BINDER BD 260
BINDER ·Made in Germany · In Active Production

BINDER BD 260

BINDER BD 260 — Incubators with Temperature 5…100 °C.

The BD 260 is a 253 L natural-convection incubator with a +5 to 100 °C working range, 0.2 K fluctuation and 0.3 K uniformity, sized between the BD 115 and BD 400 siblings. Its 0.85 kW single-phase 230 V design and 75 min heat-up to 100 °C suit standard 37 °C incubation and warm storage where forced-air airflow is undesirable.

Temperature
5…100 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER BD Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeating up time to 100 c, minHeight, cmInterior volume, LLoad per rack, kg
BD 260 Current 5–10076759425340
BD 115 5–10060.59070.511230
BD 23 5–100525949.52012
BD 400 5–10076.5120102.540035
BD 56 5–10056.57562.55730
BD 720 5–100879015973745

Full specifications

Core Performance & Functional Specifications
Temperature 5...100 °C
Heating-up time to 100 °C 75 min
Interior volume 253 L (BD260-230V / BD260UL-120V)
Load per rack 40 kg (BD260-230V / BD260UL-120V)
Nominal power 0.85 kW
Permitted load 270 kg (BD260-230V / BD260UL-120V)
Temperature fluctuation 0.2 K
Temperature uniformity 0.3 K
Article Number mfr 9010
Recovery time after door was opened for 30 s at 37°C mfr 16 min
Dimensions, Weight & Mechanical Design
Depth 760 mm (BD260-230V / BD260UL-120V)
Height 940 mm (BD260-230V / BD260UL-120V)
Width 810 mm (BD260-230V / BD260UL-120V)
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 230 V
Unit fuse 8.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Behavior under load and in operation

Installation, connection and control

Integration & Operating Notes

Electrical
External 8.0 A time-lag fuse must be provided upstream

The chamber does not carry its own branch protection for this size; the 8.0 A T 5×20 mm / 250 V fuse is external. Size the circuit and fuse accordingly, and verify supply voltage stays within ±10 % of 230 V nominal before connection.

Quick Start Guide, Electrical connection data and installation, p.18
Process Limits
Load usable volume to no more than half with 10 % wall clearance

Natural convection depends on free air movement; overloading or placing samples closer than 10 % of the interior dimensions to walls prevents homogeneous temperature distribution and invalidates the 0.3 K uniformity figure. Respect 40 kg per rack and 270 kg total.

Quick Start Guide, Placing the charging material in the usable volume, p.19
Control & I/O
Safety controller set-point must track every temperature change

The safety controller acts as an absolute maximum limit; set it 2–5 °C above the process set-point and re-check after every set-point change. Its permissible range is 10–100 °C on BD. After an overtemperature trip the chamber must cool below the reset threshold before restart.

Quick Start Guide, Set-point entry for temperature / Setting the safety controller set-point, p.20–22
Mounting
Install indoors at 18–40 °C ambient, max 2000 m altitude

Operation outside +18 to +40 °C ambient, above 70 % r.h. non-condensing, or above 2000 m altitude is outside approved conditions. Do not install in unventilated recesses or explosive areas; the 260 size requires four people lifting at the lower corners.

Quick Start Guide, Requirements for the location of installation, p.18; Lifting and transporting, p.17
Maintenance
Annual electrical inspection and BINDER-authorized maintenance

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; repeat electrical testing per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended before first start and after repair. Replace the door gasket only when cold.

Operating Manual BD-ED-FD (E2), Sections 10.1–10.2, p.36
Maintenance
Hot-air disinfection limited to 100 °C on BD

BD chambers can be hot-air disinfected at 100 °C / 212 °F, not sterilized at 190 °C as the ED/FD ovens can. Remove all flammable goods first. Use neutral cleaning agent 1002-0016 and disinfectant spray 1002-0022; avoid neutral cleaner on zinc-coated hinge parts and rear wall.

Quick Start Guide, Cleaning and decontamination, p.23

Selection & Integration Q&A

When should I choose the BD 260 rather than the BD 115 or BD 400?

All BD siblings share +5 to 100 °C, so pick by volume and load: the 260 offers 253 L with 270 kg total and 40 kg per rack. Choose the 115 for smaller batches or the 400 when 253 L and the 810 mm width are insufficient; the 260 fits between them.

What upstream electrical provision and control interface does the BD 260 need?

A 230 V 1N~ 50 Hz supply within ±10 %, protected by an external 8.0 A time-lag 5×20 mm fuse, connected via the IEC/shockproof plug. For remote monitoring, the optional RS 422 interface works with APT-COM 4; an analog temperature output and extra Pt100 are also optional.

What ambient and duty limits apply when installing the BD 260 in a lab?

Install indoors on a non-flammable surface at +18 to +40 °C, ≤70 % r.h. non-condensing, ≤2000 m altitude, with ventilation and no unventilated recess. Keep 10 % wall clearance and load no more than half the usable volume; the 260 needs four people to lift.

What maintenance and consumables are required beyond the base delivery?

Schedule annual maintenance by BINDER-authorized personnel plus repeat electrical testing per EN 50678/50699. Consumables not included: 8.0 A T fuses, door gasket (replace cold), neutral cleaner 1002-0016 and disinfectant 1002-0022. RS 422/APT-COM 4, Pt100 and alarm are options.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model BD 260 en
Datasheet ·668,2 КБ ·7 pp.
View PDF
PDF
QuickStartGuide BD BF ED FD FED E31 03 2023 dt en fr
Manual ·1,7 МБ ·35 pp.
View PDF
PDF
BD ED FD (E2) 09 2026 en
Manual ·12,9 МБ ·71 pp.
View PDF