BINDER BD 56
BINDER BD 56 — Incubators with Temperature 5…100 °C.
The BD 56 is a 57 L natural-convection incubator (BD Avantgarde.Line, Art. 9010-0323/0324) with a 5–100 °C working range and 0.1 K display resolution. It is the mid-size standard model between the BD 23 and BD 115, distinguished by class 3.1 overtemperature protection and 0.3 K uniformity. It is intended for incubation of cultures at 37 °C and warm storage of liquids.
Direct drop-in replacements
BINDER BD Series — Model Comparison & Selection
Full specifications
| Temperature | 5...100 °C |
| Heating-up time to 100 °C | 75 min |
| Interior volume | 57 L (BD056-230V / BD056UL-120V) |
| Load per rack | 30 kg (BD056-230V / BD056UL-120V) |
| Nominal power | 0.3 kW |
| Permitted load | 70 kg (BD056-230V / BD056UL-120V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.3 K |
| Back wall clearance | 160 mm (BD056-230V / BD056UL-120V) |
| Side wall clearance | 100 mm (BD056-230V / BD056UL-120V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 15 min |
| Depth | 565 mm (BD056-230V / BD056UL-120V) |
| Height | 625 mm (BD056-230V / BD056UL-120V) |
| Width | 560 mm (BD056-230V / BD056UL-120V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 6.3 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Stainless-steel interior (1.4301/AISI 304) with glass door and natural convection; no fan, so airflow is buoyancy-driven and gentle.
- Net interior 560 × 625 × 565 mm (W×H×D), 57 L usable volume; 70 kg total permitted load, 30 kg per rack.
- Within the BD series (23, 56, 115, 260, 400) the 56 is the mid-size unit; all share the 5–100 °C range, but the 56 uses the 0.30 kW / 6.3 A electrical class and class 3.1 safety per DIN 12880.
- Two voltage variants: BD056-230V (230 V 1~ 50/60 Hz, CEE 7/7) and BD056UL-120V (120 V 1~ 60 Hz, NEMA 5-15).
Behaviour under load and in operation
- Temperature fluctuation 0.1 K and uniformity 0.3 K; heating from ambient to 100 °C takes 75 min.
- Recovery after a 30 s door opening at 37 °C is 15 min — plan door-access frequency accordingly.
- Controller resolves 0.1 °C; the safety controller (class 3.1) has a set-point range of 10–100 °C and must be set 2–5 °C above the working set-point.
- Loading rule: keep samples at least 10 % of interior dimensions from walls and fill no more than half the usable volume, otherwise natural-convection uniformity degrades.
Installation, connection and operation
- Indoor, non-flammable surface; ambient 18–40 °C, ≤70 % r.h. non-condensing, altitude ≤2000 m; supply tolerance ±10 %.
- Clearances: 160 mm at the rear, 100 mm at the sides; do not install in unventilated recesses.
- IP 20, pollution degree 2, over-voltage category II (IEC 61010-1); external 6.3 A time-lag 5×20 mm fuse required upstream.
- Lift size 56 with two people; never by the door or handle. Hot-air disinfection at 100 °C is permitted for BD; racks are removable and autoclavable.
Integration & Operating Notes
The chamber has no internal miniature circuit breaker; protection is an external 6.3 A time-lag (T) 5×20 mm / 250 V fuse. Size the branch circuit and verify supply within ±10 % of 230 V (or 120 V for the UL variant) before first start-up.
Natural convection depends on free air circulation around the housing. Installing closer than the specified clearances, or in a non-ventilated recess, risks overheating and invalid temperature uniformity; the manual explicitly forbids unventilated recesses.
Samples must remain inside the usable volume with at least 10 % of interior dimensions as wall clearance. Overloading blocks the buoyancy-driven airflow, so the 0.3 K uniformity and 0.1 K fluctuation figures no longer apply to the load.
The class 3.1 safety controller is an absolute maximum limit (10–100 °C range). Set it 2–5 °C above the process set-point and re-check after every set-point change; otherwise a legitimate set-point change can trip the overtemperature protection and cut heating.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, including repeat electrical tests per EN 50678/VDE 0701 and EN 50699/VDE 0702. Replace the door gasket only when the chamber is cold to avoid damaging it.
Unlike ED/FD ovens (190 °C sterilization), the BD 56 permits hot-air disinfection only at 100 °C. Remove all flammable goods first. Racks may be autoclaved; use neutral cleaner 1002-0016 and disinfectant 1002-0022, and never apply neutral cleaner to zinc-coated hinge parts or the rear wall.
Selection & Integration Q&A
Pick the 56 when 57 L accommodates your rack load (≤30 kg/rack, 70 kg total) at 37 °C. The 23 is smaller; the 115 gives more volume but the same 5–100 °C range and 0.1 K resolution. All share natural convection, so size is the deciding factor.
A 230 V 1~ 50/60 Hz supply within ±10 % (or 120 V/60 Hz for BD056UL-120V) protected by an external 6.3 A time-lag 5×20 mm fuse. Nominal power is 0.30 kW; the chamber has no internal circuit breaker, so the fuse is the only overcurrent protection.
No. The manual forbids unventilated recesses. Install indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance, ambient 18–40 °C, ≤70 % r.h. non-condensing, altitude ≤2000 m, and IP 20 means no water exposure.
Annual maintenance by BINDER-authorized personnel plus repeat electrical tests (EN 50678/50699). Consumables: 6.3 A time-lag fuses, door gasket (replace cold), neutral cleaner 1002-0016 and disinfectant 1002-0022. APT-COM 4, Pt100 sensor and analog output are options, not included.