BINDER KB ECO 400 — Incubators with Temperature 0…70 °C.

The KB ECO 400 is a 400 L thermoelectric cooling incubator (Peltier technology) with a 0–70 °C working range, part of the KB ECO series alongside the 240, 720 and 1020 sizes. Its 650 mm interior width and 120 kg permitted load distinguish this mid-size modification, which is intended for refrigerated incubation and long-term storage of microorganisms at 4–37 °C in laboratory systems.

Temperature
0…70 °C

Direct drop-in replacements

⚖️ Compare
⚖️ Compare
⚖️ Compare
⚖️ Compare

BINDER KB ECO Series — Model Comparison & Selection

Model Temperature, °CDepth, cmHeight, cmInterior volume, LInterior width, mmLoad per rack, kg
KB ECO 400 Current 0–7079.6194.640065030
KB ECO 1020 0–70114.9192.5102097345
KB ECO 240 0–7079.6146.124765030
KB ECO 720 0–7088.5192.570097345

Full specifications

Core Performance & Functional Specifications
Temperature 0...70 °C
Interior volume 400 L
Load per rack 30 kg
Max. heat compensation 350 W
Nominal power 1.2 kW
Permitted load 120 kg (KBECO400-230V / KBECO400-120V)
Temperature fluctuation 0.1 K
Temperature uniformity 0.1 K
Back wall clearance 100 mm (KBECO400-230V / KBECO400-120V)
Side wall clearance 100 mm (KBECO400-230V / KBECO400-120V)
Article Number mfr 9020
Recovery time after door was opened for 30 s at 37°C mfr 6 min
Dimensions, Weight & Mechanical Design
Depth 796 mm (KBECO400-230V / KBECO400-120V)
Height 1946 mm (KBECO400-230V / KBECO400-120V)
Interior width 650 mm (KBECO400-230V / KBECO400-120V)
Width 925 mm (KBECO400-230V / KBECO400-120V)
Net weight of the unit (empty) mfr 196 kg
Electrical, Power & Drive Specifications
Power frequency 50 Hz
Rated voltage 200...230 V
Unit fuse 16.0 A
Phase (Nominal voltage) mfr 1

Description

Construction and series positioning

Thermal behaviour and load

Control, installation and operation

Integration & Operating Notes

Mounting
Size 400 requires the flexible tilt protection kit

The KB ECO 400 must be fitted with the flexible tilt protection kit described in chapter 4.2 before operation; omitting it leaves the tall 1,946 mm cabinet without the prescribed anti-tilt safeguard, which is a safety-relevant installation defect.

Instruction Manual KB ECO (E6.2), chap. 4.2, p. 22
Mounting
Keep 100 mm clearance at back and sides, never recessed

The chamber needs 100 mm wall clearance at the back and sides and must not be installed in unventilated recesses; the Peltier cooling rejects heat to the ambient air, so restricted airflow causes overheating and loss of cooling performance.

Instruction Manual KB ECO (E6.2), chap. 1.8 and 3.4
Electrical
Supply and fusing must be sized for the actual voltage variant

The 230 V variant draws 5.8 A at 200–230 V/50 Hz (1.2 kW nominal, 16 A unit fuse), while the UL 120 V variant draws 7.5 A at 100–120 V; the branch circuit, plug type (CEE 7/7 vs NEMA 5-20) and protective device must match the ordered variant.

Data Sheet KB ECO 400; Instruction Manual chap. 1.6 Type plate
Control & I/O
Configure and function-check the class 3.1 safety controller

The overtemperature safety controller class 3.1 per DIN 12880:2007 must be set to the correct mode and value and function-checked before routine use; an unconfigured safety controller will not protect the load if the RD4 setpoint or sensor fails.

Instruction Manual KB ECO (E6.2), chap. 11.2, p. 38
Process Limits
Do not load corrosive or explosive materials

Loading material must not contain acids or halides that attack stainless steel, aluminum or copper, and the chamber must not be used in explosive atmospheres or with combustible substances; such use voids BINDER liability and can damage the Peltier heat exchangers.

Instruction Manual KB ECO (E6.2), chap. 1.8–1.9
Maintenance
Annual maintenance by authorized personnel is mandatory

Maintenance must be performed at least once a year by licensed electricians or BINDER-authorized experts, with an annual electrical repeat inspection per EN 50678/VDE 0701 and EN 50699/VDE 0702; the controller's Service Reminder shows days remaining and displays 'Service due!' after one year.

Instruction Manual KB ECO (E6.2), chap. 24.1–24.3, pp. 64–65
Cooling & Media
Clean the Peltier fan grid several times a year in dusty air

With increased dust in the ambient air, the Peltier fan grid must be cleaned by suction or blowing several times a year; a clogged grid reduces heat rejection and can cause the chamber to lose cooling capacity or overheat.

Instruction Manual KB ECO (E6.2), chap. 24.2, p. 65

Selection & Integration Q&A

When should I choose the KB ECO 400 rather than the 240, 720 or 1020 models?

Pick the 400 when 400 L interior volume and a 120 kg permitted load (30 kg per rack) fit your batch, since all KB ECO sizes share the 0–70 °C range and 0.1 K uniformity. The 240 suits smaller loads; the 720 and 1020 are for larger batches but need more floor space and the same 100 mm clearances.

What supply and protective device does the KB ECO 400 need, and does the plug type differ between variants?

The 230 V variant (Art. 9020-0465) needs 200–230 V/50 Hz, draws 5.8 A at 1.2 kW nominal and uses a CEE 7/7 plug with a 16 A unit fuse. The UL 120 V variant (Art. 9020-0466) needs 100–120 V, draws 7.5 A and uses a NEMA 5-20 plug; size the branch circuit accordingly.

What installation and environmental limits apply to the KB ECO 400?

It must stand with 100 mm clearance at the back and sides, never in an unventilated recess, and the flexible tilt protection kit for size 400 must be mounted. IP 20 means no water exposure, and explosive atmospheres or corrosive loads (acids, halides) are prohibited.

What maintenance and accessories are required beyond the base delivery?

Annual maintenance by BINDER-authorized personnel and an annual electrical repeat inspection are mandatory; the Service Reminder tracks the one-year interval. The Peltier fan grid needs cleaning several times a year in dusty air. APT-COM™ 4 software (up to 100 chambers) and the DL1 data recorder functions are separate accessories/options.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Data Sheet Model KB ECO 400 en
Datasheet ·499,2 КБ ·7 pp.
View PDF
PDF
KB ECO (E6.2) 07 2026 en
Manual ·9,3 МБ ·91 pp.