BINDER KB ECO 720 — Incubators with Temperature 0…70 °C.
The KB ECO 720 is a 700 L thermoelectric cooling incubator (Peltier technology) with a 0–70 °C working range and 0.1 K uniformity, positioned as the mid-large chamber between the 400 and 1020 siblings. Its 500 W heat compensation and 45 kg per-rack load capacity suit high-density incubation and long-term storage at 4–37 °C.
Direct drop-in replacements
BINDER KB ECO Series — Model Comparison & Selection
| Model | Temperature, °C | Depth, cm | Height, cm | Interior volume, L | Interior width, mm | Load per rack, kg |
|---|---|---|---|---|---|---|
| KB ECO 720 Current | 0–70 | 88.5 | 192.5 | 700 | 973 | 45 |
| KB ECO 1020 | 0–70 | 114.9 | 192.5 | 1020 | 973 | 45 |
| KB ECO 240 | 0–70 | 79.6 | 146.1 | 247 | 650 | 30 |
| KB ECO 400 | 0–70 | 79.6 | 194.6 | 400 | 650 | 30 |
Full specifications
| Temperature | 0...70 °C |
| Interior volume | 700 L |
| Load per rack | 45 kg |
| Max. heat compensation | 500 W |
| Nominal power | 1.3 kW |
| Permitted load | 150 kg (both variants) |
| Temperature fluctuation | 0.1 K |
| Temperature uniformity | 0.1 K |
| Back wall clearance | 100 mm (both variants) |
| Side wall clearance | 100 mm (both variants) |
| Article Number mfr | 9020 |
| Recovery time after door was opened for 30 s at 37°C mfr | 4 min |
| Depth | 885 mm (both variants) |
| Height | 1925 mm (both variants) |
| Interior width | 973 mm (both variants) |
| Width | 1250 mm (both variants) |
| Net weight of the unit (empty) mfr | 272 kg |
| Power frequency | 50 Hz |
| Rated voltage | 200...230 V |
| Unit fuse | 16.0 A |
| Phase (Nominal voltage) mfr | 1 |
Description
Construction and series positioning
- Peltier (thermoelectric) cooling/heating, no compressor; interior volume 700 L, net width 1,250 mm, depth 885 mm, height 1,925 mm.
- Sits between KB ECO 400 and KB ECO 1020 in the series; all share the 0–70 °C range, but the 720 offers 700 L with 45 kg per rack and 150 kg total permitted load.
- Two voltage variants: 230 V (200–230 V, 50/60 Hz, CEE 7/7) and UL 100–120 V (NEMA 5-20); this model code covers the 230 V standard version (Art. 9020-0424).
Behavior under load and in different modes
- Temperature fluctuation and uniformity are both 0.1 K; recovery after a 30 s door opening at 37 °C is 4 min.
- Maximum heat compensation is 500 W, allowing the chamber to hold setpoint when warm samples are introduced; nominal power is 1.3 kW.
- Intended application range for microorganism cultivation is 4–37 °C, covering refrigerated storage, 20–25 °C incubation and 37 °C incubation, plus alternating temperature programs via automatic setpoint switchover.
- Defrosting and anti-condensation protection functions are available for refrigerating operation.
Installation, control and operation
- RD4 chamber controller with menu-driven setpoint entry, automatic setpoint switchover (date or cyclic), Idle mode, tolerance alarms, and password-protected access levels.
- Temperature safety device: over-temperature protective device class 1 and overtemperature safety controller class 3.1 per DIN 12880:2007, with alarm state and function check.
- Ethernet interface for network configuration and APT-COM™ 4 Multi Management Software (up to 100 chambers); USB interface for data export/import; DL1 data recorder with configurable storage rate.
- Installation requires 100 mm clearance at the back and sides; must not be installed in unventilated recesses. IP20: splash-proof setup required.
Integration & Operating Notes
The 230 V variant draws a nominal current of 5.8 A and 1.3 kW nominal power; the unit fuse is 16 A. Ensure the branch circuit and upstream protection are rated for this continuous load and that the CEE 7/7 plug is compatible with the site outlet. UL variants (100–120 V, 7.5 A) are separate models.
The chamber must not be installed in unventilated recesses; prescribed minimum distances of 100 mm at the rear and sides must be observed to allow heat dispersal from the Peltier cooling system. Failure to do so risks overheating damage and invalidates performance.
Loading material must not contain acids or halides that attack stainless steel, aluminum or copper. The chamber has no explosion protection and must not be used with combustible dust, solvents or explosive atmospheres. It is not a medical device per EU 2017/745 and must not be used for sterilization.
Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts; warranty is void if non-authorized personnel perform it. An electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702 is required before initial start-up and after any maintenance or repair. The service reminder displays remaining days when OK is held for 5 s.
With increased dust in the ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year to maintain cooling performance. The door gasket must only be replaced when cold; replacing it warm may damage it.
Selection & Integration Q&A
Choose the 720 when you need 700 L interior volume and up to 150 kg total load with 45 kg per rack, between the 400 and 1020. All share 0–70 °C and 0.1 K uniformity; the 720 balances capacity and footprint (1,250 mm wide).
The 230 V model requires 200–230 V, 50/60 Hz, 1-phase supply with nominal current 5.8 A and nominal power 1.3 kW; unit fuse is 16 A. Use a CEE 7/7 outlet. The UL variant (100–120 V, 7.5 A, NEMA 5-20) is a separate model.
Install with 100 mm clearance at back and sides, not in unventilated recesses. IP20 means splash-proof setup only. Ambient conditions per chap. 3.4 apply; dusty air requires frequent fan-grid cleaning. Do not operate in explosive areas.
Annual maintenance by BINDER-authorized personnel and electrical repeat inspection per EN 50678/50699 are mandatory. APT-COM™ 4 software (up to 100 chambers) and USB stick for data export are optional accessories not included. Service reminder is built-in.