This operating manual (issue 09/2026, Art. No. 7001-0026) covers BINDER BD/BD-UL incubators with natural convection, ED/ED-UL drying and heating ovens with natural convection, and FD/FD-UL drying and heating ovens with forced convection, all with microprocessor temperature controller. Covered models: BD 23 (BD023-230V, Art. 9010-0187/9110-0187), BD 23-UL (BD023UL-120V, 9010-0189/9110-0189), BD 400 (BD400-230V, 9010-0073/9110-0073), BD 400-UL (BD400UL-120V, 9010-0176/9110-0176), ED 23 (ED023-230V, 9010-0190/9110-0190 and 9010-0191/9110-0191), ED 23-UL (ED023UL-120V, 9010-0192/9110-0192 and 9010-0193/9110-0193), ED 400 (ED400-230V, 9010-0075/9110-0075), ED 400-UL (ED400UL-208V, 9010-0168/9110-0168), FD 23 (FD023-230V, 9010-0194/9110-0194), FD 23-UL (FD023UL-120V, 9010-0196/9110-0196). Options include disconnectable audible over-temperature alarm, APT-COM 4 software, additional Pt100 sensor (BD), analog temperature output, and suction plant connection.
Section-by-Section Technical Breakdown
Specifies power plug types, current types, and voltages for each model variant. ED 400 uses a 5-pole CEE plug with 3N~; UL models use NEMA plugs at 115 V or 208 V. Connection must follow local regulations.
- BD 23, BD 400, ED 23, FD 23: grounded plug, 1N~
- ED 400: CEE plug 5 poles, 3N~
- UL models: NEMA 5-15P at 115 V/60 Hz (1N~) or NEMA L21-20P at 208 V/60 Hz (3N~) for ED 400-UL
- Observe national regulations when connecting
Covers turning on the chamber, heating operation display, and air change. Includes main power switch operation for ED 400 and display indications.
- Turning on the chamber
- Heating operation display
- Air change
Describes temperature set-point entry with and without ramp function, time functions (continuous and timer operation), and user level settings including °C/°F change, ramp entry, and chamber addressing.
- Display/entry of temperature set-point without ramp
- Display/entry of temperature set-point with selected temperature ramp
- Switching between continuous operation and timer operation
- Setting tempering time in timer operation
- Temperature unit change between °C and °F
- Entering a temperature ramp
- Chamber addressing
Describes temperature safety device class 2 (DIN 12880) for ED and FD, and class 3.1 (DIN 12880) for BD (option for ED, FD). These provide overtemperature monitoring with visual and audible signals.
- Class 2 safety device for ED, FD
- Class 3.1 safety device for BD, optional for ED/FD
- Visual and audible (buzzer) signals indicate temperature exceeding
Lists optional equipment: disconnectable audible over-temperature alarm, APT-COM 4 Multi Management Software for BD and ED, additional Pt100 temperature sensor for BD, and analog output for temperature.
- Disconnectable audible over-temperature alarm (option)
- APT-COM 4 Multi Management Software (accessory for BD and ED)
- Additional Pt100 temperature sensor (option for BD)
- Analog output for temperature (option)
Provides procedures for cleaning and chemical disinfection/decontamination of the chamber. Emphasizes preventing water penetration and using appropriate agents.
- Cleaning procedures
- Decontamination / chemical disinfection
- Do not allow chamber to become wet
Covers maintenance intervals (at least annually), personnel qualification, simple troubleshooting by operating personnel, and requirements for repair and electrical testing. Includes contact information for BINDER service.
- Maintenance at least once a year by licensed electricians or BINDER-authorized experts
- Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 recommended
- Simple troubleshooting does not require technical intervention
- Repair only by BINDER Service or qualified service partners
- Replace door gasket only when cold
Provides factory calibration and adjustment information, definition of usable volume, overcurrent protection, and detailed technical data for BD, ED, and FD models, plus equipment/options extracts and accessories/spare parts.
- Factory calibration and adjustment
- Definition of usable volume
- Over current protection
- BD technical data
- ED technical data
- FD 23 technical data
- Equipment and options extracts for BD, ED, FD
- Accessories and spare parts extract
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| BD 23 nominal temperature 100 °C (212 °F), nominal power 0.20 kW, nominal current 0.9 A, nominal voltage 230 V / 50 Hz or 230 V / 60 Hz, current type 1N~. p.8 ↗ | Verify power supply matches 230 V ±10% at indicated frequency; ensure circuit can supply 0.9 A. | BD 23 |
| Temperature safety device class 3.1 (DIN 12880) for BD; class 2 (DIN 12880) for ED and FD. p.29 ↗ | Select model with appropriate safety class for application; class 3.1 provides additional protection. | BD (class 3.1); ED, FD (class 2) |
| IP protection type 20 according to EN 60529. p.8 ↗ | Chamber is protected against solid objects >12.5 mm but not water; do not install in damp areas or expose to splashing water. | All models |
| Electrical connection: ED 400 requires CEE plug 5 poles, 3N~; ED 400-UL requires NEMA L21-20P, 208 V at 60 Hz, 3N~; other UL models use NEMA 5-15P, 115 V at 60 Hz, 1N~. p.19 ↗ | Ensure appropriate receptacle and supply configuration for the specific model variant. | ED 400, ED 400-UL, UL models |
| Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. p.36 ↗ | Schedule annual maintenance to maintain warranty and safety. | All models |
| Annual repeat electrical inspection and test prior to initial startup and after maintenance/repair recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.36 ↗ | Include electrical safety testing in maintenance schedule. | All models |
| Replace door gasket only when cold; otherwise gasket may be damaged. p.36 ↗ | Allow chamber to cool before gasket replacement. | All models |
| Inner chamber, pre-heating chamber, and inside of doors made of stainless steel 1.4301 (AISI 304). p.15 ↗ | Avoid corrosive substances (acids, halides) that may damage stainless steel. | All models |
| ED and FD at temperatures above 150 °C / 302 °F may cause discoloration of metallic surfaces due to oxidation; harmless and does not impair function. p.15 ↗ | Discoloration is normal at high temperatures; no action required. | ED, FD |
| BD indicates temperature with accuracy of a tenth of a degree; ED and FD indicate with accuracy of one degree. p.15 ↗ | Choose BD for applications requiring finer temperature readout. | BD vs ED/FD |
| Chambers are not classified as medical devices and are not qualified for sterilization of medical devices. p.12 ↗ | Do not use for medical device sterilization. | All models |
| Non-ionizing radiation: maintain 30 cm distance between field source and active implants (pacemakers, defibrillators) to exclude influence. p.14 ↗ | Personnel with active implants should keep 30 cm distance. | All models |
| Chamber must not be installed in unventilated recesses; observe prescribed minimum distances (chap. 3.4). p.9 ↗ | Ensure sufficient ventilation and clearance to prevent overheating. | All models |
| Serial interface RS 422 for computer communication via APT-COM 4 software (accessory for BD and ED). p.15 ↗ | Use RS 422 interface for remote monitoring/control with APT-COM 4. | BD, ED (option) |
| Additional Pt100 temperature sensor available as option for BD. p.32 ↗ | Can add independent temperature measurement point. | BD |
| Analog output for temperature available as option. p.33 ↗ | Enables connection to external data logging or control systems. | All models (option) |
| Disconnectable audible over-temperature alarm available as option. p.32 ↗ | Provides audible warning of overtemperature; can be disconnected if not needed. | All models (option) |
| Chamber fuse may respond; check and replace if appropriate. If it responds again, contact BINDER service. p.37 ↗ | A blown fuse may indicate fault; do not repeatedly replace without investigating. | All models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Chamber without function. p.37 ↗ | No power supply. | Check connection to power supply. ED 400: Check whether the chamber is turned on at the main power switch. | Chapter 10.3 |
| Chamber without function. p.37 ↗ | Wrong voltage. | Check power supply for correct voltage (chap. 4.1). | Chapter 10.3 |
| Chamber without function. p.37 ↗ | Chamber fuse has responded. | Check chamber fuse and replace it if appropriate. If it responds again, contact BINDER service. | Chapter 10.3 |
| Chamber without function. p.37 ↗ | Controller defective. | Contact BINDER service. | Chapter 10.3 |
| Set-point temperature is not reached after specified time. p.37 ↗ | Chamber door not properly closed. | Completely close chamber door. | Chapter 10.3 |
| Set-point temperature is not reached after specified time. p.37 ↗ | Door gasket defective. | Replace door gasket. | Chapter 10.3 |
| Set-point temperature is not reached after specified time. p.37 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chapter 10.3 |
| Set-point temperature is not reached after specified time. p.37 ↗ | FD: The fan doesn’t turn or turns too slowly. Fan defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber heating permanently, set-point not held. p.37 ↗ | Controller defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber heating permanently, set-point not held. p.37 ↗ | Pt 100 sensor defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber heating permanently, set-point not held. p.37 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber heating permanently, set-point not held. p.37 ↗ | Controller not adjusted. | Calibrate and adjust controller. | Chapter 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is lit. p.37 ↗ | Heating element defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is lit. p.37 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.37 ↗ | Timer run off. | Program the timer or change to time function Continuous operation (chap. 6.3). | Chapter 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.37 ↗ | Semiconductor relay defective. | Contact BINDER service. | Chapter 10.3 |
| Chamber doesn’t heat up. Red heating control light in the display is not lit. Controller display working. p.37 ↗ | Controller defective. | Contact BINDER service. | Chapter 10.3 |