KB ECO (E6.2) 07 2026

Technical Reference Summary

Covered Models & Document Scope

Covers KB ECO and KB ECO-UL cooling incubators with RD4 controller and Peltier (thermoelectric) technology, model sizes 240, 400, 720 and 1020. 230 V variants: KBECO240-230V (9020-0423, 9120-0423), KBECO400-230V (9020-0465, 9120-0465), KBECO720-230V (9020-0424, 9120-0424), KBECO1020-230V (9020-0425, 9120-0425). UL variants (100-120 V): KBECO240UL-120V (9020-0426, 9120-0426), KBECO400UL-120V (9020-0466, 9120-0466), KBECO720UL-120V (9020-0427, 9120-0427), KBECO1020UL-120V (9020-0428, 9120-0428). Document edition E6.2, issue 07/2026, Art. No. 7001-0422.

Section-by-Section Technical Breakdown

pp. 22-24
Installation and Connections

Covers spacers for wall distance, the flexible tilt protection kit for size 400 chambers, the condensate collection pan and the electrical connection. Installation must respect the prescribed minimum distances and the ambient conditions of chap. 3.4.

  • Spacers are provided for wall distance (chap. 4.1).
  • Flexible tilt protection kit must be mounted on chambers of size 400 (chap. 4.2).
  • Condensate collection pan is part of the installation (chap. 4.3).
  • Electrical connection per chap. 4.4; chamber must not be installed in unventilated recesses.
pp. 25-27
Functional Overview of the RD4 Chamber Controller

Describes the RD4 controller menu structure and access levels and the controller performance during and after power failures. This is the basis for all set-point, alarm and network configuration described in later chapters.

  • Menu structure and access levels (chap. 5.1).
  • Performance during and after power failures (chap. 5.2).
pp. 28-29
Start Up and Temperature Set-Point Entry

Covers chamber start-up and entry of the temperature set-point on the RD4 controller. Set-point entry is the prerequisite for automatic setpoint switchover and tolerance/alarm functions.

  • Start-up procedure (chap. 6).
  • Temperature set-point entry (chap. 7.1).
pp. 29-34
Automatic Setpoint Switchover

Describes automatic switching between temperature set-points, either controlled by date or by a regular cyclic program, plus entry of the alternative temperature set-point and activation/deactivation of the Idle mode function.

  • Control by date (chap. 8.1).
  • Regular cyclic control (chap. 8.2).
  • Entry of the alternative temperature set-point (chap. 8.3).
  • Activating/deactivating Idle mode (chap. 8.4).
pp. 35
Special Controller Functions – Idle Mode

Describes the Idle mode special controller function. Idle mode is activated/deactivated via the automatic setpoint switchover menu (chap. 8.4).

  • Idle mode is a special controller function (chap. 9).
  • Activation/deactivation is described in chap. 8.4.
pp. 36-37
Password

Covers the controller password request and how to assign and modify the User password and the Admin password. Passwords protect the controller access levels.

  • Password request (chap. 10.1).
  • Assign and modify the User password (chap. 10.2.1).
  • Assign and modify the Admin password (chap. 10.2.2).
pp. 38-40
Temperature Safety Devices

Describes the over temperature protective device (class 1) and the overtemperature safety controller class 3.1, including setting the safety controller mode and value, the message and measures in the alarm state, and the function check.

  • Over temperature protective device class 1 (chap. 11.1).
  • Overtemperature safety controller class 3.1 (chap. 11.2).
  • Setting the safety controller mode (chap. 11.2.1) and value (chap. 11.2.2).
  • Message and measures in the state of alarm (chap. 11.2.3).
  • Function check (chap. 11.2.4).
pp. 41-44
General Controller Settings

Covers controller menu language, temperature unit, date and time setting, language selection at restart, chamber address and display brightness.

  • Menu language (chap. 12.1) and temperature unit (chap. 12.2).
  • Current date (chap. 12.3) and time (chap. 12.4).
  • Language selection at restart (chap. 12.5).
  • Chamber address (chap. 12.6) and display brightness (chap. 12.7).
pp. 44-45
Tolerance Range Settings

Covers the delay time for the tolerance range alarm and the temperature tolerance range setting. These settings define when a tolerance alarm is triggered.

  • Setting the delay time for tolerance range alarm (chap. 13.1).
  • Setting the temperature tolerance range (chap. 13.2).
pp. 45-47
Notification and Alarm Functions

Describes information messages, alarm messages and activation/deactivation of the audible alarm buzzer.

  • Information messages (chap. 14.1).
  • Alarm messages (chap. 14.2).
  • Activating/deactivating the audible alarm (chap. 14.3).
pp. 47-48
Program Start / Stop Function

Covers the start time program, start week program and stop program, plus controller performance after power failures.

  • Start time program (chap. 15.1).
  • Start week program (chap. 15.2).
  • Stop program (chap. 15.3).
  • Performance after power failures (chap. 15.4).
pp. 48-53
Ethernet Network Settings

Covers showing the network settings (MAC address, IP address, subnet mask, standard gateway, DNS server address, DNS chamber name) and changing the network configuration (automatic/manual IP and DNS assignment, IP address, subnet mask, standard gateway, DNS server address).

  • Showing MAC address, IP address, subnet mask, standard gateway, DNS server address and DNS chamber name (chap. 16.1.1–16.1.6).
  • Selecting automatic/manual IP address assignment (chap. 16.2.1) and DNS server assignment (chap. 16.2.2).
  • Assigning IP address (chap. 16.2.3), subnet mask (chap. 16.2.4), standard gateway (chap. 16.2.5) and DNS server address (chap. 16.2.6).
pp. 54-55
Data Recorder

Describes the recorded data, storage capacity, setting the storage rate for the DL1 recorder data and deleting the data recorder.

  • Recorded data (chap. 17.1).
  • Storage capacity (chap. 17.2).
  • Setting the storage rate for DL1 recorder data (chap. 17.3).
  • Deleting the data recorder (chap. 17.4).
pp. 56-58
USB Menu: Data Transfer via USB Interface

Covers connecting a USB stick, the import function, export functions, ongoing data transfer, errors during data transmission and removing the USB stick.

  • Connecting the USB stick (chap. 18.1).
  • Import function (chap. 18.2) and export functions (chap. 18.3).
  • Ongoing data transfer (chap. 18.4) and error during data transmission (chap. 18.5).
  • Removing the USB stick (chap. 18.6).
pp. 58-59
Self-Test Function

Describes activation and deactivation of the controller self-test function.

  • Activating the self-test function (chap. 19.1).
  • Deactivating the self-test mode (chap. 19.2).
pp. 60
Defrosting at Refrigerating Operation and Anti-Condensation Protection

Covers defrosting during refrigerating operation and the anti-condensation protection function.

  • Defrosting at refrigerating operation (chap. 20).
  • Anti-condensation protection (chap. 21).
pp. 61
Options and Accessories

Lists options and accessories, including the APT-COM™ 4 Multi Management Software accessory for networking up to 100 chambers and connecting them to a PC.

  • APT-COM™ 4 Multi Management Software (accessory) permits networking up to 100 chambers and connecting them to a PC (chap. 22.1).
pp. 61-63
Cleaning and Decontamination

Covers cleaning and decontamination/chemical disinfection of the chamber.

  • Cleaning (chap. 23.1).
  • Decontamination / chemical disinfection (chap. 23.2).
pp. 64-67
Maintenance and Service, Troubleshooting, Repair, Testing

Covers personnel qualification, maintenance intervals and service, the service reminder, simple troubleshooting, and returning the chamber to BINDER GmbH. Regular maintenance is required at least once a year; electrical repeat inspection is required annually and before initial start-up and before resuming operation after maintenance or repair.

  • Maintenance work at least once a year by licensed electricians or experts authorized by BINDER (chap. 24.2).
  • Annual repeat electrical inspection and test before initial start-up and before resuming operation after maintenance or repair; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702 (chap. 24.1).
  • Replace the door gasket only when cold (chap. 24.2).
  • With increased dust in ambient air, clean the Peltier fan grid (by suction or blowing) several times a year (chap. 24.2).
  • Service Reminder: keep OK pressed 5 seconds to display remaining days until service; 'Service due!' message reappears after one week of operation (chap. 24.3).
  • Simple troubleshooting by operating personnel without technical intervention or disassembly (chap. 24.4).
pp. 71-78
Technical Description

Covers factory calibration and adjustment, over current protection, definition of usable volume, KB ECO / KB ECO-UL technical data, equipment/options/accessories, accessories and spare parts, and dimensions.

  • Factory calibration and adjustment (chap. 26.1).
  • Over current protection (chap. 26.2).
  • Definition of usable volume (chap. 26.3).
  • KB ECO / KB ECO-UL technical data (chap. 26.4).
  • Equipment, options and accessories extract (chap. 26.5); accessories and spare parts extract (chap. 26.6).
  • Dimensions (chap. 26.7).

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Nominal temperature of the chamber is 70 °C / 158 °F. p.1 ↗ Maximum permissible operating temperature; do not exceed when selecting set-points. KB ECO / KB ECO-UL series
IP protection is 20 according to EN 60529. p.1 ↗ Chamber is not protected against water; must be set up splash-proof and must not become wet during operation, cleaning or maintenance. KB ECO / KB ECO-UL series
Temperature safety device class 3.1 according to DIN 12880:2007. p.1 ↗ Overtemperature safety controller class 3.1 must be configured and function-checked per chap. 11.2. KB ECO / KB ECO-UL series
230 V models: nominal power 1.40 kW, nominal current 5.8 A, nominal voltage range 200-230 V / 50 Hz (+/-10%) and 208-230 V / 60 Hz, 1 N~. p.1 ↗ Electrical supply and fusing must be sized for 5.8 A at 200-230 V. KB ECO 240/400/720/1020 230 V variants
UL models: nominal power 0.90 kW, nominal current 7.5 A, nominal voltage 100-120 V / 50 Hz and 100-120 V / 60 Hz, 1 N~. p.1 ↗ UL variants require a 100-120 V supply sized for 7.5 A. KB ECO 240-UL/400-UL/720-UL/1020-UL
Intended temperature application range for cultivation of microorganisms is 4 °C / 39.2 °F to 37 °C / 98.6 °F. p.1 ↗ Typical applications: long-term storage at 4 °C, refrigerated incubation at 20-25 °C, incubation at 37 °C, and alternating temperatures (e.g. 37 °C and 4 °C). KB ECO / KB ECO-UL series
Chamber must not be installed in unventilated recesses; prescribed minimum distances must be observed (chap. 3.4). p.1 ↗ Ensure sufficient ventilation for heat dispersal to avoid overheating damage. KB ECO / KB ECO-UL series
Maintenance work must be performed at least once a year by licensed electricians or experts authorized by BINDER. p.24 ↗ Plan an annual maintenance interval; warranty becomes void if maintenance is performed by non-authorized personnel. KB ECO / KB ECO-UL series
Annual repeat electrical inspection and a test prior to initial start-up and prior to resuming operation after maintenance or repair are required; recommended per EN 50678/VDE 0701 and EN 50699/VDE 0702. p.24 ↗ Include electrical safety testing in commissioning and annual service schedule. KB ECO / KB ECO-UL series
Replace the door gasket only when cold. p.24 ↗ Replacing the gasket while warm may damage it. KB ECO / KB ECO-UL series
With increased dust in the ambient air, clean the Peltier fan grid (by suction or blowing) several times a year. p.24 ↗ Dusty environments require more frequent fan-grid cleaning to maintain Peltier cooling performance. KB ECO / KB ECO-UL series
Service Reminder: keep the OK button pressed for 5 seconds to display the remaining days until maintenance is due; after the recommended one-year interval the message 'Service due!' appears and reappears after one week of operation. p.24 ↗ Use the controller display to track service due dates; confirm messages with OK. KB ECO / KB ECO-UL series
APT-COM™ 4 Multi Management Software permits networking up to 100 chambers and connecting them to a PC. p.2 ↗ Plan Ethernet network addressing and chamber addresses (chap. 12.6, chap. 16) for multi-chamber installations. KB ECO / KB ECO-UL series
Chamber must not be operated in potentially explosive areas; no explosion protection measures are provided. p.1 ↗ Do not introduce combustible or explosive substances, combustible dust or air-solvent mixtures into the chamber. KB ECO / KB ECO-UL series
Loading material must not contain corrosive ingredients (in particular acids and halides) that may damage stainless steel, aluminum and copper components. p.1 ↗ Corrosive damage caused by such ingredients is excluded from BINDER liability. KB ECO / KB ECO-UL series
The chamber is not classified as a medical device per Regulation (EU) No 2017/745 and is not qualified to perform sterilization of medical devices. p.1 ↗ Do not use for medical-device sterilization applications. KB ECO / KB ECO-UL series
Flexible tilt protection kit must be mounted on chambers of size 400. p.4 ↗ Size 400 installations require the tilt protection kit per chap. 4.2. KB ECO 400 / KB ECO 400-UL
Data recorder storage rate for DL1 recorder data is configurable; recorded data and storage capacity are described in chap. 17. p.17 ↗ Set the storage rate according to required recording resolution and available storage capacity. KB ECO / KB ECO-UL series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Service due! p.65 ↗ Recommended maintenance interval (one year of operation) has elapsed. Press the OK button to confirm the message; arrange maintenance. The message reappears after one week of operation. chap. 24.3