BINDER MKF 720 CO₂
BINDER MKF 720 CO₂ — Climate chambers with Temperature -48…180 °C.
The MKF 720 CO₂ is a 734 L alternating climate chamber with a working range of -48…180 °C and 10…98 % RH, extending 8 K lower than the -40 °C siblings in the MKF series. Its CO₂ (R 744) refrigeration circuit and 11 kW nominal power distinguish this 720-litre modification for large-volume humidity-controlled test programs.
BINDER MKF Series — Model Comparison & Selection
| Model | Temperature, °C | Height, cm | Interior volume, L | Load per rack, kg | Max heat compensation, W | Nominal power, kW |
|---|---|---|---|---|---|---|
| MKF 720 CO₂ Current | -48–180 | 200.5 | 734 | 40 | 1100 | 11 |
| MKF 1020 | -40–180 | 200.5 | 991 | 40 | 1500 | 12.5 |
| MKF 1020 CO₂ | -48–180 | 200.5 | 991 | 40 | 1400 | 11.5 |
| MKF 115 | -40–180 | 172.5 | 115 | 30 | 400 | 4.8 |
| MKF 240 | -40–180 | 171.5 | 228 | 30 | 400 | 6.8 |
| MKF 240 CO₂ | -50...180 | 171.3 | 228 | 30 | 900 | 6.8 |
| MKF 400 | -40–180 | 171 | 417 | 30 | 800 | 8 |
| MKF 400 CO₂ | -50...180 | 172.7 | 417 | 30 | 1600 | 7.9 |
| MKF 56 | -40–180 | 144.5 | 60 | 15 | 250 | 2.8 |
| MKF 720 | -40–180 | 200.5 | 734 | 40 | 1000 | 11 |
Full specifications
| Temperature | -48...180 °C |
| Interior volume | 734 L (MKF720CO2-400V, 50 Hz) |
| Load per rack | 40 kg (MKF720CO2-400V, 50 Hz) |
| Maximum heat compensation | 1100 W |
| Nominal power | 11 kW (MKF720CO2-400V, 50 Hz) |
| Article Number mfr | 9020 |
| Depth | 1140 mm (MKF720CO2-400V, 50 Hz) |
| Height | 2005 mm (MKF720CO2-400V, 50 Hz) |
| Width | 1580 mm (MKF720CO2-400V, 50 Hz) |
| Power frequency | 50 Hz (MKF720CO2-400V, 50 Hz) |
| Rated voltage | 400 V |
| Unit fuse | 25.0 A (MKF720CO2-400V, 50 Hz) |
| Humidity range | 10...98 % RH |
| Cooling down time from 180°C to -40°C mfr | 60 min |
| Humidity fluctuation with humidity dependent on set value mfr | 2 % r.F. |
| Average heating-up rate according to IEC 60068-3-5 between -40°C and 180°C mfr | 5.1 K/min |
Description
Construction and series positioning
- The MKF 720 CO₂ is the 734 L member of the MKF (E6) alternating climate chamber series; siblings include the 115, 240 and 1020 litre models, all rated -40…180 °C, while this CO₂ variant reaches -48 °C.
- The chamber is built around a CO₂ (R 744) refrigeration circuit with 2.20 kg refrigerant fill and a maximum operating pressure of 120 bar; repairs to this circuit are currently restricted to BINDER.
- Net dimensions are 1,580 mm wide × 2,005 mm high × 1,140 mm deep, with a 40 kg load limit per rack.
- The 720-litre interior volume is the second-largest in the series, below the 1020 and above the 240 and 115.
Behavior under load and in different media
- Temperature uniformity varies from 0.1 to 1.0 K and fluctuation from 0.1 to 0.2 K depending on set value; with humidity active, the temperature range narrows to 10…95 °C and fluctuation widens to 0.1…1 K.
- Humidity control spans 10…98 % RH with a fluctuation of 2 % r.h.; tolerated error under simultaneous condensation is ±2 % r.h.
- Average heating rate per IEC 60068-3-5 between -40 °C and 180 °C is 5.1 K/min; cooling from 180 °C to -40 °C takes 60 min.
- Maximum heat compensation is 1,100 W, defining the permissible dissipated load from test specimens.
Installation, connection and control
- Electrical supply is 400 V 3 N/PE 50 Hz with 11 kW nominal power and a 25.0 A unit fuse; the type plate example also lists 6.50 kW / 12.0 A, so protection must be sized for the higher value on the actual unit.
- A type B residual current circuit breaker is mandatory, and supply voltage must stay within ±10 % of nominal at the indicated frequency.
- The MB2 chamber controller provides fixed-value, timer, time-program and week-program modes, a chart-recorder display, alarm and notification functions, and user management with password-protected authorization levels.
- Humidity control requires a freshwater supply (automatic via water pipe or manual via internal can) and a wastewater connection; a condensate recycling lever and hose burst/reflux safety kit are available as options.
- Temperature safety is provided by an over-temperature protective device class 1 and a safety controller class 2.0 per DIN 12880, with an optional over/under-temperature safety device class 2.
Selection & Integration Q&A
Pick the 720 CO₂ when you need 734 L of usable volume and the extra 8 K cold margin to -48 °C for humidity-controlled programs; the 240 is smaller but stops at -40 °C, while the 1020 adds volume at the same -48 °C only in its CO₂ version.
Humidity control needs a freshwater supply (automatic via pipe or manual via internal can) and a wastewater drain. A connection kit for the freshwater pipe and an optional hose burst/reflux safety kit are available; without them the humidification system cannot operate.
No. The room volume must be calculated per EN 378-1:2016+A1:2020 for the 2.20 kg CO₂ charge, unventilated recesses are prohibited, and minimum wall clearances must be kept. IP 20 also means the unit must not be wetted.
Annual maintenance, steam humidifier service at least yearly, condenser cleaning every 1–2 years, and the BINDER Service reset of the 8760-hour message are not included. Door gaskets, the Pt 100 sensor, RS485 and analog output options, and APT-COM 4 software must be ordered separately.