BINDER KB ECO 240
BINDER KB ECO 240 — Incubators with Temperature 0…70 °C.
The KB ECO 240 is a 247 L thermoelectric (Peltier) cooling incubator with a 0–70 °C working range and ±0.1 K uniformity, built for microbiological cultivation and long-term storage. Within the KB ECO series it is the mid-size 230 V modification (200–230 V, 0.9 kW nominal), distinguished from the 400/720/1020 sizes by volume and from the UL variant by supply voltage. It serves as a stand-alone temperature-controlled chamber in laboratory workflows requiring stable, contamination-free incubation.
Direct drop-in replacements
BINDER KB ECO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB ECO 240 Current | 0–70 | 79.6 | 146.1 | 247 | 650 | 30 |
| KB ECO 1020 | 0–70 | 114.9 | 192.5 | 1020 | 973 | 45 |
| KB ECO 400 | 0–70 | 79.6 | 194.6 | 400 | 650 | 30 |
| KB ECO 720 | 0–70 | 88.5 | 192.5 | 700 | 973 | 45 |
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 247 L |
| Load per rack | 30 kg |
| Max. heat compensation | 350 W |
| Nominal power | 0.9 kW |
| Permitted load | 100 kg (both variants) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.1 K |
| Back wall clearance | 100 mm (both variants) |
| Side wall clearance | 100 mm (both variants) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 3 min |
| Depth | 796 mm (both variants) |
| Height | 1461 mm (both variants) |
| Interior width | 650 mm (both variants) |
| Width | 925 mm (both variants) |
| Net weight of the unit (empty) mfr | 152 kg |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- The KB ECO 240 uses Peltier thermoelectric cooling rather than a compressor circuit, giving a 247 L interior (650 mm wide, 796 mm deep, 1461 mm high) in a 925 mm wide cabinet.
- It is the 240-size member of the KB ECO family; siblings 400, 720 and 1020 share the 0–70 °C range but differ in chamber volume. The 240 is the smallest 230 V variant listed and the only one without a tilt-protection requirement (that kit applies to size 400).
- Two supply variants exist: KBECO240-230V (200–230 V, CEE 7/7) and KBECO240UL-120V (100–120 V, NEMA 5-20). This card covers the 230 V model.
Behavior and working range
- The intended biological application band is 4 °C to 37 °C, covering refrigerated storage, 20–25 °C incubation and 37 °C cultivation, as well as alternating set-points (e.g. 37 °C / 4 °C) via automatic setpoint switchover.
- Temperature fluctuation and uniformity are each specified at ±0.1 K for the 230 V variant; maximum heat compensation is 350 W, which bounds the heat load the Peltier stage can offset at low set-points.
- The chamber is rated to a nominal 70 °C; set-points must not exceed this. The overtemperature safety controller is class 3.1 per DIN 12880:2007, with a class 1 over-temperature protective device as backup.
Control, installation and operation
- The RD4 chamber controller manages set-point entry, tolerance-range alarms with configurable delay, automatic date- or cycle-based setpoint switchover, Idle mode, a DL1 data recorder, Ethernet networking and USB import/export.
- Installation requires 100 mm clearance at the rear and sides and prohibits unventilated recesses; the unit is IP 20 and must be kept splash-proof.
- Electrical supply must be sized for the 230 V variant's nominal current and 16 A unit fuse; annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required.
Integration & Operating Notes
The 230 V model draws a nominal current that must be accommodated by the branch circuit and the 16 A unit fuse. Undersized supply wiring or shared circuits will trip under Peltier run-up; verify voltage stays within 200–230 V at 50 Hz or 208–230 V at 60 Hz.
Loading material must not contain acids or halides that attack stainless steel, aluminum or copper components, and the chamber has no explosion protection. Corrosive damage from such ingredients is excluded from BINDER liability, so verify sample chemistry before loading.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, and an electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial start-up and after any repair. Non-authorized maintenance voids the warranty.
With increased dust in ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year. A clogged grid reduces heat rejection and can prevent the chamber from reaching low set-points such as 4 °C.
Holding the OK button for 5 seconds shows remaining days until maintenance is due; after the one-year interval the message 'Service due!' appears and reappears weekly. This lets operators schedule service without external tracking.
Selection & Integration Q&A
Pick the 240 when 247 L interior (650 mm wide) fits your batch and you want the smallest 230 V KB ECO. Step up to 400/720/1020 only for larger loads; note the 400 additionally requires the flexible tilt protection kit, which the 240 does not.
The 230 V variant requires 200–230 V / 50 Hz or 208–230 V / 60 Hz, 1 N~, with a 16 A unit fuse. Size the branch circuit for the nominal current stated on the type plate and keep voltage within tolerance to avoid Peltier controller faults.
Only with 100 mm clearance at the rear and both sides and never in an unventilated recess. The Peltier fan grid rejects heat to ambient; blocked airflow causes overheating and loss of cooling at low set-points. IP 20 also means keep it splash-proof.
Annual maintenance by BINDER-authorized personnel plus electrical repeat inspection per EN 50678/VDE 0701 are mandatory. For networking up to 100 chambers, APT-COM™ 4 software is a separate accessory; USB stick and Ethernet cabling are not included.