel refinery gas analyzer is a multi-dimensional Micro GC configuration with three backflush channels (10 m CP-Molesieve 5 Å, 10 m CP-PoraPLOT U, 10 m CP-AL2O3/KCl) and one straight 8 m CP-Sil 5CB channel. It targets fixed gases, C1–C6 hydrocarbons and sulfur compounds in refinery gas streams, using argon carrier gas at 80 °C column temperature and 300 kPa column head pressure.
Agilent Technologies Energy Chemical GC Analyzers Series — Model Comparison & Selection
| Model | Typical quantitation limit, mol% | Carrier gas | Analysis time, min |
|---|---|---|---|
| Refinery gas analyzer, four channel multi-dimensional system Current | — | Argon | — |
| Analyzers for biodiesel, biofuels | 0.01 | — | — |
| Gasoline analyzer, benzene analysis | 0.01 | — | — |
| GCxGC-based analyzers, aviation turbine fuels | — | Hydrogen | — |
| natural gas analyzer, fast analysis times | 0.01 | — | 20 |
| NGA Analyzer, natural gas analysis | 0.01 | — | 20 |
Full specifications
| Carrier gas | Argon |
| Column temperature | 80 °C |
| Channel 1 column mfr | 10 backflush |
| Channel 1 compounds analyzed mfr | 2 |
| Channel 2 column mfr | 10 backflush |
| Channel 2 compounds analyzed mfr | 2 S |
| Channel 3 column mfr | 10 backflush |
| Channel 3 compounds analyzed mfr | 3 olefins |
| Channel 4 column mfr | 8 straight |
| Channel 4 compounds analyzed mfr | 6 hydrocarbons |
| Injection time mfr | 40 ms |
| Column head pressure mfr | 200 kPa |
| Depth | 32.94 cm |
| Height | 28.28 cm |
| Weight | 7.3 kg |
| Width | 14.5 cm |
| Operating temperature | 0 °C |
Description
Construction and positioning
- Four independent GC channels in one Micro GC chassis: three 10 m backflush columns (CP-Molesieve 5 Å RTS, CP-PoraPLOT U, CP-AL2O3/KCl) plus one 8 m CP-Sil 5CB straight column.
- This is the four-channel RGA configuration; the three-channel RGA configuration (backflush-to-detector on the AL2O3/KCl channel) is a separate variant of the same 990 Micro GC platform.
- Sibling analyzer packages (biodiesel, gasoline/benzene, GCxGC aviation fuels, complete RGA workflow) share the platform but differ in column sets and target compound lists.
Behavior under load and media
- Argon carrier gas; column temperature 80 °C; injector temperature 110 °C; column head pressure 300 kPa; injection time 40 ms.
- Channel 1 (backflush, 10 m) and Channel 2 (backflush, 10 m) each analyze 2 compounds; Channel 3 (detector, 10 m) analyzes 3 hydrocarbons; Channel 4 (straight, 8 m) analyzes 6 hydrocarbons.
- Backflush operation on Channels 1–3 protects the analytical columns from heavy ends; the straight Channel 4 provides the C6 hydrocarbon separation.
Installation, control and operation
- Site preparation follows the 990 Micro GC Site Preparation Checklist (G3588-90007, Rev 1.0, January 2022), which also covers the Mobile Micro GC Quad channel and power supply components.
- The application note (5994-1043EN) defines the two RGA configurations and their column sets; configuration must be fixed before ordering columns and method setup.
- Operating temperature range starts at 0 °C; unit mass 7.3 kg, width 14.5 cm, height 28.28 cm, depth 32.94 cm.
Integration & Operating Notes
The four-channel configuration is specified with argon carrier gas. Substituting another carrier changes response on the Molesieve and PoraPLOT U channels and breaks comparability with the published RGA configuration; confirm argon supply purity and delivery pressure before commissioning.
Bench space, power supply components and environmental conditions must be prepared per document G3588-90007 Rev 1.0 (January 2022), which covers the 990 Micro GC, Mobile Micro GC Quad channel and power supply. The checklist is the controlling document for utilities and clearance before installation.
The method uses a 40 ms injection time. The injection event must be synchronized across the four channels so that backflush and straight-column runs start together; any change to injection duration alters peak areas and must be re-validated against the RGA configuration.
Selection & Integration Q&A
Pick the four-channel version when the 8 m CP-Sil 5CB straight column is needed for the six-hydrocarbon separation alongside the three backflush channels. The three-channel RGA configuration uses a backflush-to-detector AL2O3/KCl channel and omits the straight Sil 5CB channel, so it covers fewer hydrocarbon compounds.
Supply argon carrier regulated to 300 kPa column head pressure. The method runs at 80 °C column temperature and 110 °C injector temperature with a 40 ms injection; the gas panel must hold pressure stable at the specified setpoint, otherwise retention times on all four channels drift.
Follow the 990 Micro GC Site Preparation Checklist G3588-90007 Rev 1.0 (January 2022), which defines bench space, power supply components and environmental conditions for the 990 Micro GC and Mobile Micro GC Quad channel. The unit operates from 0 °C ambient and weighs 7.3 kg at 14.5 cm width.
The base unit ships without the argon carrier supply, pressure regulation to 300 kPa, and the site-preparation items listed in G3588-90007 (power supply components, bench utilities). Column sets are configuration-specific: the four-channel RGA set comprises three 10 m backflush columns and one 8 m straight column, ordered as a defined configuration.