BINDER CB 260
BINDER CB 260 — Incubators with Temperature 5…50 °C.
The BINDER CB 260 is a 267 L CO2 incubator with a 200–230 V single-phase supply and a temperature range of +5 to 50 °C, distinguished within the CB series by its larger interior volume compared to the CB 170 and CB 56. It provides stable culture conditions with 0.1 K fluctuation and 0.4 K uniformity, and is intended for cell and tissue culture in research and production laboratories.
Direct drop-in replacements
BINDER CB Series — Model Comparison & Selection
Full specifications
| Temperature | 5...50 °C |
| CO2 recovery time | 6 min (all models) |
| Interior volume | 267 L (all models) |
| Load per rack | 10 kg (all models) |
| Nominal power | 1.5 kW (all models) |
| Permitted load | 40 kg (all models) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.4 K |
| Back wall clearance | 100 mm (all models) |
| Side wall clearance | 50 mm (all models) |
| Article Number mfr | 9040 |
| Recovery time after door was opened for 30 s at 37°C mfr | 6 min |
| Depth | 785 mm (all models) |
| Height | 1020 mm (all models) |
| Width | 740 mm (all models) |
| Power frequency | 50/60 Hz (all models) |
| Rated voltage | 200...230 V (CB260-230V / CB260-230V-O) |
| Unit fuse | 10.0 A (CB260-230V / CB260-230V-O) |
| Humidity range | 90...95 % RH |
Description
Construction and series positioning
- The CB 260 uses a 267 L interior chamber with net dimensions of 740 mm (W) × 1,020 mm (H) × 785 mm (D), making it the largest standard size in the CB series (siblings: CB 170 and CB 56).
- It is available in four versions: CB260-230V (standard, 200–230 V), CB260-230V-O (with O2 control 0.2–20 vol.-%), CB260UL-120V (100–120 V, NEMA 5-20), and CB260UL-120V-O (100–120 V with O2 control). This card describes the standard 230 V model (article 9040-0207).
- The chamber is equipped with a sterilizable NDIR CO2 sensor and the MB2 touchscreen program controller, which supports fixed-value, timer, and week programs, as well as alarm and notification functions.
Behavior under load and in operation
- Temperature control is maintained within ±0.1 K fluctuation and 0.4 K uniformity across the usable volume. After a 30 s door opening at 37 °C, the chamber recovers to setpoint within 6 minutes.
- The Permadry water pan provides humidity control in the range of 90 to 95 % RH, which is essential for preventing culture desiccation.
- The chamber supports a maximum load of 40 kg, with each rack rated for 10 kg. The interior volume of 267 L allows multiple shelves, but the load must be distributed evenly to avoid temperature stratification.
- Hot-air sterilization is available: a complete cycle lasts 10 hours (4 hours effective sterilization at ≥160 °C, followed by 6 hours cooling to 37 °C). CO2 control is automatically deactivated during sterilization and reactivated only after the chamber cools below 80 °C and the user confirms the sterilization success message.
Installation, connection, and control
- The chamber requires a single-phase 200–230 V, 50/60 Hz supply with a 10 A fuse. Nominal power is 1.5 kW. A wall clearance of 100 mm at the back and 50 mm at the sides is mandatory for adequate ventilation.
- Gas connections for CO2 (and O2/N2 for -O variants) are made at the rear. Gas cylinder connection kits are available as accessories. The CO2 sensor is factory-adjusted; after replacement, CO2 (and O2 if applicable) control adjustment must be repeated.
- The MB2 controller offers RS485 and Ethernet interfaces for network integration, as well as USB for data transfer and firmware updates. A zero-voltage relay output is provided for external alarm signaling; it resets when the alarm is acknowledged or the condition clears.
- The door is equipped with a DuoDoor lock; an optional divided inner door is available to reduce gas loss during sample access.
Integration & Operating Notes
Insufficient clearance restricts heat dissipation from the chamber walls and can cause overheating, reduced temperature uniformity, or premature component failure. Ensure the bench or floor stand can support the chamber weight and that the rear is accessible for gas and power connections.
The unit fuse is 10.0 A for 200–230 V models. Connecting to a supply outside the rated voltage range or using an undersized fuse will compromise safety and may damage the controller. Verify that the supply can deliver the nominal 1.5 kW without significant voltage drop.
During hot-air sterilization, CO2 and O2 control are automatically deactivated. They remain off until the chamber temperature falls below 80 °C and the user confirms the 'Successful sterilization' message. Do not expect gas control to function during this period; plan cell culture work accordingly.
The tolerance range alarm (e.g., ±2 °C around 37 °C) does not trigger until the setpoint has been reached at least once. This prevents false alarms during warm-up. Configure the alarm delay to avoid nuisance alarms during door openings or transient deviations.
Replacing the door gasket while the chamber is warm can deform or damage the gasket, leading to leaks and loss of temperature/humidity control. Allow the chamber to cool to ambient temperature before replacement. Also, after replacing the CO2 sensor, perform a CO2 control adjustment.
Perform electrical safety tests according to EN 50678/VDE 0701 and EN 50699/VDE 0702 before initial startup and after any maintenance or repair. Schedule at least annual maintenance; a maintenance message appears after 8760 operating hours or two years and cannot be permanently cleared by the user.
Selection & Integration Q&A
Choose the CB 260 when you need a larger working volume (267 L) for high-throughput culture or multiple users. The CB 170 (170 L) and CB 56 (56 L) are better suited for smaller labs or lower sample numbers. All share the same temperature range and controller features.
Use the gas cylinder connection kits (accessory) for CO2, O2, and N2. For standard CO2 control, only CO2 is required. For -O variants with O2 control, both O2 and N2 are needed. Ensure the gas supply pressure matches the specifications in the manual and that the connections are leak-tested.
The base delivery does not include gas cylinder connection kits, the BINDER-Water-Set for humidity, additional shelves, or the divided inner door. Also, annual electrical safety testing and sensor replacement (after 8760 h or 2 years) are not included. Order these accessories and plan service accordingly.