BINDER BD 720 — Incubators with Temperature 5…100 °C.
The BD 720 is a 737 L natural-convection incubator with a +5 to 100 °C working range, 0.1 K display resolution and 0.7 K uniformity, distinguished within the BD series by its size-720 chamber and 1.65 kW single-phase heating. It serves incubation of cultures at 37 °C and precise tempering of harmless materials in large-volume laboratory workflows.
Direct drop-in replacements
BINDER BD Series — Model Comparison & Selection
Full specifications
| Temperature | 5...100 °C |
| Heating-up time to 100 °C | 90 min |
| Interior volume | 737 L (BD720-230V / BD720UL-240V) |
| Load per rack | 45 kg (BD720-230V / BD720UL-240V) |
| Nominal power | 1.65 kW |
| Permitted load | 315 kg (BD720-230V / BD720UL-240V) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.7 K |
| Back wall clearance | 160 mm (BD720-230V / BD720UL-240V) |
| Side wall clearance | 100 mm (BD720-230V / BD720UL-240V) |
| Article Number mfr | 9010 |
| Recovery time after door was opened for 30 s at 37°C mfr | 23 min |
| Depth | 870 mm (BD720-230V / BD720UL-240V) |
| Height | 1590 mm (BD720-230V / BD720UL-240V) |
| Width | 1165 mm (BD720-230V / BD720UL-240V) |
| Power frequency | 50 Hz |
| Rated voltage | 230 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
The BD 720 is the largest chamber in the BD Avantgarde.Line incubator series, which also includes the BD 23, BD 115, BD 260 and BD 400. All share the +5 to 100 °C range and natural-convection architecture, but the BD 720 provides 737 L of interior volume in a 1,165 mm wide, 1,590 mm high, 870 mm deep cabinet. Unlike the forced-convection BF series, it has no fan; temperature distribution relies on natural convection, giving gentle airflow suited to incubation. The standard variant is BD720-230V (Art. 9010-0331, 230 V 1~ 50/60 Hz, CEE 7/7 plug); a UL variant BD720UL-240V (Art. 9010-0332, 240 V, NEMA 6-20 plug) exists for North America.
Thermal behavior
Heating from ambient to 100 °C takes 90 minutes at 1.65 kW nominal power. Temperature fluctuation is 0.1 K and uniformity 0.7 K across the usable volume, with display resolution of 0.1 °C. After a 30 s door opening at 37 °C, recovery takes 23 minutes. The overtemperature protective device class 1 trips at 120 °C, and the safety controller (class 3.1 per DIN 12880 for BD) has a settable range of 10–100 °C with 0.1 °C accuracy. Loading is limited to 45 kg per rack and 315 kg total; the usable volume must not be loaded beyond half, and samples need at least 10 % of interior dimensions as wall clearance.
Installation and control
Install indoors on a non-flammable surface with 160 mm rear and 100 mm side clearance; ambient must stay within +18 to +40 °C, ≤70 % r.h. non-condensing, at ≤2000 m altitude. Supply tolerance is ±10 % of nominal voltage; the chamber fuse is external 16.0 A. The controller supports temperature set-point entry, timer/continuous operation, and °C/°F selection. Options include an additional Pt100 sensor, analog temperature output, RS 422 interface for APT-COM 4 software, and a disconnectable audible over-temperature alarm.
Integration & Operating Notes
The chamber must not be lifted by the door handle or door. Use a fork lifter applied only from the rear in the middle of the chamber, and transport only in original packaging with straps; incorrect lifting can damage the cabinet or cause injury.
Natural convection depends on free airflow. Load the usable volume to no more than half, maintain at least 10 % of interior dimensions as wall clearance, and respect 45 kg per rack and 315 kg total to preserve the specified 0.7 K uniformity.
The safety controller set-point is an absolute maximum limit; set it approximately 2 °C to 5 °C above the working temperature and re-check after every set-point change. Its permissible range on BD is 10–100 °C with 0.1 °C accuracy; values outside this range cannot be entered.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts. Annual repeat electrical inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is recommended, and the door gasket must only be replaced when the chamber is cold.
Selection & Integration Q&A
Choose the BD 720 when you need 737 L of usable volume for large batch incubation at 37 °C; the BD 400 and BD 260 offer the same +5 to 100 °C range and natural convection but smaller chambers. The BD 720 requires 1.65 kW and 160 mm rear clearance, so verify room space and supply capacity.
Provide 230 V 1~ 50/60 Hz with ±10 % tolerance and an external 16.0 A time-lag (T) 5x20 mm / 250 V chamber fuse. The 1.65 kW load should be on a dedicated circuit; the CEE 7/7 plug is standard, and the UL variant BD720UL-240V uses a NEMA 6-20 plug at 240 V.
Annual maintenance by BINDER-authorized personnel and electrical safety testing per EN 50678/EN 50699 are not included. Optional accessories include the additional Pt100 sensor, analog temperature output, RS 422/APT-COM 4 software, and disconnectable audible over-temperature alarm; cleaning agents 1002-0016 and 1002-0022 are ordered separately.